| Measuring geometry |
Reflectance: d/8 (diffuse illumination/8º viewing angle); SCI (specular component included)/ SCE (specular component excluded) switchable; meets ISO and DIN standards for d/8 geometry; also meets CIE and ASTM standards for d/0 geometry. Transmittance: d/0 (diffuse illumination/0º viewing angle) |
| Detector |
Silicon photodiode array with flat holographic grating |
| Wavelength range |
360 to 740nm |
| Wavelength pitch |
10nm |
| Half bandwidth |
Approx. 14nm average |
| Photometric range |
0 to 200%; Resolution: 0.001% |
| Light source |
Pulsed xenon arc lamp |
| Measurement time |
0.6 to 0.8 sec. (to start of data output) |
| Illumination/measurement areas |
Reflectance: Changeable between LAV, MAV, and SAV LAV: ø28mm illumination/ø25.4mm measurement MAV: ø11mm illumination/ø8mm measurement SAV: 5x7mm illumination/3x5mm measurement Transmittance: Approx. ø20mm |
| Repeatability |
When white calibration plate is measured 30 times at 10-sec. intervals after white calibration has been performed: Spectral reflectance: Standard deviation within 0.05% Chromaticity: Standard deviation within DE*ab 0.005 When black tile (BCRA Series ||; reflectance: 1%) is measured 30 times at 10-second intervals after white calibration has been performed: Spectral reflectance: 380 to 740nm: Standard deviation within 0.02% 360 and 370nm: Standard deviation within 0.04% Chromaticity: Standard deviation within ΔE*ab 0.05 |
| Inter-instrument agreement (LAV) |
mean ΔE*ab 0.08 (typical) based on
12 BCRA Series || color tiles. Max ΔE*ab 0.3 (corresponds to approx.
ΔE* cmc 0.2) for any of 12 BCRA Series II color tiles compared
to values measured with master body. |
| Temperature drift |
Spectral reflectance: Within ±0.10%/ºC Color difference: Within ±E*ab 0.05/ºC |
| UV adjustment |
Computer controlled; continuously variable |
| Specimen conditions for transmittance measurements |
Sheet, plate, or liquid form up to a maximum thickness of approximately 50mm |
| Interface |
RS-232C standard; baud rates; 1200, 2400, 4800, or 9600 |
| Power |
AC 100V/120V/230V/ 50/60Hz (using included AC adapter) |
| Operation temperature/humidity range |
13 to 33ºC (53 to 84ºF); less than 80% relative humidity with no condensation |
| Storage temperature/humidity range |
0 to 40ºC (32 to 96ºF); less than 80% relative humidity with no condensation |
| Dimensions |
271 x 259 x 500mm (10-11/16 x 10-3/16 x 19-11/16in.) |
| Weight |
18kg (39.7lb.) |
| Standard accessories |
White Calibration Plate CM-A90; Target Mask (3x5mm) CM-A91; Target Mask (ø8mm) CM-A92; Target Mask (ø25.4mm) CM-A93; Zero Calibration Box CM-A94; AC Adapter AC-A12; RS-232C Cable (2m/6.6 ft., 9-pin, for IBM PC) CM-A52 |
| Optional accessories |
Transmittance Specimen Holder CM-A96; Cell (2mm) CM-A97; Cell (10mm) CM-A98; Cell (20mm) CM-A99; Transmittance Zero Calibration Plate CM-A100; RS-232C Cable (for IBM PC)CM-A53 to CM-A55; RS232C Cable (5m/16.4ft., for NEC PC)CM-A56 |