Our agents are available:
9:00 AM - 6:00 PM EST
Fluke 726 Multifunction Precision Process Calibrator
   Search: HACKER SAFE certified sites prevent over 99.9% of hacker crime.
   

   
   




Fluke 726 Multifunction Precision Process Calibrator




Measurement Accuracy
Voltage DC 30.000 V 0.01%+ 2 mV (upper display)
20.000 V 0.01%+ 2 mV (lower display)
90.00 mV 0.01%+ 10μV
-10.00 mV to 75.00 mV 0.025 % + 1 count (via TC connector)
Current DC 24.000 mA 0.01%+ 2 mA
Resistance 0.0 to 4000.0 Ω +/- 0.015%
Frequency 2.0 to 1000.0 CPM 0.05 % + 0.1 cpm
1.0 to 1000.0 Hz 0.05 % + 0.1 Hz
1.00 to 15.00 kHz 0.05 % + 0.01 kHz
Sensitivity 1 V peak-to-peak-minimum
Pressure Accuracy from 0.025% of range using any of 29 pressure modules. (for detailed specifications refer to pressure modules in options and accessories) Modules available for differential, gage, vacuum, absolute, dual and high pressure.
Source Accuracy
Voltage DC 100.00 mV 0.01%+10 μV
20.000 V 0.01%+2 mV
-10.00 mV to 75.00 mV 0.025 % + 1 count (via TC connector)
Current DC 24.000 mA (Source) 0.01%+ 2 μA
24.000 mA (Simulate) 0.01%+ 2 μA
Resistance Range, Accuracy (Percent of reading + floor) and Excitation Current
5 to 400 Ω, 0.015% + 0.1 Ω, 0.1 to 0.5mA
5 to 400 Ω, 0.015% + 0.05 Ω, 0.5 to 3mA
400 to 1500 Ω, 0.015% + 0.5 Ω, 0.05 to 0.8 mA
1500 to 4000 Ω, 0.015% + 0.5 Ω, 0.05 to 0.4 mA
Frequency 2.0 to 1000.0 CPM 0.05%
1.0 to 1000.0 Hz 0.05%
1.0 to 10.0 kHz 0.25%
10.0 to 15.0 kHz 0.5%
Waveform 5 V p-p squarewave, -0.1 V offset
RTDs and Thermocouples
Measure and Source Accuracy Cu10 1.8 °C
NI-120 0.15 °C
PT-100 (385) 0.45 °C
PT-100 (392) 0.35 °C
PT-100 (391-JIS) 0.35 °C
PT-200 (385) 0.95 °C
PT-500 (385) 0.55 °C
PT-1000 (385) 0.35 °C
RTD Resolution Cu10 0.1 °C
J 0.4 °C
K 0.5 °C
T 0.4 °C
E 0.4 °C
R 1.4 °C
S 1.4 °C
B 1.4 °C
L 0.45 °C
U 0.45 °C
N 0.6 °C
XK 0.4 °C
BP 1.1 °C
TC Resolution J, K, T, E, L, N, U: 0.1 °C, 0.1 °F B, R, S: 1 °C, 1 °F
Ramp Functions
Source functions Voltage, current, resistance, frequency, temperature
Ramps Slow ramp, Fast ramp, 25% step-ramp
Loop Power Function
Voltage 24 V
Accuracy 10%
Maximum current 22 mA, short circuit protected
Step Functions
Source functions Voltage, current, resistance, frequency, temperature
Steps 25% of range, 100% of range
Environmental Specifications
Operating Temperature -10 °C to +55 °C
Storage Temperature -20 °C to +71 °C
Operating Altitude 3000 m
Safety Specifications
Agency Approvals EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992
Mechanical & General Specifications
Size 200mm x 96mm x 47mm (7.9 in x 3.8 in x 1.9 in)
Weight 650 g (23 oz.)
Batteries 4 AA alkaline batteries
Battery Replacement Separate battery compartment, accessible without breaking calibration seall
Side port connections Pressure module connector, also used for remote real-time programming



   
  Welcome | Companies | Search by Type | How To Order | Leasing | About Us  
  Tequipment.NET
205 Westwood Ave
Long Branch, NJ 07740
For our Electronic Whiteboard division visit www.touchboards.com
BBBOnline Reliability Seal. Tequipment.NET
Phone: 1-877-742-TEST (8378) / 732-222-7077
Fax: (732) 222-7088
E-mail: salesteam@Tequipment.NET