Fluke 743B Documenting Process Calibrator
- Source Accuracy
- Measurement Accuracy
- Temperature, RTDs, and Thermocouples
- General Specifications
- User's Manuals & Other Documentation
| Source Accuracy |
| Range (full scale) |
Accuracy
(% of reading + % of full scale) |
| 110.000 mV |
0.01% + 0.005% |
| 1.10000V |
0.01% + 0.005% |
| 15.0000V |
0.01% + 0.005% |
| Source 22.000 mA |
0.01% + 0.015% |
| Simulate 22.000 mA |
0.02% + 0.03% |
| 11.000W |
0.01% + 20 mW |
| 110.00W |
0.01% + 40 mW |
| 1.1000 kW |
0.02% + 0.5W |
| 11.000 kW |
0.03% + 5W |
| 0.00 to 10.99 Hz |
0.01 Hz |
| 11.00 to 109.99 Hz |
0.1 Hz |
| 110.0 to 1099.9 Hz |
0.1 Hz |
| 1.100 to 21.999 kHz |
2 Hz |
| 22.000 to 50.000 kHz |
5 Hz |
| Measurement Accuracy |
| Range (Full scale) |
Accuracy
(% of reading + % of full scale) |
| 110.000 mV DC |
0.025% + 0.015% |
| 1.10000V DC |
0.025% + 0.005% |
| 11.0000V DC |
0.025% + 0.005% |
| 110.000V DC |
0.05% + 0.005% |
| 300.00V DC |
0.05% + 0.005% |
| V AC, 20 to 40 Hz |
2% + 10 counts |
| V AC, 40 to 500 Hz |
0.5% + 5 |
| V AC, 500 to 1 kHz |
2% + 10 |
| V AC, 1 kHz to 5 kHz |
10% + 20 |
| 30.000 mA DC |
0.01% + 0.015% |
| 110.00 mA DC |
0.01% + 0.015% |
| 11.000W |
0.05% + 50 mW |
| 110.00W |
0.05% + 50 mW |
| 1.1000 kW |
0.05% + 0.5W |
| 11.000 kW |
0.1% + 10W |
| 1.00 to 109.99 Hz |
0.05 Hz |
| 110.0 to 1099.9 Hz |
0.5 Hz |
| 1.100 to 10.999 kHz |
5 Hz |
| 11.00 to 50.00 kHz |
50 Hz |
| Temperature, RTDs, and Thermocouples* |
| Device |
|
| 10W Cu (427) |
Measure Accuracy: 3°C
Source Accuracy: 1°C |
| 100W Pt (3916) |
Measure Accuracy: 0.3°C
Source Accuracy: 0.1°C |
| 100W Pt (3926) |
Measure Accuracy: 0.3°C
Source Accuracy: 0.1°C |
| 100W Pt (385) |
Measure Accuracy: 0.3°C
Source Accuracy: 0.1°C |
| 200W Pt (385) |
Measure Accuracy: 0.3°C
Source Accuracy: 0.1°C |
| 500W Pt (385) |
Measure Accuracy: 0.3°C
Source Accuracy: 0.1°C |
| 1000W Pt (385) |
Measure Accuracy: 0.3°C
Source Accuracy: 0.1°C |
| 120W Ni (672) |
Measure Accuracy: 0.3°C
Source Accuracy: 0.1°C |
| E |
Measure Accuracy: 0.3°C
Source Accuracy: 0.2°C |
| N |
Measure Accuracy: 0.5°C
Source Accuracy: 0.3°C |
| J |
Measure Accuracy: 0.3°C
Source Accuracy: 0.2°C |
| L |
Measure Accuracy: 0.3°C
Source Accuracy: 0.2°C |
| K |
Measure Accuracy: 0.3°C
Source Accuracy: 0.3°C |
| T |
Measure Accuracy: 0.3°C
Source Accuracy: 0.3°C |
| U |
Measure Accuracy: 0.3°C
Source Accuracy: 0.3°C |
| B |
Measure Accuracy: 0.9°C
Source Accuracy: 0.8°C |
| R |
Measure Accuracy: 1.0°C
Source Accuracy: 0.9°C |
| S |
Measure Accuracy: 0.9°C
Source Accuracy: 0.9°C |
| C |
Measure Accuracy: 0.6°C
Source Accuracy: 0.6°C |
* Resolution: 0.1°C, except 1°C for 10W Cu. Best case, mid-range accuracies. Sensor inaccuracies not included. For 2-wire and 3-wire RTD measurements, add 0.4°C. Thermocouple accuracies with external cold junction. For internal junction add 0.2°C.
| General Specifications |
| Dimensions |
130 x 236 x 61 mm |
| Weight |
1.4 kg (3 lb 1 oz) |
| Internal Battery Pack |
NiCd, 7.2V, 1700 mAh |
| Battery Life |
Typical usage > eight hours |
| Battery Replacement |
Via snap-shut door without opening calibrator; no tools required |
| Side Port Connections |
- Pressure module connector
- RS-232 connector to interface to your PC
- Connection for optional battery eliminator
|
| Safety |
Complies with CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993 |
| Data Storage Capacity |
1 week of calibration results |
| 90 Day Specifications |
The standard specification interval for the 743B is 1 or 2 years. Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2. Floor specifications, expressed as "% of f.s.", "counts" or "ohms", remain constant. |
|