Fluke 725 Multifunction Process Calibrator
Summary Specifications (18°C to 28°C for one year)* |
| Voltage (max load, 1 mA) |
0 to 100 mV:
Resolution: 0.01 mV
Accuracy:.02% Rdg + 2 LSD
0 to 10V (source):
Resolution: 0.001V
Accuracy:.02% Rdg + 2 LSD
0 to 30V (measure):
Resolution: 0.001V
Accuracy:.02% Rdg + 2 LSD
|
| mA (max load, 1000Ω) |
0 to 24 mA:
Resolution:0.001 mA
Accuracy: .02% Rdg + 2 LSD
|
| mV (TC terminals) |
-10.00 mV to +75.00 mV:
Resolution: .01 mV
Accuracy: .025% Rdg + 1 LSD
|
| Resistance |
0Ω to 3200Ω (measure):
Resolution: 0.01Ω to 0.1Ω
Accuracy: 0.10Ω to 1.0Ω
15Ω to 3200Ω (source):
Resolution: 0.01Ω to 0.1Ω
Accuracy: 0.10Ω to 1.0Ω
|
| Frequency* * |
1 CPM to 10 kHz (measure):
Resolution: 5 digits
Accuracy: 0.05% Rdg + 1 LSD
2.0 to 1000.0 CPM (source):
Resolution: 0.1 CPM
Accuracy: ±.05%
1 to 1000 Hz (source):
Resolution: 1 Hz
Accuracy: ±.05%
1.0 to 10.0 kHz (source):
Resolution: 0.1 kHz
Accuracy: ±.25%
|
| Loop Supply |
24V dc:
Resolution: N/A
Accuracy: 10% |
* Temperature coefficient, -10°C to 18°C, 28°C to 55°C,±.005% of range per °C.
* * (For frequency source, waveform is 5V p-p squarewave, -0.1V offset)
| Thermocouple Accuracy Specifications* |
| J |
-200 to 0ºC:
Accuracy: 1.0ºC
0 to 1200ºC:
Accuracy: 0.7ºC
|
| K |
-200 to 0ºC:
Accuracy: 1.2ºC
0 to 1370ºC:
Accuracy: 0.8ºC
|
| T |
-200 to 0ºC:
Accuracy: 1.0ºC
0 to 400ºC:
Accuracy: 0.8ºC
|
| E |
-200 to 0ºC:
Accuracy: 0.9ºC
0 to 950ºC:
Accuracy: 0.7ºC
|
| R |
-20 to 0ºC:
Accuracy: 2.5ºC
0 to 500ºC:
Accuracy: 1.8ºC
500 to 1750ºC:
Accuracy: 1.4ºC
|
| S |
-20 to 0ºC:
Accuracy: 2.5ºC
0 to 500ºC:
Accuracy: 1.8ºC
500 to 1750ºC:
Accuracy: 1.5ºC
|
| B |
600 to 800ºC:
Accuracy: 2.2ºC
800 to 1000ºC:
Accuracy: 1.8ºC
1000 to 1800ºC:
Accuracy: 1.4ºC
|
| L |
-200 to 0ºC:
Accuracy: 0.85ºC
0 to 900ºC:
Accuracy:0.7ºC
|
| U |
-200 to 0ºC:
Accuracy: 1.1ºC
0 to 400ºC:
Accuracy: 0.75ºC
|
| N |
-200 to 0ºC:
Accuracy: 1.5ºC
0 to 1300ºC:
Accuracy: 0.9ºC
|
* (Accuracy specifications include 0.2°C cold junction uncertainty)
| Thermocouple Resolution |
| J, K, T, E, L, N, U |
0.1ºC, 0.1ºF |
| B, R, S |
1ºC, 1ºF |
| RTD Ranges and Accuracy Specifications |
Ni 120
|
-80ºC to 260ºC:
Measure: 0.2ºC
Source: 0.2ºC
|
| Pt 100 - 385 |
-200ºC to 800ºC:
Measure: 0.33ºC
Source: 0.33ºC |
| Pt 100 - 3926 |
-200ºC to 630ºC:
Measure: 0.3ºC
Source: 0.3ºC |
Pt 100 - 3916
(JIS) |
-200ºC to 630ºC:
Measure: 0.3ºC
Source: 0.3ºC |
| Pt 200 - 385 |
-200ºC to 250ºC:
Measure: 0.2ºC
Source: 0.2ºC
250ºC to 630ºC:
Measure: 0.8ºC
Accuracy: 0.8ºC |
| Pt 500 - 385 |
-200ºC to 500ºC:
Measure: 0.3ºC
Source: 0.3ºC
500ºC to 630ºC:
Measure: 0.4ºC
Accuracy: 0.4ºC |
| Pt 1000-385 |
-200ºC to 100ºC
Measure: 0.2ºC
Source: 0.2ºC
100ºC to 630ºC
Measure: 0.3ºC
Source: 0.2ºC
|
| RTD Resolution |
| RTD |
0.1ºC, 0.1ºF |
| General Specifications |
| Maximum voltage |
30V |
| Storage Temperature |
-20ºC to 71ºC |
| Operating Temperature |
-10ºC to 55ºC |
| Relative Humidity |
90% (10ºC to 30ºC); 75% (30ºC to 40ºC);
45% (40ºC to 50ºC); 35% (50ºC to 55ºC) |
| Vibration |
Random, 2g, 5-500 Hz |
| Safety |
CSA C22.2 No. 1010.1:1992 |
| EMC |
EN50082-1:1992 and EN55022:1994 Class B |
| Size |
96 x 200 x 47 mm (3.8 x 7.9 x 1.9 inches) |
| Weight |
650g (23oz) |
| Battery |
Four AA alkaline batteries. Battery life: 25 hours typical |
| Warranty |
Three years |
|