Fluke N4K PP30 Power Phase Analyzer

Click image to enlarge
Fluke N4K PP30
|
|
Features
- 0.3 V – 1000 V, 0.03 A – 10 A direct
- Sampling rate 100 kHz
- Bandwidth dc to 1 MHz
- 0.3 % basic accuracy for 2 Years
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| N4K PP30 |
EA1300Z, POWER PHASE VOLT/CUR NORMA 4000 & 5000 |
$695.00 |
| $656.78 |  |
|
| N4K PP40 |
EA1400Z, POWER PHASE VOLT/CUR NORMA 4000 & 5000 |
$2095.00 |
| $1979.78 |  |
|
| N4K PP42 |
EA1420Z, POWER PHASE VOLT/CUR NORMA 4000 & 5000 |
$2395.00 |
| $2263.28 |  |
|
| Voltage |
| 8 ranges |
0.3 – 1 – 3 – 10 – 30 – 100 – 300 – 1000 V Upeak = 2 x range |
| Input impedance |
2 MOhm / 20 pF |
| CMR Common Mode Rejection |
80 dB at 100 kHz |
| Current |
| 6 ranges |
30 – 100 mA – 0.3 – 1 – 3 – 10 A Ipeak = 2 x range; max level 150 % at sine wave (limit of error as at 100 %). |
| Input impedance with integrated shunts |
| Ranges 30 mA, 100 mA |
1 Ohm |
| Ranges 0.3 A, 1 A |
0.1 Ohm |
| Ranges 3 A, 10 A |
0.01 Ohm |
| Current overload |
max. 15A continuous 30 A < 5 sec / 15 sec no load 100 A < 0.1s / 30 sec no load |
| Input for external shunt or probe |
| BNC terminal |
100 kOhm / 30pF 30 – 100 mV – 0.3 – 1 – 3- 10 V |
| Overload |
max. 20 Vrms |
| CMR common mode rejection |
80 dB at 100 kHz |
| Basic accuracy PP30 |
| Sum limit of error |
U |
I |
| Range |
0.15% |
0.15% |
| Reading |
0.15% |
0.15% |
| Voltage and current depending on the input level at 45 – 65 Hz frequency range |
| Input level |
Sum limit of error U |
Sum limit of error I |
| in % |
in % |
in % |
| 100 |
0.30 |
0.30 |
| 50 |
0.45 |
0.45 |
| 30 |
0.65 |
0.65 |
| 10 |
1.65 |
1.65 |
| 5 |
3.15 |
3.15 |
| 3 |
5.15 |
5.15 |
| 1 |
15.15 |
15.15 |
| |
U and I via BNC |
I direct measured |
| Bandwidth -3dB |
1 MHz |
0.3 MHz |
| Current measurement accuracy depending on the frequency (reading + range) |
| Frequency |
Limits of error current [ % ] |
| [Hz] |
Internal shunt |
BNC input |
| 0 |
0.30 |
-0.30 |
0.30 |
-0.30 |
| 10 |
0.30 |
-0.30 |
0.30 |
-0.30 |
| 45 |
0.30 |
-0.30 |
0.30 |
-0.30 |
| 65 |
0.30 |
-0.30 |
0.30 |
-0.30 |
| 500 |
0.30 |
-0.30 |
0.30 |
-0.30 |
| 1000 |
0.30 |
-0.30 |
0.30 |
-0.30 |
| 2000 |
0.36 |
-0.54 |
0.35 |
-0.48 |
| 5000 |
0.44 |
-0.86 |
0.40 |
-0.72 |
| 10000 |
0.50 |
-1.10 |
0.45 |
-0.90 |
| 20000 |
0.56 |
-1.34 |
0.50 |
-1.08 |
| 50000 |
0.64 |
-1.66 |
0.55 |
-1.32 |
| 100000 |
0.70 |
-1.90 |
0.60 |
-1.50 |
| Voltage measurement accuracy depending on the frequency (reading + range) |
| Frequency in Hz |
Limits of error voltage in % |
| 0 |
0.30 |
-0.30 |
| 10 |
0.30 |
-0.30 |
| 45 |
0.30 |
-0.30 |
| 65 |
0.30 |
-0.30 |
| 500 |
0.30 |
-0.30 |
| 1000 |
0.30 |
-0.30 |
| 2000 |
0.35 |
-0.48 |
| 5000 |
0.40 |
-0.72 |
| 10000 |
0.45 |
-0.90 |
| 20000 |
0.50 |
-1.08 |
| 50000 |
0.55 |
-1.32 |
| 100000 |
0.60 |
-1.50 |
| Valid for averaged values at 23 ±1 °C ambient temperature, sine waveform and after 1 hour turn on time with measuring signal. |
| PP30 |
Between U and IBNC |
Between U and IShunt |
| Angular Error |
0.1° +0.1° / kHz |
0.1° +0.1° / kHz |
| Error Power |
| Frequency in Hz |
Limit of error power in % at Lamda = 1 |
| 45 |
0.49 |
-0.49 |
| 65 |
0.49 |
-0.49 |
|