| Input Section |
| HF INPUT |
(1-7 modulation signal input) |
| Input Coupling |
AC |
| Measurement Voltage Range |
0.1 to 2.0 V p-p |
| Input Range |
3 ranges |
| 0.3 V range |
0.1 to 0.3 V p-p |
| 0.9 V range |
0.3 to 0.9 V p-p |
| 2.0 V range |
0.9 to 2.0 V p-p |
| Input Impedance |
50 Ω |
| Auto Slicer |
Response time constant: 10 kHz |
| Maximum Input Voltage |
± 2.5 V |
| Measurement Control (ARMING IN/INHIBIT IN) |
| Input Impedance |
10k Ω |
| Input Signal Level |
0/+5 V |
| Maximum Input Voltage |
-0.7 V/+5.7 V |
Equalizer Section Conforms to Blu-ray Disc (Part 1, version 1.0) standards |
| Applicable format |
1-7 PP modulation |
| Channel Bit Rate |
66 MHz |
| Equalizer Mode |
Conventional Equalizer Mode |
| Gain Variable Range |
0 to 8 dB (36 steps) |
| Gain Accuracy |
±0.5 dB |
| Group Delay Deviation |
[2.0 nsp-p (5.8 dB: 3 MHz to 16.5 MHz) |
| Jitter Measurement Section |
| Applicable Speed |
Clock frequency: 66 MHz ±4 % |
| Measurement Mode |
DATA to CLOCK DATA to CLOCK 2T eliminate mode |
| Measurement Resolution |
25 ps |
| Unit Displayed |
% (Sigma value with respect to clock period) |
| Measurement Accuracy |
| Meter Indication |
± 5% of full scale ± residual jitter |
| Digital Display |
± 5% ± residual jitter |
| Polarity Section |
| CLOCK |
fixed |
| Measurement Time Constant |
0.04 s to 5 s |
| Measurement Control Section (ARMING/INHIBIT function) |
| Measurement Control Mode |
ARMING mode/INHIBIT mode |
| Electrical Characteristics |
| Input Impedance |
10k Ω |
| Input Signal Level |
0/+5 V |
| Maximum Input Voltage |
-0.7 V/+5.7 V |
| INHIBIT |
| ENABLE |
HIGH/LOW |
| COUNT |
0 to 50, in 1 steps |
| ARMING |
| SLOPE |
RISE/FALL |
| COUNT |
0 to 50, in 1 steps |
| START DLY |
0 to 99.99 ms (in 0.01 steps min.) |
| LENGTH MODE |
SAMPLE/TIME |
| LENGTH |
0.01 to 99.99 ms (in 0.01 steps min.)(TIME only) |
Clock Regenerator (DATA to CLOCK mode only) Regenerates reference clock signal from DATA signal input. |
| HF |
66 MHz ± 8% |
| Clock Frequency Measurement Section (DATA to CLOCK mode only) |
| Measurement Range |
60.72 MHz to 71.28 MHz |
| Measurement Accuracy |
± 0.1% |
Judgement Section Outputs GO/NO GO results of jitter and frequency measured with respect to the preset value. |
| GO/NO GO LED |
Indicates results corresponding to meter indications. |
| Dedicated Remote |
| Control Connector |
Outputs judgement results for jitter measurement. |
| Output Section |
| MONITOR OUT |
To Monitor the HF Signal Input |
| Output Impedance |
50 Ω |
| Output Amplitude |
Up to 2.0 V p-p ± 30% (into 50 Ω, in proportion to input signal) |
| Output Connector |
BNC |
EQUALIZER OUT To Monitor the HF signal input. |
| Output Impedance |
50 Ω |
| Output Amplitude |
0.8 V p-p ± 30% (into 50 ) |
| Output Connector |
BNC |
DIGITAL OUT (DATA, CLOCK) Outputs binary DATA signal and clock signal regenerated by PLL. |
| Output Signal |
DATA signal, CLOCK signal |
| Output Amplitude |
0.15 V p-p (into 50 ) |
| Output Offset Voltage |
0.34 V (into 50 ) |
| Output Connector |
BNC |
| DC OUT |
| Output Voltage |
0.05 V/% |
| Output Accuracy |
± 2% |
GATE MONITOR To monitor arming/inhibit control signals. |
| Output Amplitude |
TTL Level |
| Output Impedance |
1k Ω |
| Remote Control Section |
| Dedicated Remote Control Connector |
Communicates judgement results and front panel settings. |
| Input Level |
0/5 V (pulled-up with 47k Ω resistor) |
| Maximum Input Voltage |
-0.7/+5.7 V |
| Judgement Results Output Pins |
GO: 5 V (open drain output, pulled-up with |
47k Ω resistor)
NO GO |
0 V |
| Maximum Current Output |
10 mA |
| RS-232C Interface |
| Communication |
Controls function, outputs data. |
| Baud Rate |
38400 bps max. |
| General Specifications |
| Operating |
Temperature : 0 to 40° C Humidity : 85% RH (without condensation) |
| Spec-Guaranteed |
Temperature : 10 to 30° C Humidity : 85% RH (without condensation) |
| Operating Environment |
Indoor Use |
| Operating Altitude |
Up to 2,000 m |
| Overvoltage Category |
II |
| Pollution Degree |
2 |
| Power Requirements |
90 to 250 VAC, 50/60 Hz Max. 55 W, 80 VA max |
| Dimension (WxHxD) |
83/8 x 51/4 x 1113/16 in. 213 x 132 x 300 mm |
| Weight |
8.8 lbs., 4.0 kg |
| Accessories |
Power Cord: 1 |
| Instruction Manual |
1 |
Option OP70 Limit Equalizer (For Blu-ray HF) Conforms to Blu-ray Disc (part 1, version 1.0 standards) |
| Applicable Format |
1-7 PP modulation |
| Channel Bit Rate |
66 MHz |
| Equalizer Mode |
Conventional Equalizer Mode or Limit, Equalizer Mode, selectable |
| Gain Variable Range |
0 to 8 dB (36 steps) |
| Gain Accuracy |
± 0.5dB |
| Group Delay Deviation |
2.0 nsp-p (5.8 dB: 3 MHz to 16.5 MHz) |
Option OP71 DVD/CD Measurement Input Section DATA INPUT (EFM/8-16 modulation signal input) |
| Input Coupling |
AC (2 Hz/1 kHz, selectable) |
| Measure Voltage Range |
50 mV to 5 Vp-p |
| Input Impedance |
1M Ω/50 Ω, selectable |
| Slice Level VARIABLE |
± 2.5V |
| AUTO (ASYMMETRY ON) |
20 Hz/1 kHz/5 kHz/10 kHz, selectable |
| Maximum Input Voltage |
± 5V |
| Applicable Speed |
DVD: Clock Frequency x1 speed: 27 MHz ± 10% x2 speed: 54 MHz ± 10% CD: x1, x2, x4, x8, x10, x12 speed |
| Measurement Mode |
DVD: PERIOD mode, sum of all -T data in PERIOD mode, PULSE WIDTH mode, sum of all -T data in PULSE WIDTH mode,DATA to CLOCK |
| CD |
PULSE WIDTH mode, sum of all -T data
in PULSE WIDTH mode Unit Displayed : ns,% |
| Measurement Resolution |
50 ps |
| Display Resolution |
0.01 ns |
| Measurement Accuracy Sigma Value |
±4 % ±0.15 ns Average Value: ±1 ns |
| Option OP73 GPIB (IEEE 488.2) |
| Function |
Transfers data, controls front panel settings. |
| Option OP74 LAN |
| Function |
Transfers data, controls front panel settings. |