| Electrical Characteristics |
| Current Range |
2000 A ACRMS or DC |
| Measuring Range |
± 2800 A |
| Output Sensitivity |
1 mV/A |
| Accuracy (at +25°C) |
± 1% of reading ± 500 mA |
| Resolution |
± 100 mA |
| Load Impedance |
> 10 k Ohms and ≤ 100 pF |
| Conductor Position Sensitivity |
± 1.5% relative to centre reading |
| Frequency Range……(small signal) |
DC to 10 kHz (- 1 dB) |
| Temperature Coefficient |
± 0.1% of reading / °C |
| Power Supply |
9 V Alkaline, MN1604/PP3 75 Hours, low battery indicator |
| Working Voltage (see Safety Standards section) |
600 V ACRMS or DC |
| General Characteristics |
| Maximum Conductor Size |
50 mm diameter |
| Output Connection |
4mm safety plugs |
| Output Zero |
Manual adjust via thumbwheel |
| Cable Length |
1.5 meters |
| Operating Temperature Range |
0 to +50 °C |
| Storage Temperature Range (with battery removed) |
-20 to +85 °C |
| Operating Humidity |
15% to 85% (non condensing) |
| Weight |
570 g |
| Safety Standards |
BSEN61010-1: 1993 and Amendment A2: July 1995 BSEN61010-2-032: 1995 BSEN61010-2-031: 1995 600 VRMS, Category III, Pollution Degree 2
Use of the probe on uninsulated conductors is limited to 600 V ACRMS or DC and frequencies below 1 kHz. |
| EMC Standards |
EN61326 :1998 |