| Measurement Accuracy |
| Voltage DC |
30.000 V 0.01%+ 2 mV (upper display) 20.000 V 0.01%+ 2 mV (lower display) 90.00 mV 0.01%+ 10μV -10.00 mV to 75.00 mV 0.025 % + 1 count (via TC connector) |
| Current DC |
24.000 mA 0.01%+ 2 mA |
| Resistance |
0.0 to 4000.0 Ω +/- 0.015% |
| Frequency |
2.0 to 1000.0 CPM 0.05 % + 0.1 cpm 1.0 to 1000.0 Hz 0.05 % + 0.1 Hz 1.00 to 15.00 kHz 0.05 % + 0.01 kHz |
| Sensitivity |
1 V peak-to-peak-minimum |
| Pressure |
Accuracy from 0.025% of range using any of 29 pressure modules. (for detailed specifications refer to pressure modules in options and accessories) Modules available for differential, gage, vacuum, absolute, dual and high pressure. |
| Source Accuracy |
| Voltage DC |
100.00 mV 0.01%+10 μV 20.000 V 0.01%+2 mV -10.00 mV to 75.00 mV 0.025 % + 1 count (via TC connector) |
| Current DC |
24.000 mA (Source) 0.01%+ 2 μA 24.000 mA (Simulate) 0.01%+ 2 μA |
| Resistance |
Range, Accuracy (Percent of reading + floor) and Excitation Current 5 to 400 Ω, 0.015% + 0.1 Ω, 0.1 to 0.5mA 5 to 400 Ω, 0.015% + 0.05 Ω, 0.5 to 3mA 400 to 1500 Ω, 0.015% + 0.5 Ω, 0.05 to 0.8 mA 1500 to 4000 Ω, 0.015% + 0.5 Ω, 0.05 to 0.4 mA |
| Frequency |
2.0 to 1000.0 CPM 0.05% 1.0 to 1000.0 Hz 0.05% 1.0 to 10.0 kHz 0.25% 10.0 to 15.0 kHz 0.5% |
| Waveform |
5 V p-p squarewave, -0.1 V offset |
| RTDs and Thermocouples |
| Measure and Source Accuracy |
Cu10 1.8 °C NI-120 0.15 °C PT-100 (385) 0.45 °C PT-100 (392) 0.35 °C PT-100 (391-JIS) 0.35 °C PT-200 (385) 0.95 °C PT-500 (385) 0.55 °C PT-1000 (385) 0.35 °C |
| RTD Resolution |
Cu10 0.1 °C |
| J |
0.4 °C |
| K |
0.5 °C |
| T |
0.4 °C |
| E |
0.4 °C |
| R |
1.4 °C |
| S |
1.4 °C |
| B |
1.4 °C |
| L |
0.45 °C |
| U |
0.45 °C |
| N |
0.6 °C |
| XK |
0.4 °C |
| BP |
1.1 °C |
| TC Resolution |
J, K, T, E, L, N, U: 0.1 °C, 0.1 °F B, R, S: 1 °C, 1 °F |
| Ramp Functions |
| Source functions |
Voltage, current, resistance, frequency, temperature |
| Ramps |
Slow ramp, Fast ramp, 25% step-ramp |
| Loop Power Function |
| Voltage |
24 V |
| Accuracy |
10% |
| Maximum current |
22 mA, short circuit protected |
| Step Functions |
| Source functions |
Voltage, current, resistance, frequency, temperature |
| Steps |
25% of range, 100% of range |
| Environmental Specifications |
| Operating Temperature |
-10 °C to +55 °C |
| Storage Temperature |
-20 °C to +71 °C |
| Operating Altitude |
3000 m |
| Safety Specifications |
| Agency Approvals |
EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 |
| Mechanical & General Specifications |
| Size |
200mm x 96mm x 47mm (7.9 in x 3.8 in x 1.9 in) |
| Weight |
650 g (23 oz.) |
| Batteries |
4 AA alkaline batteries |
| Battery Replacement |
Separate battery compartment, accessible without breaking calibration seall |
| Side port connections |
Pressure module connector, also used for remote real-time programming |