Fluke 52 Series II Thermometer
| Specifications |
| Measurement accuracy |
Above -100 °C (-148 °F):
J, K, T, E, and N-type: ±[0.05% + 0.3°C (0.5°F)]*
R and S-type: ±[0.05% + 0.4°C (0.7°F)]*
Below -100 °C (-148 °F):
J, K, E, and N-types: ±[0.20% + 0.3°C (0.5°F)]*
T-type: ±[0.50% + 0.3°C (0.5°F)] |
| Display resolution |
0.1°C / °F / K < 1000°
1°C / °F / K ³ 1000° |
| Measurement range |
J-type: -210°C to 1200°C (-346°F to 2192°F)
K-type: -200°C to 1372°C (-328°F to 2501°F)
T-type: -250°C to 400°C (-418°F to 752°F)
E-type: -150°C to 1000°C (-238°F to 1832°F)
N-type: -200°C to 1300°C (-328°F to 2372°F)*
R and S-type: 0°C to 1767°C (32°F to 3212°F)* |
| Temperature scale |
ITS-90 |
| Applicable standards |
NIST-175 |
| Environmental |
Operating temperature:
-10°C to 50°C (14°F to 122°F)
Storage temperature:
-40°C to 60°C (-40°F to 140°F)
Humidity:
0% to 90%; 0°C to 35°C (32°F to 95°F)
0% to 70%; 0°C to 50°C (32°F to 122°F) |
| General |
Unit weight: 400g (14 oz)
Dimensions: 17.3 cm L x 8.6 cm W x 3.8 cm D (6.8 in L x 3.4 in W x 1.5 in D)
Battery: 3 AA batteries; typical 1000-hour life
Certification: CE, CSA, TÜV (pending)
Safety: CSA C22.2 No. 1010.1 1992; EN 61010 Amendments 1,2 |
* Only the Fluke Models 53 and 54 Series II thermometers are capable of measuring N, R, and S-type thermocouples.
| User's Manuals & Other Documentation |
| Product Overview |
51, 52, 53, & 54 Product Overview (PDF, 205kB) |
| User's Manual |
51 & 52 User's Manual (PDF, 158kB) |
| User's Manual |
53 & 54 User's Manual (PDF, 216kB) |
| Service Manual |
51, 52, 53, & 54 Service Manual (PDF, 413kB) |
| Service Manual Supplement |
51, 52, 53, & 54 Service Manual Supplement (PDF, 71kB) |
| Selection Guide |
50 Series Selection Guide (PDF, 557kB) |
|