Features:
  • Superior calibration accuracy to 0.0025% of reading
  • Direct keyboard entry or cursor entry with decade control
  • Source/Read thermocouple (13), RTD (9), Voltage, Current, Pressure (read only)
  • Custom RTD and SPRT profiles
  • Nine (9) setpoints for each output range and type
  • Beryllium-Copper binding posts reduce thermal EMFs
  • RS232, USB and IEEE-488 remote control
  • Compatible with Fluke Met/Cal® software
  • Isolated measurement channel
  • Two (2) voltage ranges: 10V and 100 V DC
  • MilliAmp range 0 to 52 mA
  • MilliAmp range with simultaneous 24 VDC power
  • Selectable 250 Ohm HART™ resistor
  • Accuracy of 0.005% of reading on voltage ranges



  Description

The Martel 3001 precision calibrator combines the power and features of the M2001 (voltage, current, TC, RTD and pressure) with a second completely isolated measurement channel for a single laboratory calibration instrument unmatched in versatility, performance, and value. As with every Martel calibrator, the 3001’s world-class performance and features are accessed through a very simple-to-use, intuitive user interface. The Martel 3001 is truly a “process calibration laboratory in a box.”

Simple, Intuitive Interface
The 3001 provides simple, front-panel entry of mode, range, and value, using either direct key-board entry (1) or cursor entry (2). Using cursor entry , the LEFT/RIGHT arrow keys are used to move the cursor under the digit in the display to be changed. The UP/DOWN arrow keys increment/decrement the value at the cursor position. Using direct keyboard entry (1), the exact value desired is entered using the numeric keys, and the ENTER key is pressed to set the output to that value. Whichever way you choose, setup is simple and fast. In the voltage output mode, the 3001 auto-ranges on the entered value for maximum accuracy at all times.

The Performance You Demand – In Any Mode
Voltage Mode
The 3001 offers four precision voltage output ranges (100mV, 1V, 10V, and 100V) all with 0.003% (30ppm) accuracy. These ranges are ideal for calibrating a broad range of DC voltage instrumentation. Additionally all voltage outputs settle to full specification in less than 200ms making the 3001 ideal for automated calibration systems.
An automatic stand-by mode (3) assures that output voltages above 30VDC must be acknowledged by the operator before the voltage appears at the output jacks. The stand-by mode is also triggered if the output current compliance is exceeded, thereby protecting the device under calibration.

Current Mode
The 3001 features a precision current output range (100mA) that offers 0.01% (100ppm) accuracy, which is ideal for calibrating process instrumentation especially 4 to 20mA equipment. With a full 12 volts of compliance at 100mA virtually any precision DC current measuring device can be calibrated using the 3001. Like the voltage ranges the current range offers quick settling time and an operate/stand-by mode.

Thermocouple Mode
The Martel 3001 can read and source any of 11 types of thermocouples. Its T/C input and output is Cold Junction Compensated, using an ultra-stable PT-1000 sensor.

RTD Mode
The 3001 can read and source 9 RTD types as well as YSI-400 and Ohms for non-standard curves. Probe coefficients (A, B, C, and R0) can be entered directly, with storage for up to five custom curves and one SPRT curve. The performance of the 3001 in the RTD mode compares to dedicated RTD measurement instruments. Unlike low-cost, less accurate RTD instruments, the display in the 3001 is always active, reading to three decimal places, using polynomial averaging to extract a high accuracy signal. The result is a very quiet, high accuracy reading.

Pressure Mode
The 3001 operates with all Martel BetaPort series pressure modules using the BPPA module adapter and covers pressure ranges from 0 to 10" H2O to 10,000 psi. Pressure can be displayed in a wide range of engineering units with up to 0.025% Full Scale accuracy. The 3001 also supports Fluke 700 series modules and Mensor 6100 precision pressure modules.

Total Setpoint Control
A SHIFT key (4) provides easy access to the setpoint controls of the 3001. Up to nine setpoints can be defined for each output mode and each thermocouple and RTD type. Setpoints are recalled individually at the touch of three buttons, SHIFT (4), SETPOINT (SPT) button and then the corresponding numeric keys 1-9. Any number of sequential setpoints can be stepped through automatically, with complete control of dwell time. Either way, for rapid setup of repeatable tests, no other instrument comes close to the Martel 3001.

Remote Control
All of the 3001 operating functions can be accessed via RS-232, IEEE-488 or USB using a standard PC running Fluke Met/Cal® software, Windows® HyperTerminal or other software using an ASCII protocol. Custom control programs may be written using programming software such as C++. Switching between LOCAL and REMOTE is as simple as touching the SHIFT (4) and LOCAL buttons.

Rock-Solid Stability
The 3001 stability and accuracy is traceable to NIST standards. The accuracy of the 3001 is specified for both 90-day and one-year intervals. Manual zero calibrations can be made on all T/C and pressure functions to eliminate offsets.

Flexible Output
Five-way copper alloy binding posts (5) provide a wide range of connection options. A standard pressure module connector is provided (6), as is the CJC T/C mini-jack (7).

