  • Source Accuracy
  • Measurement Accuracy
  • Temperature, RTDs, and Thermocouples
  • Technical Data
  • General Specifications
  • User's Manuals & Other Documentation


    Source Accuracy
    Range (full scale) Accuracy
    (% of reading+% of full scale)
    110.000 mV 0.01% + 0.005%
    1.10000V 0.01% + 0.005%
    15.0000V 0.01% + 0.005%
    Source 22.000 mA 0.01% +0.015%
    Simulate 22.000 mA 0.02% + 0.03%
    11.000W 0.01% + 20 mW
    110.00W 0.01% + 40 mW
    1.1000 kW 0.02% + 0.5W
    11.000 kW 0.03% + 5W
    0.00 to 10.99 Hz 0.01 Hz
    11.00 to 109.99 Hz 0.1 Hz
    110.0 to 1099.9 Hz 0.1 Hz
    1.100 to 21.999 kHz 2 Hz
    22.000 to 50.000 kHz 5 Hz


    Measurement Accuracy
    Range (Full scale) Accuracy
    (% of reading+% of full scale)
    110.000 mV DC 0.025% + 0.015%
    1.10000V DC 0.025% + 0.005%
    11.0000V DC 0.025% + 0.005%
    110.000V DC 0.05% + 0.005%
    300.00V DC 0.05% + 0.005%
    V AC, 20 to 40 Hz 2% + 10 counts
    V AC, 40 to 500 Hz 0.5% + 5
    V AC, 500 to 1 kHz 2% + 10
    V AC, 1 kHz to 5 kHz 10% + 20
    30.000 mA DC 0.01% + 0.015%
    110.00 mA DC 0.01% + 0.015%
    11.000W 0.05% + 50 mW
    110.00W 0.05% + 50 mW
    1.1000 kW 0.05% + 0.5W
    11.000 kW 0.1% + 10W
    1.00 to 109.99 Hz 0.05 Hz
    110.0 to 1099.9 Hz 0.5 Hz
    1.100 to 10.999 kHz 5 Hz
    11.00 to 50.00 kHz 50 Hz


    Temperature, RTDs, and Thermocouples*
    Device  
    10W Cu (427) Measure Accuracy: 3°C
    Source Accuracy: 1°C
    100W Pt (3916) Measure Accuracy: 0.3°C
    Source Accuracy: 0.1°C
    100W Pt (3926) Measure Accuracy: 0.3°C
    Source Accuracy: 0.1°C
    100W Pt (385) Measure Accuracy: 0.3°C
    Source Accuracy: 0.1°C
    200W Pt (385) Measure Accuracy: 0.3°C
    Source Accuracy: 0.1°C
    500W Pt (385) Measure Accuracy: 0.3°C
    Source Accuracy: 0.1°C
    1000W Pt (385) Measure Accuracy: 0.3°C
    Source Accuracy: 0.1°C
    120W Ni (672) Measure Accuracy: 0.3°C
    Source Accuracy: 0.1°C
    E Measure Accuracy: 0.3°C
    Source Accuracy: 0.2°C
    N Measure Accuracy: 0.5°C
    Source Accuracy: 0.3°C
    J Measure Accuracy: 0.3°C
    Source Accuracy: 0.2°C
    L Measure Accuracy: 0.3°C
    Source Accuracy: 0.2°C
    K Measure Accuracy: 0.3°C
    Source Accuracy: 0.3°C
    T Measure Accuracy: 0.3°C
    Source Accuracy: 0.3°C
    U Measure Accuracy: 0.3°C
    Source Accuracy: 0.3°C
    B Measure Accuracy: 0.9°C
    Source Accuracy: 0.8°C
    R Measure Accuracy: 1.0°C
    Source Accuracy: 0.9°C
    S Measure Accuracy: 0.9°C
    Source Accuracy: 0.9°C
    C Measure Accuracy: 0.6°C
    Source Accuracy: 0.6°C
    * Resolution: 0.1°C, except 1°C for 10W Cu. Best case, mid-range accuracies. Sensor inaccuracies not included. For 2-wire and 3-wire RTD measurements, add 0.4°C. Thermocouple accuracies with external cold junction. For internal junction add 0.2°C.


    Technical Data
    Environmental Specifications* Operating Temperature: -10ºC to 50ºC (-20ºC typical except for frequency and ac voltage measurement)
    Storage Temperature: -20ºC to 60ºC
    Operating Altitude: 2800m above mean sea level (9186 ft)
    Enclosure Protection: Designed to meet IEC529 IP52 (normal operating vacuum for dust)
    Data Log Function Measure functions: Voltage, current, resistance, frequency, temperature, pressure
    Reading rate: 1, 2, 5, 10, 20, 30, or 60 readings per minute
    Maximum record length: 8000 readings (7980 for 30 or 60 readings per minute)
    Ramp Function Source functions: Voltage, current, resistance, frequency, temperature
    Rate: 4 steps/second
    Trip Detect: Continuity or voltage (continuity detection not available when sourcing current)
    Loop Power Function Voltage: Selectable, 24V or 28V
    Accuracy: 5%
    Maximum current: 22 mA, short circuit protected
    Maximum input voltage: 30V dc
    * (all calibrator specifications apply from +18°C to +28°C unless stated otherwise)


    General Specifications
    Dimensions 130 x 236 x 61 mm
    Weight 1.4 kg (3 lb 1 oz)
    Internal Battery Pack NiMH 7.2V, 3500 mAh
    Battery Life Typical usage > eight hours
    Battery Replacement Via snap-shut door without opening calibrator; no tools required
    Side Port Connections
    • Pressure module connector
    • RS-232 connector to interface to your PC
    • Connection for optional battery eliminator
    Safety Complies with CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993
    Data Storage Capacity 1 week of calibration results
    90 day specifications The standard specification intervals for the 744 is 1 and 2 years. Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "%of output" specifications by 2. Floor specifications, expressed as "% of f.s." or "counts" or "ohms" remain constant.


    User's Manuals & Other Documentation
    User's Manual 744 User's Manual (PDF, 1.96MB)
    User's Manual Supplement 744 User's Manual Supplement (PDF, 8kB)
    User's Guide 744 HART Mode User's Guide (PDF, 584kB)
    User's Manual Supplement 744 HART Mode User's Guide Supplement (PDF, 6kB)
    Instruction Sheet 744 V2.0 Upgrade Instruction Sheet (PDF, 10kB)
    Calibration Manual 74X Calibration Manual (PDF, 429kB)
    User's Manual DPC/TRACK User's Manual (PDF, 825kB)
    Installation Guide DPC/TRACK Installation Guide (PDF, 15kB)





   
  Welcome | Companies | Search by Type | How To Order | Leasing | About Us  
  Tequipment.NET
205 Westwood Ave
Long Branch, NJ 07740
For our Electronic Whiteboard division visit www.touchboards.com
BBBOnline Reliability Seal. Tequipment.NET
Phone: 1-877-742-TEST (8378) / 732-222-7077
Fax: (732) 222-7088
E-mail: salesteam@Tequipment.NET