Hart Scientific 9102S Industrial Calibrators - Handheld Dry-Wells

Click image to enlarge
Hart Scientific 9102S
|
User guide
Additional photos
Click image to enlarge
|
Features:
- Smallest dry-wells in the world
- Ranges from –10°C to 375°C
- Accuracy to ±0.25°C, stability of ±0.05°C at 0°C
- RS-232 interface with Hart software (Model 9930) – NEW!
- We ship Hart Scientific products only within the United States and United States territories. This DOES include US military working abroad. If you have any questions about this policy please call - 1-877-742-8378.
|
Description
Hart’s line of portable dry-wells is incredible. They’re the smallest, lightest, and most portable dry-wells in the world. And now they're better than ever!
For work in the temperature range of –10°C to 122°C, Hart’s Model 9102S dry-well is another first in the industry, featuring stability to ±0.05°C.
This dry-well is only four inches high and six inches wide, achieves temperatures as low as –10°C, includes a NIST-traceable calibration, and is stable to ±0.05°C. The Model 9102S is excellent for dial gauges, digital thermometers, bulb switches, and other sensors that need calibration below ambient.
The 9102S has two wells so you can use one for a REFERENCE THERMOMETER to increase accuracy. Both wells are 1/2 inch in diameter, and each has inserts available for almost any sensor size. The 9102S also has a battery pack option that gives you approximately four hours of field use when AC power is unavailable.
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| 9102S |
Dry-Well, HH Low-Temp |
| $2350.00 |  |
|
Specifications
| |
9102S |
9100S |
| Range
|
–10°C to 122°C (14°F to 252°F) at 23°C ambient
|
35°C to 375°C (95°F to 707°F)
|
| Accuracy
|
±0.25°C |
±0.25°C at 50°C; ±0.25°C at 100°C;
±0.5°C at 375°C
|
| Stability
|
±0.05°C at 0°C
±0.1°C at 122°C
|
±0.07C at 50°C; ±0.1°C at 100°C;
±0.3°C at 375°C
|
| Well-to-Well Uniformity
|
±0.2°C with sensors of similar size at equal depths within wells
|
| Heating Times
|
ambient to 100°C: 10 minutes
|
35° to 375°C: 9.5 minutes
|
| Stabilization
|
7 minutes
|
5 minutes
|
| Cooling Times
|
ambient to 0°C: 10 minutes
|
375°C to 100°C: 14 minutes
|
| Well Depth
|
4 inches (102 mm)
|
4 inches (102 mm); 1/16" (1.6 mm) hole is 3.5" (89 mm) deep
|
| Removable Inserts
|
Available in sizes from 1/16" (1.6 mm) to 7/16" (11.1 mm)
[1/4" (6.4 mm) and 3/16" (4.8 mm) included]
|
N/A
|
| Computer Interface
|
RS-232 included with free Interface-it software
|
| Power
|
94-234 VAC (±10%), 50/60 Hz, 60 W; or 12 VDC
|
115 VAC (±10%), 1.5 A or 230 VAC (±10%), 0.8 A, specify, 50/60 Hz, 175 W
|
| Size
|
3.9" H x 5.5" W x 6.9" D (99 x 140 x 175 mm)
|
2.25" H x 4.9" W x 5.9" D (57 x 125 x 150 mm)
|
| Weight
|
4 lb. (1.8 kg)
|
2 lb. 3 oz. (1 kg)
|
| NIST-Traceable Calibration
|
Data at –10°C, 0°C, 25°C, 75°C, 55°C, 100°C, and 122°C
|
Data at 50°C, 100°C, 150°C, 200°C, 250°C, 300°C, and 375°C
|
|