Simpson F45 Temperature Digital Panel Meter

Click image to enlarge
|
|
Features:
- Full 4-1/2 Digit, Bright Red 0.56" (14.2mm) Display
- User-Programmable T/C or RTD type
(J, K, E, T, R, S, and Platinum 100 RTD)
- Jumper-Selectable Display Indicates ºC or ºF With
0.1º or 1º Display Resolution
- Linear mV Scale for Easy Calibration
- Short 2.94" (74.7mm) Deep, 1/8 DIN Case
- Screw Terminal Connector for Easy Installation
- Optional Isolated 9-32VDC Power Supply
|
Description
| The Falcon Series temperature indicator
is a high accuracy, microprocessor-based panel instrument designed
to provide the maximum flexibility in temperature measurement. The
Falcon automatically compensates for differences between unit temperature
(room temperature) and ice point (0º), and provides the complete
NIST range for six thermocouple types as well as Platinum 100 RTD.
Input ranges are userselectable and can be changed without recalibration.
The Falcon series features a standard 1/8 DIN case with a depth
of less than 3”. Screw terminals are standard for easy installation. |
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| F45 |
4-½ Digit Falcon Process Indicator Panel |
| Call |  |
|
Ordering Options
Specifications
| DISPLAY |
| Type |
7-segment, red LED |
| Height |
0.56" (14.2mm) |
| Decimal Point |
Jumper-selectable 2-position (corresponding
to resolution desired) |
| Overrange indication |
Most significant digit = “1”; other digits
blank |
| Polarity |
Automatic, with "-" indication, "+" indication
implied |
| POWER REQUIREMENTS |
| AC Voltages |
120 or 220VAC, ±10% 50/60Hz |
| DC Voltages |
9-32VDC, ±1% |
| Power Consumption |
3VA |
| ENVIRONMENTAL |
| Operating Temperature |
0 to 55ºC |
| Storage Temperature |
-10 to 60ºC |
| Relative Humidity |
0 to 85% non-condensing |
| Warm-up Time |
Less than 20 minutes |
| MECHANICAL |
| Bezel |
3.78"x 1.89" x .44"(96 x 48 x 11.2mm) |
| Depth |
2.94" (74.7mm) |
| Panel Cut-out |
3.62" X 1.77"(91.9 x 45mm 1/8 DIN) |
| Case Material |
94V-1, UL rated Noryl® |
| Weight |
9.0oz (255.1g) |
| INPUTS |
| Thermocouple |
J, K, E, T, R, and S |
| RTD |
Platinum 100 |
| Millivolt |
±84mV reading of uncompensated mV |
| Cold Junction Compensation Error |
0.1ºC/ºC |
| Input Impedance |
10M (typical) |
| Lead Resistance Effect |
4.0µV/100 |
| Conversion Rate |
2-1/2 times per second |
| Open Thermocouple Detection |
-1 on display, -40nA bias on thermocouple |
| Source Type |
Temperature
Range |
Conformance
Error |
Resolution |
| E |
-200 to 1000ºC
328 to 1832ºF |
±0.1% of
rdg ±0.1ºC
±0.1% of rdg ±0.2ºF |
0.1º or 1º
User-Selectable |
| J |
-200 to 1200ºC
-328 to 2192ºF |
±0.1% of
rdg ±0.1ºC
±0.1% of rdg ±0.2ºF |
0.1º or 1º
User-Selectable |
| K |
-200 to 1372ºC
-328 to 2501ºF |
±0.1% of
rdg ±0.1ºC
±0.1% of rdg ±0.2ºF |
0.1º or 1º
User-Selectable |
| T |
-200 to 400ºC
-328 to 752ºF |
±0.1% of
rdg ±0.1ºC
±0.1% of rdg ±0.2ºF |
0.1º or 1º
User-Selectable |
| R |
-50 to 1768ºC
-58 to 3214ºF |
±0.1% of
rdg ±0.2ºC
±0.1% of rdg ±0.4ºF |
1º
Automatic |
| S |
-50 to 1768ºC
-58 to 3214ºF |
±0.1% of
rdg ±0.2ºC
±0.1% of rdg ±0.4ºF |
1º
Automatic |
| mV |
-19.999 to
+4.000 mV |
±0.02%
of rdg*
±0.002 mV |
0.001 or 0.01
mV*
User-Selectable |
| RTD Pt 100 |
-200 to 850ºC
-328 to 1562ºF |
±0.06%
of rdg ±0.1ºC
±0.06% of rdg ±0.2ºF |
0.1º or 1º
User-Selectable |
* Usable resolution of 0.002mV.
Conformance is to NIST
Monograph 175.
|
| Temperature-Electromotive Force
Reference Functions for the Letter-Designated Thermocouple types based on the ITS-90. |
|