Minolta CA-210 LCD Color Analyzer

Click image to enlarge
|
|
Features:
- High speed measurement of luminance and chromaticity (maximum 20 times/second)
- Measures luminance as low as 0.1cd/m2
- Ideal for measuring gamma characteristic and adjustment
- Up to 5 probes for multi-point measurements
- Accuracy of ±0.002 for white and ±0.004 for R, G,B
- CIE 1931 Standard Observer XYZ filter
- Short measuring distance of 30 ±10mm
- Special optical design limits acceptance angle within ±2.5 degrees, eliminating influence of viewing angle dependency
- 4-digit LCD display for chromaticity (more precise data readings)
- Small, lightweight (4.1kg)
- USB interface for easy connection to computer
- Matrix calibration: white and R,G,B
- LED for easy positioning of the sensor
Applications:
- Flicker Measurements, Gamma Adjustments on production line
- White Balance, chromaticity adjustment, inspection on Flat Panel Display(FPD) production lines
- Quality control and shipping inspection by FPD manufacturers
- Receiving inspection of FPDs by manufacturers and resellers
|
Description
|
The CA-210 LCD Color Analyzer is Minolta's "next generation" color analyzer designed to measure color luminance, gamma characteristic and flicker on various types of Flat Panel Displays including LCD, EL, and Plasma. Faster than previous generation analyzers, the CA-210's measurement of luminance and chromaticity at up to 20 times/second makes it ideal in high production environments. (AND it can simultaneously measure up to five points on the screen. ) The CA-210's special optical design limits acceptance angle to within ± 2.5 degrees, thus eliminating the influence of the LCD's viewing angle dependency. Matrix calibration allows for high measurement of LCD displays that provide colors through additive mixing of three monochrome colors (R,G,B). A new low luminance capability enables measurements as low as 0.1cd/m2
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| CA-210 |
CA-210 Display Color Analyzer Body, Part No: 1891-212 |
| Call |  |
|
| Model |
Accessories |
List Price |
Sale Price |
| CA-H10 |
CA-H10 HOOD, Part No: 1891-710 |
| Call |  |
|
| CA-B14 |
CA-B14 MULTI-PROBE EXPANSION BOARD, Part No: 1891-711 |
| Call |  |
|
| CA-H11 |
CA-H11 LENS CAP, Part No: 1891-712 |
| Call |  |
|
| CA-P12 |
CA-P12 Probe (2M) (LCD Only), Part No: 1891-884 |
| Call |  |
|
| CA-PU12 |
CA-PU12 Universal Measuring Probe (2M) (Plasma, CRT and LCD), Part No: 1891-835 |
| Call |  |
|
Specifications
| Item |
CA-210 |
| Receptor |
Detector: Silicon photo cell |
| Measurement area |
27mm diameter |
| Angle of aperture |
±2.5 degrees |
| Pointing function |
By LED |
| Measurement distance |
30 ±10mm |
| Display range |
Luminance: 0.01 to 999.9 cd/m² Chromaticity: Displayed in 4 or 3-digit value(user selectable) |
| Luminance |
Measurement range: 0.10 to 999.9 cd/m² Accuracy: ±2% ±1 digit Repeatability: 0.10 to 0.99 cd/m² 0.2%+1 digit (2σ) 1.00 to 999.9 cd/m² 0.1%+1 digit (2σ) |
| Chromaticity |
Measurement range: 0.10 to 999.9 cd/m² Accuracy: 0.10 to 4.99 cd/m² ±0.005(for white) 5.00 to 19.99 cd/m² ±0.004(for white) 20.00 to 999.9 cd/m² ±0.003(for white) 160cd/m² ±0.002(for white) ±0.004 (for monochrome) Repeatability: 0.10 to 0.19 cd/m² 0.010 (2σ) 0.20 to 0.49 cd/m² 0.005 (2σ) 0.50 to 0.99 cd/m² 0.002 (2σ) 1.00 to 999.9 cd/m² 0.001 (2σ) |
| Flicker Contrast Method |
Measurement range: 5 cd/m² or higher Display range: 0.0 to 100% Accuracy: ±1% (30 Hz AC/DC 10% sine wave) ±2% (60 Hz AC/DC 10% sine wave) Repeatability: 1% (2σ) (AC/DC 10% sine wave) |
| Flicker JEITA method |
Measurement range: 5 cd/m² or higher Accuracy: &plumsn;0.5 (30 Hz AC/DC 10% sine wave) Repeatability: 0.3 (2σ) (30 Hz AC/DC 10% sine wave) |
| Measurement speed |
xyLv: 0.10 to 1.99cd/m² 5 measurements/sec (4.5 measurements/sec) 2 cd/m² or higher 20 measurements/sec (17 measurements/sec) Flicker Contrast: 16 measurements/sec. (16 measurements/sec.) Flicker JEITA: 0.5 measurements/sec. (0.3 measurements/sec.) |
| Display |
Digital: xyLv, XYZ, TΔuvLv, u'v'Lv RGB analyze Flicker (Contrast method) Analog: ΔxΔyΔLv,;R/G B/G ΔG, ΔR B/R G/R, Flicker (Contrast method) |
| SYNC mode |
NTSC, PAL, EXT, UNIV, INT |
| Object under measurement |
Vertical synchronizing frequency: 40 to 200 Hz (Flicker: 40 to 130 Hz) |
| Memory channel |
100 channels |
| Analyzer function |
Standard function |
| Interface |
RS-232C (38,400 bps or below), USB(1.1 conformity) |
| Multi-point expansion |
Max. 5 points with 4-probe expansion board (CA-B14) |
| Software |
SDK software (supplied as standard accessory) |
| Operating temperature/humidity range |
0 to 40°C, 32 to 104°F 85%RH or less, no condensation at 35°C /95°F) |
| Storage temperature range |
-20 to 55°C |
| Input voltage range |
100 to 240, 50/60 Hz, 50VA |
| Size |
Main body: 340 (W) x 127 (H) x 216 (D) Probe: 49 x 204 |
| Weight |
Main body: 3.58 kg Probe: 520 g |
| Operating environment |
Equipment category: 2 Pollution degree: 2 |
| Standard accessories |
AC power cord SDK Soft CA-SDK Hood CA-H10 |
| Optional accessories |
4-point expansion board CA-B14, Probe CA-P12(2m)CA-P15(5m), Hood CA-H10, Lens Cap CA-H11 |
|