| Alternating Current Output |
| Range |
3 to 30 amperes RMS, programmable |
| Resolution |
1 amp |
| Output Frequency |
Sinewave synthesis, 50 Hz to 400Hz |
| Compliance Voltage |
3 volts RMS |
| Regulation |
±0.25% no load to full load |
| Resistance Measurement |
| Resistance Limit |
User programmable max. and min. Test is automatically aborted when limit is exceeded. |
| Resistance Range |
1W at 3 amps linearly decreasing to 100 mW at 30A |
| Resistance Resolution |
10mW |
| Measurement Accuracy |
± 1.0% of reading; ± 0.5% of range |
| Accuracy |
Specifications apply for 1 year, ± 5°C from the cal temp within 20°C to 30°C |
| Calibration |
Certificate of calibration traceable to the National Institute of Standards and Technology (NIST) provided |
| Test Time |
Digital count down timer, programmable 1 to 9999 seconds |
| Terminal Configuration |
True four-wire Kelvin |
| General Specifications |
| Remote Interface |
VICL and serial com standard. PLC optional. |
| Display |
40 character dot matrix illuminated liquid crystal display |
| Indicators |
L.E.D. status indicators |
| Alarm |
Audible and visual alarms provide pass/fail indication |
| Diagnostics |
Built-in self diagnostics test ROM, RAM and output system |
| Safety |
CE mark certified to EN61010 |
| Accessories |
K-2, VIC-3, BBL-2 and operator's manual supplied with unit |
| Warranty |
Extended 5 year parts and labor with annual factory calibration and maintenance. Standard warranty one year. |
| Operating Temperature |
0°C to 50°C |
| Humidity |
90% RH 0°C to 40°C |
| Power |
115/230 VAC ± 10% 250 VA max |
| Dimensions |
89mmH x 432mmW x 457mmD (3.5"H x 17"W x 18"D) |
| Weight |
11Kg (24 lb.) Net/15Kg (34 lb.) Shipping |