Martel Electronics Beta DMC-1400 Documenting Multifunction Calibrator

Click image to enlarge
Martel DMC-1400
|
|
Features:
- Maintain calibration data for 50 instruments in non-volatile
memory
- Print calibration certificates without PC
- Communications port compatible with BetaPort-P pressure modules
included
- All source modes can be programmed with dedicated setpoints
- Built-in 250 Ohm resistor for HART compatibility with smart
transmitters and PLCs
- Direct entry of custom RTD coefficients(R0,A,B,C)
- Meets CE Requirements and is designed to IEC 1010 safety standards
|
Description
|
The DMC-1400 Calibrator fully documents any calibration easily while performing the normal job. No special software is required! It features the ability to test for pass/fail criteria, measure and simulate thermocouples, RTDs, current, voltage, and frequency, and source pulse trains.
In addition, the built-in 250 Ohm resistor for HART compatibility with smart transmitters and PLCs, full fuseless protection, and a serial communications port for full control with ASCII commands
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| DMC-1400 |
DMC-1400 Documenting Multi-function Calibrator with NIST calibration certificate. |
$2695.00 |
| $2533.30 |  |
|
| Model |
Accessories |
List Price |
Sale Price |
| 80340 |
AC Adaptor for DMC-1400 |
| $42.00 |  |
|
Specifications
| DMC-1400 Specifications (23 °C ±5 °C unless otherwise noted) |
| Pressure Ranges |
Various ranges in 29 different pressure modules – see separate Pressure Module data sheet |
| T/C Ranges |
J, K, T, E, R, S, N, B, L, U,including – 10 to 75 mV |
| RTD Ranges |
Pt 385 (100, 200, 500, 1000 Ohms) Pt 392, JIS, Ni120,Cu10, YSI400 |
| Ohms Ranges |
0 to 400.00 and 400 to 3,200 |
| mA Range |
0 to 24.000 mA |
| Voltage |
| (Low) Read |
0-20.000 VDC |
| Source |
0-20.000 VDC |
| (High) Read only |
0-250 VDC or VAC |
| Frequency |
0.000 to 24.000 mA |
| CPM |
| Source |
2.0 - 1,000.0 |
| Measure |
0.0 - 1,000.0 |
| Hz |
| Source |
0 - 1,000 |
| Measure |
0.0 - 1,000.0 |
| kHz |
| Source |
1.0 -10.0 |
| Measure |
0.00 -10.00 |
| Continuity |
0 to 400.0 Ohms;audible tone when under 10 Ohms |
| Accuracy |
[±x.x °C total error] |
| T/C Type J |
±0.3 °C ±10 µV; [±0.4 °C] |
| RTD Pt 100 |
±0.1 °C ±0.075 ; [±0.3 °C] |
| Ohms (400) |
±0.1 |
| Ohms (3,200) |
±1.0 |
| mA Compliance |
Up to 1,000 load |
| Voltage Drive Capability |
1 mA |
| RTD IEX-Range |
0.01 mA to 3 mA |
| RTD Frequency Response |
10 ms – works with all pulsed transmitters Environmental |
| Operating Temperature |
10 °C to +50 °C |
| Storage Temperature |
20 °C to 70 °C |
| Power Requirements |
6 VDC Batteries 4 AA alkaline batteries, 25 hours typical operating time |
| Mechanical |
| Dimensions |
8.7” H x 4.2” W x 2.3” D (220.9 x 106.6 x 58.4 mm) |
| Weight |
30.5 ounces (863 gms) |
| Accessories Included |
Test leads, 4 AA alkaline batteries, NIST Certifi cate, and instruction manual |
| Optional Accessories |
Carrying Case; Model CC572 |
|