PIE Model 512 Read & Source RTD Calibrator
   Search: HACKER SAFE certified sites prevent over 99.9% of hacker crime.
   
   
   



PIE Model 512 Read & Source RTD Calibrator



PIE Model 512
Click image to enlarge

PIE 512

Datasheet
Instructions

Features:
  • Accurate to ±0.25°C (±0.4°F) with 0.1° Resolution
  • Simulate/Read RTD sensors
  • Calibrate/Read directly in temperature for your RTD curve or Ω
  • Select from 7 RTD types or Ω (custom curves available)
  • Patent Pending Detection of 2, 3, or 4 Wire Connection Indicators
  • Guaranteed to Work with All Pulsed Instruments
  • Works with a wide variety of transmitters including popular Rosemount and Honeywell Models
  • Compatible with devices using pulsed excitation currents including PLCs, DCS, Recorders, and all others
  • EZ-Dial Knob Easily adjust output for full temperature range by 0.1°
  • EZ-Check Switch for easy 0% and 100% span adjustments
  • Store new EZ-Check values by pressing the EZ-Dial Knob
  • Uses a standard 9V Alkaline Battery
  • Superior battery life of 45 hours under typical continuous usage
  • Easy access to battery compartment
  • Lightweight, Rugged and Reliable
  • Small, tough and protected to 60V
  • NIST Traceable Calibration Certificate included Free



  Description

The Practical Instrument Electronics Model 512 RTD Calibrator provides direct temperature calibration to all types of instruments such as transmitters, recorders, controllers, alarms, data acquisition, and computer systems. Also, the Model 512 reads RTD outputs and displays in temperature. It is compatible with pulsed systems and transmitters (like the Rosemount 3144.) 2, 3, or 4 wire connections are detected automatically. The Model 512 is a superior replacement for decade boxes, eliminating the need for lugging around large equipment and the possibility of misreading RTD tables.

Select from 7 RTD types to source/read in °C or °F with 0.1 ° resolution. Or, select Ω for direct resistance source/read capability.
Use the EZ-Check Switch to quickly switch between three stored temperature/ Ω outputs. It’s easy to customize these values to your application. In read mode, the EZ-Check Switch recalls minimum and maximum readings. Store/Clear memory with a press of the EZ-Dial Knob.

The Practical Instrument Electronics Model 512 offers the highest performance and functions in its class by exceeding the accuracy and functions of many higher priced RTD calibrators. It is a low cost solution for checkout and calibration of all RTD instruments in the field, shop or control room. Contact Practical Instruments Electronics for custom RTD curves, ranges, or special requirements not provided by the Model 512.




  Models, Options, & Accessories


Model Description List Price Sale Price
512 RTD Read & Source (7 Types, ?/0.1) $925.00
$878.75 
Model Accessories List Price Sale Price
020-0205 Carrying Case w/PIE LOGO for all RTD & T/C models $40.00
$38.00 



  Specifications

General Specifications: (Unless otherwise indicated all specifications are rated from a nominal 23 °C, 70 % RH for 1 year from calibration)
Temperature Range -25 to 60 °C (-10 to 140 °F)
Relative Humidity Range 10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing
Size 4.9 X 3.15 X 1.82 inches (125.5 X 80 X 46.2 mm)
Weight 9.1 oz (258 grams)
Battery 9V Alkaline provides 45 hours of continuous use
Miscellaneous High contrast graphic liquid crystal display with 0.357” (9.07 mm) high digits
Low battery indication with nominal 1 hour of operation left
Protection to 60V for up to 30 seconds in duration
Resolution 0.1°C or 0.1°F / 0.01 Ω
Accuracy ±(0.015 % of Ω + 0.05) Ω (see accuracy tables for temperature error)
Span 0.00-400.00 Ω
Temperature Coefficient ±0.01 % of span in Ω/°C ambient
RTD Simulation Specifications
Allowable Excitation Current 100 µA to 10.2 mA, steady or pulsed/intermittent/smart
for accuracies below 100µA add ±10µV/Excitation Current (units are in &Omega)
Pulsed Excitation Current Compatibility DC to 0.01 second pulse widths
RTD Read Specifications
Excitation Current 1 mA nominal


   
  Welcome | Companies | Search by Type | How To Order | Leasing | About Us  
  Tequipment.NET
205 Westwood Ave
Long Branch, NJ 07740
For our Electronic Whiteboard division visit www.touchboards.com
BBBOnline Reliability Seal. Tequipment.NET
Phone: 1-877-742-TEST (8378) / 732-222-7077
Fax: (732) 222-7088
E-mail: salesteam@Tequipment.NET