Isolated Measurement Channel
The 3001 features a fully isolated measurement channel which allows the user to calibrate process transmitters and signal isolators. In reality it’s like having two instruments in one! This channel also incorporates a 24 volt loop power supply to power 2-wire transmitters and a HART interface resistor enabling direct connection to HART communicators. Key features are:

  • Two voltage ranges 10V and 100V DC
  • Milliamp range 0 to 52mA
  • Milliamp range with simultaneous 24 volt power (0 to 24ma)
  • Selectable 250 ohm HART resistor
  • Accuracy of 0.005% of reading on all ranges




  Models, Options, & Accessories


Model Description List Price Sale Price
3001 3001 BENCH STANDARD CALIBRATOR with Ultra-High Accuracy, Isolated Readback Circuit, IEEE interface (488) & 24 VDC instrument power supply (Available 1 June 2007) $4295.00
$4037.30 
3001-120 3001 BENCH STANDARD CALIBRATOR [120] with Ultra-High Accuracy, Isolated Readback Circuit, IEEE interface (488) & 24 VDC instrument power supply (Available 1 June 2007) $4295.00
$4037.30 
3001-240 3001 BENCH STANDARD CALIBRATOR [240] with Ultra-High Accuracy, Isolated Readback Circuit, IEEE interface (488) & 24 VDC instrument power supply (Available 1 June 2007) $4295.00
$4037.30 
Model Accessories List Price Sale Price
BPPA100 BPPA100 BetaPort-P Pressure Module Adapter $250.00
$235.00 
IBP-1 IBP-1 High Accuracy PT100 RTD Probe, for M2001, 3001 $695.00
$653.30 
LCA-1 LCA-1 Spare RTD Lemo Connector for M2001, 3001 $49.00
$46.06 
PTL-1 PTL-1 Low EMF Beryllium Copper Test Leads $40.00
$37.60 
URA-1 URA-1 Universal RTD Adapter for M2001, 3001, BetaGauge 311/321 $129.00
$121.26 
80029 Thermocouple Connection Kit for Types J, K, T & E $89.00
$83.66 
80036 Thermocouple Connection Kit for Types R, S, B & N $110.00
$103.40 



  Specifications

Output Voltage
Range 0 to 100.000 mV
0 to 1.00000 V
0 to 10.0000 V
0 to 100.000 V
Resolution
0 to 100 mV Range 1 µV
0 to 1 V Range 10 µV
0 to 10 V Range 100 µV
0 to 100 V Range 1 mV
Accuracy (% of reading)
0 to 100 V Range ±0.003% (30ppm) ± 3 µV
0 to 1 V Range ±0.003% (30ppm) ± 10 µV
0 to 10 V Range ±0.003% (30ppm) ± 100 µV
0 to 100 V Range ±0.003% (30ppm) ± 1 mV
Maximum Burden (~ 1 Ohm output impedance)
0 to 100 mV Range 10 mA
0 to 1 V Range 10 mA
0 to 10 V Range 10 mA
0 to 100 V Range 1 mA
Output Current
Range 0 to 100.000 mA
Resolution 1 µA
Accuracy (% of reading) ± 0.005% ± 1 Count
Maximum Burden 10 V
Thermocouples
Output
Types J, K, T, E, R, S, N, B, L, U, C, BP, XK
Range mV
Resolution 0.1 °C/°F
Accuracy 0.14 °C; Type J, typical
Input
Types J, K, T, E, R, S, N, B, L, U, C, BP, XK
Range mV
Resolution 0.01 °C/°F
Accuracy 0.14 °C; Type J, typical
RTD
Output
Range Pt385 (100, 200, 500, 1000), Pt392, Pt3916 (JIS), Ni120, Cu 10, YS I400
Resolution 0.01 °C/°F; Pt385-1 00, typical
Accuracy ±0.05 °C; Pt385-100, typical
Input (All RTD inputs are 4 wire)
Range Pt385 (100, 200, 500, 1000), Pt392, PT3916 (JIS), Ni120, Cu10, YSI400, 25 Ohm SPRT
Resolution 0.001 °C/°F; Pt385-100, typical
Accuracy ±0.02 °C; Pt385-100, typical
Ohms
Output
Range 5 to 4000.0 O
Resolution 5 to 400.00 O 0.001 O
5 to 4000.0 O 0.01 O
Accuracy 5 to 400.00 O ±0.05 O
5 to 4000.0 O ±0.3 O
Input (4 wire connection)
Range 0 to 4000.00 O
Resolution 0 to 400.00 O 0.001 O
0 to 4000.0 O 0.01 O
Accuracy 0 to 400.00 O 40 PPM ±0.002 O
0 to 4000.00 O 40 PPM ±0.02 O
Pressure
Range 0 to 1 inch H2O; to 10,000 psi
Compatibility All BetaPort modules using the BPPA adapter and all Fluke 700 and Mensor 6100 Series Pressure Modules
Isolated Measurement Channel
Range Accuracy
0-10.0000V ±0.005% ± 0.2mV
0-100.000V ±0.005% ± 2.0mV
0-52.0000mA ±0.01% ± 1µA
Loop power 24 V ± 10%
HART™ resistor 250O ± 3%
Maximum current 24 mA
Stability
Warm-up Time 30 minutes to rated accuracy
Temp Co. (~18°C/>28°C) 10% of accuracy spec/°C
Environmental
Operating Temperature 0°C to +50°C
Storage Temperature -20°C to +70°C
Humidity
Operating <80% to 30°C
<70% to 40°C
<40% to 50°C
Storage <95%, non-condensing
Power Requirements
Voltage Range 90 to 240 VAC
<15 VA
Mechanical
Dimensions 5"h x 19"w x 11"d
(17.7 cm x 48.26 cm x 27.96 cm)
Weight 10.5 lbs. (4.8 kg)
Display (2) Large character 16 by 2 line alphanumeric backlit LCDs



   
  Welcome | Companies | Search by Type | How To Order | Leasing | About Us  
  Tequipment.NET
205 Westwood Ave
Long Branch, NJ 07740
For our Electronic Whiteboard division visit www.touchboards.com
BBBOnline Reliability Seal. Tequipment.NET
Phone: 1-877-742-TEST (8378) / 732-222-7077
Fax: (732) 222-7088
E-mail: salesteam@Tequipment.NET