Hioki 3156-01 Leak Current HiTester
   Search: HACKER SAFE certified sites prevent over 99.9% of hacker crime.
   
   
   




Hioki 3156-01 Leak Current HiTester



Hioki 3156-01
Click image to enlarge

 Hioki 3156-01

Features:
  • Leakage Current Measurement Essential for Electrical Safety
  • Stores data for 100 units
  • Power supply separation
  • Remote Control
  • Automatic measurement function
  • Standard support for a variety of Networks
  • Measurement that complies with standards
  • RS-232C, GP-IB, and EXT I/O interfaces standard
  • Save up to 30 sets of measuring conditions
  • Wiring check functions
Additional Photos
Click image to enlarge




  Description

Because of the risks it poses, many standards stipulate leak current test methods, the performance of test equipment, limits for leak current, and other factors related to leak current. The bench-type 3156 Leak Current HiTester by itself, without any additional equipment, is capable of performing tests that comply with standards for a wide range of equipment, from general purpose electrical equipment to medical electrical equipment. The 3156 also is equipped with functions and an interface that support embedded test lines.

In order to avoid dangers such as electric shock, electrical equipment is designed so that those portions that are likely to be touched by people are insulated from the power source. However, the insulation resistance is not infinite, so there is always some leak current present. Furthermore, the extent of the leak current changes over time as the insulation deteriorates. To determine their safety level, the 3156 Leak Current HiTester can prove to be highly valuable, both as a tester on the production line as well as a maintenance and inspection tool.

The power supply voltage for the 3156 does not need to be changed, even if the supply voltage for the equipment being tested is different. The difference can be managed by controlling the line supply voltage for the equipment being tested. This helps prevent malfunctions due to the wrong voltage being input for the 3156 power supply.





  Models, Options, & Accessories


Model Description List Price Sale Price
3156-01 Leakage Current Tester $3495.00
$3180.45 
Model Accessories List Price Sale Price
9637 RS-232C Cable $40.00
$36.40 
9638 RS-233C Cable $55.00
$50.05 
9151-02 GP-IB Connector Cable $299.00
$272.09 
9151-04 GP-IB Connector Cable $299.00
$272.09 
9442 Printer $649.00
$590.59 
9443-01 Please call or email us for pricing and availability.
9443-02 Please call or email us for pricing and availability.
9443-03 AC Adapter $129.00
$117.39 
9444 Connection Cable $89.00
$80.99 
1196 Recording Paper $99.95
$90.95 
9686 Carrying Case $795.00
$723.45 
9267 Software Package for the Safety Test $169.00
$153.79 



  Specifications

Measurement Mode: Earth leakage current / Leakage current between enclosure and earth / Leakage current between enclosure and enclosure / Leakage current between enclosure and line / Patient leakage current I / Patient leakage current II / Patient leakage current III / Patient auxiliary current
Target Current: DC / AC / AC+DC (25mA max), AC peak (75mA max)
Measurement Range: DC / AC / AC+DC mode ; 50 µ A / 500 µA / 5 mA / 25 mA
AC peak mode ; 500 µA / 1 mA / 10 mA / 75 mA
Measurement System: Indication of a current value calculated based the measured drop in voltage caused by simulated resistance of the human body
Measurement of true effective value
The measurement section ; chassis-grounded and floating
Accuracy: DC / AC / AC+DC mode: ± (2.0 % rdg. + 6 dgt.)
AC peak mode: ± (2.0 % rdg. + 2 dgt.)
DC mode: ± (2.0 % rdg. + 3 dgt.)
Input Resistance: 1 M-ohm ± 1 % (Excluding voltmeter section, simulated resistance of the human body)
Network: For medical electrical equipment / For IEC 60990 / For JIS / For UL / General-purpose 1 / General-purpose 2
Functions: 110 % voltage application function / Wiring check function / Automatic measurement function / Application line selection function / Ground fault prevention function / Setting of single-fault condition / Switching power supply polarity / Setting of measuring time / Measurement delay / Maximum value hold / Allowable value judgement / Data save / Clock / Data back up / etc
Interface: EXT I/O , RS-232C , GP-IB
Power Supply: 100, 120, 220, 240 V AC (default setting) (50/60 Hz, Rated power ; 30VA)
Dimensions: 320 mm(12.6 in)W × 110 mm(4.33)H × 263 mm(10.35)D
Weight: 4.0 kg(141 oz)
Accessories: 9170 TEST LEAD(2), 9195 ENCLOSURE PROBE(1), 9399 CARRYING CASE(1), Alligator clip(3)(2 red, 1 black), AC Power cord(2), Spare fuse(2)



   
  Welcome | Companies | Search by Type | How To Order | Leasing | About Us  
  Tequipment.NET
205 Westwood Ave
Long Branch, NJ 07740
For our Electronic Whiteboard division visit www.touchboards.com
BBBOnline Reliability Seal. Tequipment.NET
Phone: 1-877-742-TEST (8378) / 732-222-7077
Fax: (732) 222-7088
E-mail: salesteam@Tequipment.NET