Fluke 724 Temperature Calibrator
| Summary Specifications (18ºC to 28ºC for one year)* |
|
Voltage (max load, 1 mA)
|
0 to 100 mV:
Resolution: 0.01 mV
Accuracy:.02% Rdg + 2 LSD
0 to 10V (source):
Resolution: 0.001V
Accuracy:.02% Rdg + 2 LSD
0 to 30V (measure):
Resolution: 0.001V
Accuracy:.02% Rdg + 2 LSD
|
|
mA (max load, 1000W)
|
0 to 24 mA:
Resolution:0.001 mA
Accuracy: .02% Rdg + 2 LSD
|
|
mV (TC terminals)
|
-10.00 mV to +75.00 mV:
Resolution: .01 mV
Accuracy: .025% Rdg + 1 LSD
|
|
Resistance
|
0W to 3200W (measure):
Resolution: 0.01W to 0.1W
Accuracy: 0.10W to 1.0W
15W to 3200W (source):
Resolution: 0.01W to 0.1W
Accuracy: 0.10W to 1.0W
|
|
Loop Supply
|
24V dc:
Resolution: N/A
Accuracy: 10%
|
* Temperature coefficient, -10°C to 18°C, 28°C to 55°C,±.005% of range per °C.
| Thermocouple Accuracy Specifications (measure or source) |
|
J
|
-200 to 0ºC:
Accuracy: 1.0ºC
0 to 1200ºC:
Accuracy: 0.7ºC
|
|
K
|
-200 to 0ºC:
Accuracy: 1.2ºC
0 to 1370ºC:
Accuracy: 0.8ºC
|
|
T
|
-200 to 0ºC:
Accuracy: 1.0ºC
0 to 400ºC:
Accuracy: 0.8ºC
|
|
E
|
-200 to 0ºC:
Accuracy: 0.9ºC
0 to 950ºC:
Accuracy: 0.7ºC
|
|
R
|
-20 to 0ºC:
Accuracy: 2.5ºC
0 to 500ºC:
Accuracy: 1.8ºC
500 to 1750ºC:
Accuracy: 1.4ºC
|
|
S
|
-20 to 0ºC:
Accuracy: 2.5ºC
0 to 500ºC:
Accuracy: 1.8ºC
500 to 1750ºC:
Accuracy: 1.5ºC
|
|
B
|
600 to 800ºC:
Accuracy: 2.2ºC
800 to 1000ºC:
Accuracy: 1.8ºC
1000 to 1800ºC:
Accuracy: 1.4ºC
|
|
L
|
-200 to 0ºC:
Accuracy: 0.85ºC
0 to 900ºC:
Accuracy: 0.7ºC
|
|
U
|
-200 to 0ºC:
Accuracy: 1.1ºC
0 to 400ºC:
Accuracy: 0.75ºC
|
|
N
|
-200 to 0ºC:
Accuracy: 1.5ºC
0 to 1300ºC:
Accuracy: 0.9ºC
|
* (Accuracy specifications include 0.2°C cold junction uncertainty)
| Thermocouple Resolution |
|
J, K, T, E, L, N, U
|
0.1ºC, 0.1ºF
|
|
B, R, S
|
1ºC, 1ºF
|
| RTD Ranges and Accuracy Specifications* |
|
Ni 120
|
-80ºC to 260ºC:
Measure: 0.2ºC
Source: 0.2ºC
|
|
Pt 100 - 385
|
-200ºC to 800ºC:
Measure: 0.33ºC
Source: 0.33ºC
|
|
Pt 100 - 3926
|
-200ºC to 630ºC:
Measure: 0.3ºC
Source: 0.3ºC
|
Pt 100 - 3916 (JIS)
|
-200ºC to 630ºC:
Measure: 0.3ºC
Source: 0.3ºC
|
|
Pt 200 - 385
|
-200ºC to 250ºC:
Measure: 0.2ºC
Source: 0.2ºC
250ºC to 630ºC:
Measure: 0.8ºC
Accuracy: 0.8ºC
|
|
Pt 500 - 385
|
-200ºC to 500ºC:
Measure: 0.3ºC
Source: 0.3ºC
500ºC to 630ºC:
Measure: 0.4ºC
Accuracy: 0.4ºC
|
|
Pt 1000-385
|
-200ºC to 100ºC:
Measure: 0.2ºC
Source: 0.2ºC
100ºC to 630ºC:
Measure: 0.3ºC
Source: 0.2ºC
|
* Note: Measure (4 wire)
| RTD Resolution |
|
RTD
|
0.1ºC, 0.1ºF
|
| General Specifications |
|
Maximum Voltage:
|
30V
|
|
Storage Temperature:
|
-20ºC to 71ºC
|
|
Operating Temperature:
|
-10ºC to 55ºC
|
|
Relative Humidity:
|
90% (10ºC to 30ºC); 75% (30ºC to 40ºC); 45% (40ºC to 50ºC); 35% (50ºC to 55ºC)
|
|
Vibration:
|
Random, 2g, 5-500 Hz
|
|
Safety:
|
CSA C22.2 No. 1010.1:1992
|
|
EMC:
|
EN50082-1:1992 and EN55022:1994 Class B
|
|
Size:
|
96 x 200 x 47 mm (3.8 x 7.9 x 1.9 inches)
|
|
Weight:
|
650g (23oz)
|
|
Battery:
|
Four AA alkaline batteries. Battery life: 25 hours typical
|
|
Warranty:
|
Three years
|
|