| Withstanding voltage testing |
| Test Voltage |
Output Voltage: |
0 ~ 2.5 kV/0 ~ 5 kV, AC 2 ranges |
| Output capacity: |
500 VA (5 kV,100 mA) |
| >30 min. if output is more than 50 mA |
| Wave shape: |
Commercial power sign wave |
| Power voltage fluctuation: |
<15% (No load to maximum load) |
| Voltage output: |
Zero cross switch |
| Voltage output setting: |
Voltage slider adjust |
| Voltage measurement |
Voltage: |
Average current display |
| Analog: |
Scale = AC0 ~ 5 kV |
| Accuracy: |
±5% (JIS C1102) |
| Digital: |
Scale = AC 0.00 ~ 6.00 kV Green LED |
| Display output voltage during the testing, and show the
test voltage used after testing. |
| Measurement accuracy: |
± 1.5% of f.s. |
| Current measurement |
Current |
Average current display |
| Measure range |
0.01~ 199.9mA, 2 ranges |
| Display |
Digital 3 and 1/2 digits display Green LED |
| Display the leak current during the test, and show the leak
current at the evaluation after the testing. |
| Resolution: |
0.01 mA(0.1~ 9.9mA),
0.1 mA(10.0 ~ 110.0mA) |
| Accuracy |
± (5%+20μ) of Maximum setting |
| Test result evaluation |
Evaluation method |
Maximum limit analog comparator, Maximum,
minimum limit digital comparator |
| Maximum limit |
0.1~110.0mA
(Minimum limit + more than +1 digit) |
| Minimum limit: |
0.0 ~109.0mA
(Maximum limit - less than -1 digit) |
| Resolution: |
0.1 mA, Enable minimum limit setting ON/OFF |
| Evaluation factor: |
Maximum limit setting > leak current > Minimum limit setting - GOOD |
| Maximum limit setting ≤ Leak current -- NG |
| Minimum limit setting ≥ Leak current -- NG |
| Testing time |
Setting range: |
0.5 ~999 sec.(Timer off capable) |
| Setting resolution: |
0.1 sec (0.5 ~99.9 sec.),
1 sec. (< 100 sec.) |
| Time display: |
Timer ON: Display the remaining time
|
| Timer OFF: Display the progressed time |
| Accuracy: |
± 20 msec(0.5 ~99.9 sec.),
±200 msec. (<100sec.) |