Martel Electronics M2000/M2001 Lab standard Multi-Function calibrators

Click image to enlarge
Martel M2000A
|
|
Features:
- Superior Calibration accuracy to 0.0025% of reading
- Direct keyboard entry or cursor entry with decade control
- Automatic standby function protects device under test
- Source current, voltage, T/C, RTD (Ohms)
- Read T/C, RTD and pressure
- Source/read eleven (11) thermocouple types
- Source/read nine (9) standard RTD types
- Custom RTD/SPRT profile
- Nine (9) setpoints for each output range and type
- Local or RS232 remote control
- Optional IEEE-488 control available
- Compatible with Fluke Met/Cal® software
- Optional rack/panel mount kit available
Additional Photos
Click image to enlarge
|
Description
|
The Lab Calibrator combines the power and features of a voltage/current calibrator, with the addition of thermocouple, RTD, and pressure calibration, for a single laboratory calibration instrument unmatched in versatility, performance, and value. As with every Martel lab calibrator, the calibration equipment's world-class performance and features are accessed through a very simple-to-use, intuitive user interface. The calibration test equipment is truly a "process calibration laboratory in a box."
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| M2000A |
High Accuacy V/I Lab Standard with IEEE-488 Communication 120= 1919092, |
$1995.00 |
| $1875.30 |  |
|
| M2000A |
High Accuacy V/I Lab Standard with IEEE-488 Communication 120= 1919092, |
$1995.00 |
| $1875.30 |  |
|
| M2000A-120 |
M2000A [120] High Accuacy V/I Lab Standard with IEEE-488 Communication Interface |
$1995.00 |
| $1875.30 |  |
|
| M2000A-240 |
M2000A [240] High Accuacy V/I Lab Standard with IEEE-488 Communication Interface |
$1995.00 |
| $1875.30 |  |
|
| M2001 |
High accuracy process lab standard 120= 1919038, 240= 1919139 |
$3295.00 |
| $3097.30 |  |
|
| M2001A |
M2001A High Accuracy Process Lab Standard with IEEE-488 Communication Interface |
$3595.00 |
| $3379.30 |  |
|
| M2001A-120 |
M2001A [120] High Accuracy Process Lab Standard with IEEE-488 Communication Interface |
$3595.00 |
| $3379.30 |  |
|
| M2001A-240 |
M2001A [240] High Accuracy Process Lab Standard with IEEE-488 Communication Interface |
$3595.00 |
| $3379.30 |  |
|
Specifications
| Specifications for M2000(1 year at 23°C ±5°C unless otherwise noted)) |
| Output Voltage |
| Range |
0 to 100.000 mV 0 to 1.00000 V 0 to 10.0000 V 0 to 100.000 V |
| Resolution |
| 0 to 100 mV Range |
1 µV 0 to 1 V Range :10 µV 0 to 10 V Range: 100 µV 0 to 100 V Range :1 mV |
| Accuracy (% of reading) |
| 0 to 100 mV Range |
±0.003% (30ppm) ± 3.0 µV 0 to 1 V Range :±0.003% (30ppm) ± 20.0 µV 0 to 10 V Range :±0.003% (30ppm) ± 200.0 µV |
| 0 to 100 V Range |
±0.003% (30ppm) ± 2.0 mV |
| Maximum Burden (≤ 1 Ohm output impedance) |
| 0 to 100 mV Range |
10 mA |
| 0 to 1 V Range |
10 mA |
| 0 to 10 V Range |
10 mA |
| 0 to 100 V Range |
1 mA |
| Output Current |
| Range |
0 to 100.000 mA |
| Resolution |
1 µA |
| Accuracy (% of reading) |
±0.01% ± 2 µA |
