Fluke 707 Loop Calibrator |
|
|
| |
|
|
| |
|
Fluke 707 Loop Calibrator
| Summary Specifications (18° to 28° C, one year) |
|
Voltage Measure
|
Range:
0 to 28V
Resolution:
0.001V
Accuracy:
±(.015% Rdg + 2 counts)
|
|
mA Measure
|
Range:
0 to 24
Resolution:
0.001 mA
Accuracy:
.015% Rdg + 2 LSD
|
|
MA source1
|
Range:
0 to 24 mA
Resolution:
0.001 mA
Accuracy:
.015% Rdg + 2 LSD
|
|
MA simulate2
|
Range:
0 to 24 mA
Resolution:
0.001 mA
Accuracy:
.015% Rdg + 2 LSD
|
|
Loop Supply
|
Range:
24V dc
Resolution:
n/a
Accuracy:
24V ± 10%
|
Temperature Coefficient, -10 to 18 șC, 28 to 55 șC, ± .005% of range per șC;
1 Max load, 1200 Ohms; 950 Ohms at 20 mA in HART mode
2 Max applied voltage for simulation, 30 V
| Operating Modes |
|
Measure current
|
|
Measure current/Loop power
|
|
Measure voltage
|
|
Simulate current; manual adjust; scroll up/down
|
|
Simulate current; 25% steps; 0 - 100% steps
|
|
Simulate current; Auto ramp (smooth/step)
|
|
Source current; manual adjust; scroll up/down
|
|
Source current; 25% steps; 0 - 100% steps
|
|
Source current; Auto ramp (smooth/step)
|
| General Specifications |
|
Maximum voltage
|
30V
|
|
Storage temperature
|
-30 to 60șC
|
|
Operating temperature
|
-10 to 55șC
|
|
Relative humidity
|
0 to 90% (0 to 35 șC); 0 to 70% (35 to 55 șC)
|
|
Operating Altitude
|
3,000 meters max.
|
|
Shock and vibration
|
Per MIL-T-28800 for a Class 2 Instrument
|
|
Safety
|
CSA C22.2 No. 1010.1: 1992
|
|
EMC
|
EN50082-1:1992 and EN55022:1994 Class B
|
|
Size/weight
|
6.0 in L x 2.8 in W x 1.6 in D (152 mm L X 70 mm W x 41 mm D) 8 oz (224 g)
|
|
Size/weight with C10 Holster
|
6.4 in L x 2.9 in W x 1.9 in D (164 mm L x 75 mm W x 47 mm D) 12.3 oz. (350 g)
|
|
Battery
|
One 9 V alkaline
|
|
Battery life
|
18 hours typical, at 12 mA
|
|
Warranty
|
Three years
|
|
|
|
| |
Tequipment.NET
205 Westwood Ave
Long Branch, NJ 07740
|
|
Tequipment.NET
Phone: 1-877-571-7901 / 732-222-7077
Fax: (732) 222-7088
E-mail: salesteam@Tequipment.NET
|
|
|
|