Martel Electronics Beta TC-100 Precision Thermocouple Calibrator

Click image to enlarge
Martel TC100
|
|
Features:
- High accuracy ±0.3 °C (Type J T/C - all errors combined)
- Ten(10) common T/C Types plus mV
- Accepts both T/C mini-plug and bare T/C wires
- Simple decade control of output
- Three setpoints for each T/C type
- MIN/MAX recall in measure mode
- Input protection to 240VAC
- Supplied in full rubber boot; optional carrying case available
|
Description
|
The Thermocouple Calibrator provides high accuracy and measurement of ten common thermocouples, as well as mV. The TC-100 thermocouple calibrator's accuracy of ±0.3 °C for Type J T/Cs includes all errors, at resolutions of 0.01 °C or °F in measure mode, and 0.1 °C or °F in source mode. Features including MIN/MAX recall in measure mode, three setpoints per thermocouple range, a large knob for decade control of the output in source mode, and the ability to accept bare T/C wires in addition to mini-plug inputs, makes the thermocouple calibrator an accurate, easy-to-use instrument for all your thermocouple calibration needs.
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| TC100 |
TC100 Precision Thermocouple Calibrator, includes Protective Sleeve Case, Batteries, Instruction Manual, NIST Cal Cert
|
$740.00 |
| $695.60 |  |
|
| PTC8001 |
PTC8001 Temperature calibrator for T/Cs and RTDs, includes protective rubber boot, test leads, batteries, instruction manual and NIST Cal Cert |
$1045.00 |
| $982.30 |  |
|
Specifications
| TC-100 Specifications (1 year at 23 °C ±5 °C; % of reading unless otherwise noted) |
| Output Voltage |
| Range |
-10 to +75.000 mV |
| Resolution |
1 µV |
| Accuracy |
±0.007 % of rdg, ±10 µV |
| Output Impedance |
≤1 Ohm |
| Input Voltage |
| Range |
-10 to +75.000 mV |
| Resolution |
1 µV |
| Accuracy |
±0.007 % of rdg, ±10 µV |
| Input Impedance |
> 1 MegOhm |
| Thermocouple Source/Measure |
| Types |
J, K, T, E, R, S, B, L, U, C |
| Range |
mV |
| Resolution |
| Source |
±0.1°C or °F |
| Measure |
±0.01°C or °F |
| Accuracy |
| J |
±0.5 °C; -210 °C to -100 °C ±0.3 °C; -100 °C to +1,200 °C |
| K |
±0.6 °C; -200 °C to -100 °C ±0.35 °C; -100 °C to +1,000 °C ±0.5 °C; +1,000 °C to +1,372 °C T ±0.7 °C; -200 °C to -150 °C ±0.3 °C; -150 °C to +400 °C |
| E |
±0.5 °C; -200 °C to -100 °C ±0.3 °C; -100 °C to +1,000 °C |
| R |
±1.8 °C; 0 °C to 250 °C ±1.0 °C; 250 °C to +1,767 °C |
| S |
±1.8 °C; 0 °C to 250 °C ±1.0 °C; 250 °C to +1,767 °C |
| B |
±1.7 °C; 600 °C to 1,000 °C ±1.2 °C; 1,000 °C to 1,820 °C |
| L |
±0.5 °C; -200 °C to -100 °C ±0.4 °C; -100 °C to +900 °C |
| U |
±0.7 °C; -200 °C to 0 °C ±0.3 °C; 0 °C to +600 °C |
| C |
±0.4 °C; 0 to °C 1,000 °C ±0.7 °C; 1,000 °C to +1,800 °C ±1.2 °C; +1,800 °C to +2,316 °C |
| CJC Temp. Offset |
±0.05 °C/°C outside of 23±5 °C |
| Warm-up Time |
1 minute to specification |
| Environmental |
| Operating Temperature |
-10 °C to +55 °C |
| Storage Temperature |
-20 °C to 70 °C |
| Power Requirements |
9 VDC |
| BatterY |
9 V alkaline; 006P/ IEC 6F22/NEDA1604 Optional NiCad Optional AC adapter/charger |
| Mechanical |
| Dimensions |
5.7” H x 3.15” W x 1.43” D(144.7 x 80.0 x 36.3 mm) |
|