| Operation Mode |
Independent: |
Two independent outputs and 5V fixed output
Outputs from 0V to rated voltage and 0A to rated current |
| Series: |
Output from 0V to ± rated voltage at rated current
Output from 0V to double rated voltage at rated current |
| Parallel: |
Output from 0A to double rated current at rated voltage |
| 5V Fixed Output |
Line Regulation: |
£5mV |
| Load Regulation: |
£10mV |
| Ripple and Noise: |
£2mVrms |
| Voltage Accuracy: |
5V ±0.25V |
| Output Current: |
3A |
| Tracking Operation |
Tracking Error: |
£0.5% + 10mV of the master |
| Series Regulation: |
£300mV |
| Constant Voltage Operation |
Line Regulation: |
£0.01% +3mV |
| Load Regulation: |
£0.01% +3mV (rating current £3A)
£0.02% +5mV (rating current £10A) |
| Ripple and Noise: |
£1mVrms 5Hz ~ 1MHz |
| Recovery Time: |
£100µs (50% load change, minimum load 0.5A) |
| Constant Current Operation |
Line Regulation: |
£0.2% +3mA |
| Load Regulation: |
£0.2% +3mA |
| Ripple Current: |
£3mArms |
| Meter |
Meter Type: |
Dual 3½ digit 0.5" LED displays
(Quad 3½ digit 0.39" LED displays on GPC-3030DQ) |
| Accuracy: |
±(0.5% of reading + 2 digits) |
| Insulation |
Chassis and Terminal: |
20MW or above (500V DC) |
| Chassis and AC Cord: |
30MW or above (500V DC) |
| Other Details |
Power Source: |
AC 100V/120V/220V/240V ±10%, 50/60Hz |
| Accessories: |
Power cord x 1, Instruction manual x 1, Test lead (type GTL-105, type GTL-104 with GPC-1850) x 3 |
| Dimensions: |
255mm (w) x 145mm (h) x 335mm (d) |
| Weight: |
11.5kg |