Live chat by BoldChat
Our agents are available:
9:00 AM - 6:00 PM EST
Fluke 705
Loop Calibrator
   Search: HACKER SAFE certified sites prevent over 99.9% of hacker crime.
   

   
   




Fluke 705
Loop Calibrator





Summary Specifications (18°C to 28°C for one year)
Voltage Measure Range: 0 to 28V

Resolution: 0.001V

Accuracy: .025% Rdg + 1 LSD
mA Measure Range: 0 to 24

Resolution: 0.001 mA

Accuracy: .02% Rdg + 2 LSD
MA source Range: 0 to 24 mA

Resolution: 0.001 mA

Accuracy: .02% Rdg + 2 LSD
Notes: Max load, 1000 Ohms, Max applied voltage for simulation, 30V
Loop Supply Range: 24 Vdc

Resolution: 10%
Temperature coefficient -10ºC to 18ºC, 28ºC to 55ºC, ±.005% of range per ºC


General Specifications
Maximum voltage 30V
Storage temperature -40ºC to 60ºC
Operating temperature -10ºC to 55ºC
Relative humidity 0 to 90% (0ºC to 35ºC); 0 to 70% (35ºC to 55ºC)
Shock and vibration per MIL-T-28800 for a Class 2 Instrument
Safety CSA C22.2 No. 1010.1:1992
EMC EN50082-1:1992 and EN55022:1994 Class B
Size/weight 152 mm L x 70 mm W x 41 mm D (6.0 in L x 2.8 in. W x 1.6 in D) 340g (12oz)
Battery One 9V battery
Warranty Three years


User's Manuals & Other Documentation
Instruction Sheet 705 Instruction Sheet (PDF, 519kB)
Calibration Information 705 Calibration Information (PDF, 245kB)



   
  Welcome | Companies | Search by Type | How To Order | Leasing | About Us  
  Tequipment.NET
205 Westwood Ave
Long Branch, NJ 07740
For our Electronic Whiteboard division visit www.touchboards.com
BBBOnline Reliability Seal. Tequipment.NET
Phone: 1-877-742-TEST (8378) / 732-222-7077
Fax: (732) 222-7088
E-mail: salesteam@Tequipment.NET