Minolta LCS-II Spectrophotometer

Click image to enlarge
|
|
Features:
- Measures all Color Shades of Optically Clear Liquids
- Color Scales include Gardner, Iodine, Hazen (APHA), Lovibond,
Pharmacopoeia Also, CIELab, XYZ and Yxy under Illuminant C and
2 degree standard observer.
- Easy and intuitive.
- Stand alone unit with built-in display so there is no need
for a PC
- Can be used with either glass or disposable plastic cells/cuvettes
- The instrument automatically detects a cell or cuvette and
automatically measures
- Tungsten Lamp has expected lifetime of 2000 hours; easily
replaced by user.
- Open sample compartment for easy placement and viewing
- Integrated test media control for self checks according to
ISO 9000 and GLI
- Equipped with LCS addista ® color standards, power cable,
interface cable, and Spectral QC Software
- Spectral QC Software
- Handles large data quantities over extended periods of
time
- Allows for individual sample ID and storage
- Provides both statistical and graphical analyses
|
Description
The LCS-II precisely measures the color
of all optically clear liquids and is ideal for beverages, food,
petroleum, pharmaceuticals, cosmetics, chemicals. In addition to
standard color scales, the LCS-II includes Gardner, APHA or Hazen,
ASTM D1500, Lovibond, and Pharmacopoeia. That means you can use
it ithe LCS-II in applications as varied as checking water contaminants
at industrial sites, or measuring the color of injectable drugs
and medications. The LCS-II comes equipped with software to give
statistical evaluations as well as spectral graphs, and the instrument
can be operated as a stand-alone – no PC is required.
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| LCS-II |
Please call or email us for pricing and availability. |
Specifications
| Geometry |
0°/&de180 rectillnear |
| Spectral Range(Colorimetric |
380 to 720 nm 10 nm resolution |
| Repeatability |
0.1ΔE, 1 std. dev |
| Spectral Range(Photometric) |
340 to 900nm 1nm resolution 10 nm resolution |
| Inter-Instrument agreement |
*+ 0.2% transmission |
| Color Scale |
CIELab: CIELCH; XYZ; Yxy; Lab(h) |
| Color Differences |
ΔE* and component differences Text descriptor, CMC tolerances |
| Indices |
Gardner, Hazen, Iodine, Lovibond, Hess-Ives, European and US Phamacopoeia, Klett number, Sybolt, ASTM D 1500 |
| Spectral |
% transmission % absorbance Absorption |
| Illumin./Observer |
C/2° |
| Illumination |
Tungsten Halogen Lamp |
| Measuring compartment |
10mm x 10 mm 50mm x 10mm 11mm round |
| Operating Temperature |
50°F to 108°F(10°C to 42°C) |
| Humidity |
5 to 85% non condensing |
| Dimensions |
16.4 x 6.5 x 14.6 in (415 x 165 x 370mm) |
| Weight |
19.8 lbs. (9kg) |
| Interface |
RS-232/printer |
* Referred to distilled water
|