| Technical Data |
| DC Current Output (Internal battery operation) |
Range and Resolution
0.000 to 20.000 mA or 4.000 to 20.000 mA (selectable at power-up). Over-range to 24.000 mA
Drive Capability
12 V compliance, or 500 ohms @20 mA
Accuracy (% of span)
0.05 %
|
| DC Current Simulate (External 24 volt loop supply) |
Range and Resolution
0.000 to 20.000 mA or 4.000 to 20.000 mA (selectable at power-up). Over-range to 24.000 mA
Drive Capability
1000 ohms @20 mA
Accuracy (% of span)
0.05 %
|
| Current Adjustment Modes |
Range and Resolution
Manual: Coarse Fine: 25% step
Automatic: Slow Ramp, Fast Ramp, 25% step |
| Measurement Accuracy |
| VDC |
Range and Resolution:
400.0 mV, 4.000V, 40.00V, 400.0V, 1000V
Best Accuracy (% of reading + LSD):
0.1% + 1
|
VAC (true RMS) to 500 Hz * |
Range and Resolution:
400.0 mV, 4.000V, 40.00V, 400.0V, 1000V
Best Accuracy (% of reading + LSD):
0.7% + 2
|
| mA DC |
Range and Resolution:
30.000 mA
Best Accuracy (% of reading + LSD:
0.05% + 2
|
| A DC |
Range and Resolution:
1.000 A (0.440 A continuous)
Best Accuracy (% of reading + LSD):
0.2% + 2
|
| A AC |
Range and Resolution:
1.000 A (0.440 A continuous)
Best Accuracy (% of reading + LSD):
1% + 2
|
| Resistance |
Range and Resolution:
400.0 Ohms, 4.000 kOhms, 40.00 kOhms, 400.0 kOhms, 4.000 MOhms, 40.00MOhms
Best Accuracy (% of reading + LSD):
0.2% + 1
|
| Frequency (0.5 Hz to 20 kHz) |
Range and Resolution:
199.99 Hz, 1999.9 Hz, 19.999 kHz, 199.99 kHz
Best Accuracy (% of reading + LSD):
0.005% + 1
|
| Diode Test |
Range and Resolution:
2.400 V (measures diode voltage drop)
Best Accuracy (% of reading + LSD):
2% + 1
|
| Continuity |
Range and Resolution
Beeps for resistance < approx. 100 ohms |
*For measurement of adjustable speed motor drive output, consider the
Fluke 39, 41 or 43 power quality analyzers.
| General Specifications |
| Maximum voltage applied between any jack and Earth Ground |
1000 V rms |
| Storage Temperature |
-40 ºC to 60 ºC |
| Operating Temperature |
-20 ºC to 55 ºC |
| Temperature Coefficient |
0.05 x (specified accuracy) per ºC (for temperatures < 18 ºC or > 28 ºC) |
| Relative Humidity |
95% up to 30 ºC; 75% up to 40 ºC; 45% up to 50 ºC; 35% up to 55 ºC |
| Vibration |
Random, 2g, 5-500 Hz |
| Shock |
1 meter drop test |
| Safety |
Designed in accordance with IEC 1010-1, ANSI/ISA S82.01-1994 and CAN/CSA C22.2 No. 1010.1-92 Overvoltage Category III |
| Size |
32 mm H x 87 mm W x 187 mm L (1.25 in H x 3.41 in W x 7.35 in L) |
| Size, with holster and Flex-Stand |
52 mm H x 98 mm W x 201 mm L (2.06 in H x 3.86 in W x 7.93 in L) |
| Weight |
369 g (13 oz); with holster and Flex-Stand 638 g (22.5 oz) |
| Battery |
Single 9V Alkaline battery (ANSI/NEDA 1604A or IEC 6LR61) |
| Battery Life |
50 hours typical (measurement), 12 hours typical (sourcing 12 mA) |
| Warranty |