Fluke 337A 1000 Amp True Rms Ac/Dc Digital Clamp Meter With Backlight
| Voltage DC |
| Range |
0 - 600.0 V |
| Accuracy |
1% ± 5 counts |
| Voltage AC |
| Range |
0 - 600.0 V |
| Accuracy |
1% ± 5 counts (20-100 Hz), 6% ± 5 counts (100-400 Hz) |
| AC Response |
True-rms |
| Current DC |
| Range |
0 - 999.9 A |
| Accuracy |
2% ± 3 counts |
| Current AC |
| Range |
0 - 999.9 A |
| Accuracy |
2% ± 5 counts (10-100 Hz), 6% ± 5 counts (100 - 400 Hz) |
| Crest Factor(50/60 Hz) |
3 @ 500A, 2.5 @ 600A, 1.42@1000 A, add 2% for CF>2 |
| AC Response |
True-rms |
| Resistance |
| Range |
0 - 600.0 Ω, 6001 - 6000 Ω |
| Accuracy |
1.5% ± 5 counts |
| Frequency |
| Range |
5 - 400Hz |
| Accuracy |
0.5% ± 5 counts |
| Trigger Level |
10-100 Hz >= 5A, 5-10 Hz, 100-400 Hz >=10A |
| Inrush Current |
| Integration Time |
100 ms |
| Continuity |
<= 30 Ω |
| MIN/MAX |
Yes |
| Operating Temperature |
-10 °C to +50 °C, -14 °F to 122 °F |
| Storage Temperature |
-40ºC to +60ºC, -22 °F to 140 °F |
| Operating Altitude |
2,500 meters max, 6,500 ft max |
| Temperature De-rating |
Add 0.1 x specified accuracy for each ºC above 28ºC or below 18ºC (above 82.4°F or below 64.4°F) |
| Safety Rating |
IEC 1010-2-032, 600V CAT III |
| Size |
250mm x 80mm x 40mm (9.875" x 3.125" x 1.625") |
| Weight |
310 g (11oz) |
| Warranty |
Three years |
| Battery Life |
Alkaline approximately 150 hours continuous Automatic low battery indicator |
| Backlight |
Yes |
| Display Hold |
Yes |
|