| Maximum Burden |
10 V |
| Stability |
| Warm-up Time |
30 minutes to rated accuracy |
| Temp Co. (≤18°C/>28°C) |
10% of accuracy spec/°C |
| Temperature Range |
| Operating |
0°C to +50°C |
| Storage |
-20°C to +70°C |
| Power Requirements |
| Voltage range |
90 to 240 VAC <15 VA |
| Mechanical |
| Dimensions |
11.5” x 4.7” x 8.75”(29.21 cm x 11.93 cm x 22.00cm) |
| Weight |
Approximately 5 pounds (2.27 kg) |
| Display |
16 x 2 large character alphanumeric, backlit LCD |
| Specifications for M2001(1 year at 23°C ±5°C; % of reading, unless otherwise noted) |
| Output Voltage |
| Range |
0 to 100.000 mV 0 to 1.00000 V 0 to 10.0000 V 0 to 100.000 V |
| Resolution |
| 0 to 100 mV |
Range 1 µV |
| 0 to 1 V Range |
10 µV |
| 0 to 10 V Range |
100 µV |
| 0 to 100 V Range |
1 mV |
| Accuracy (% of reading) |
| 0 to 100 mV Range |
±0.003% (30ppm) ± 3 µV |
| 0 to 1 V Range |
±0.003% (30ppm) ± 20 µV |
| 0 to 10 V Range |
±0.003% (30ppm) ± 200 µV |
| 0 to 100 V Range |
±0.003% (30ppm) ± 2 mV |
| Maximum Burden (= 1 Ohm output impedance) |
| 0 to 100 mV Range |
10 mA |
| 0 to 1 V Range |
10 mA |
| 0 to 10 V Range |
10 mA |
| 0 to 100 V Range |
1 mA |
| Output Current |
| Range |
0 to 100.000 mA |
| Resolution |
1 µA |
| Accuracy (% of reading) |
±0.01% ± 2 Counts |
| Maximum Burden |
10 V |
| Thermocouples |
| Output |
| Types |
J, K, T, E, R, S, N, B, L, U, C |
| Range |
mV |
| Resolution |
0.1 °C/°F |
| Accuracy |
0.14 °C; Type J, typical |
| Input |
| Types |
J, K, T, E, R, S, N, B, L, U, C |
| Range |
mV |
| Resolution |
0.01 °C/°F |
| Accuracy |
0.14 °C; Type J, typical |
| RTD |
| Output |
| Range |
PT385 (100, 200, 500, 1000),PT392, PT3916 (JIS), Ni120, Cu10, YSI400 |
| Resolution |
0.1 °C/°F; PT385-100, typical |
| Accuracy |
±0.1 °C; PT385-100, typical |
| Input (All ?/RTD inputs are 4 wire) |
| Range |
PT385 (100, 200, 500, 1000),PT392,PT3916 (JIS), Ni120,Cu10,YSI400, 25 Ohm SPRT |
| Resolution |
0.001 °C/°F; PT385-100, typical |
| Accuracy |
±0.020 °C; PT385-100, typical |
| Ohms |
| Output |
| Range |
5 to 4000.0 Ohms |
| Resolution |
| 5 to 400.00 |
0.01 |
| 5 to 4000.0 |
0.1 |
| Accuracy |
| 5 to 400.00 |
±0.03 |
| 5 to 4000.0? ±0.3? |
| Input |
| Range |
0 to 4000.00 Ohms |
| Resolution |
| 0 to 400.00 |
0.001 |
| 0 to 4000.0 |
0.01 |
| Accuracy |
| 0 to 400.00 |
40 PPM ±0.002? |
| 0 to 4000.0 |
40 PPM ±0.02? |
| Pressure |
| Range |
0 to 1 inch H2O; to 10,000 psi |
| Compatibility |
All Fluke 700 Series Pressure Modules |
| Stability |
| Warm-up Time |
30 minutes to rated accuracy |
| Temp Co. (=18°C/>28°C) |
10% of accuracy spec/°C |
| Environmental |
| Operating Temperature |
0°C to +50°C |
| Storage Temperature |
-20°C to +70°C |
| Humidity |
| Operating |
<80% to 30°C <70% to 40°C <40% to 50°C |
| Storage |
<95%, non-condensing |
| Power Requirements |
| VoltageRange |
90 to 240 VAC <15 VA |
| Mechanical |
| Dimensions |
11.5" x 4.7" x 8.75"(29.21 cm x 11.93 cm x22.00cm) |
| Weight |
Approximately 5 pounds(2.27 kg) |
| Display |
16 x 2 large character alphanumeric, backlit LCD |
|