Fluke 190 ScopeMeter Series
|
Oscilloscope Mode
|
|
Vertical Deflection
|
Bandwidth:
- Fluke 199= 200 MHz
- Fluke 196= 100 MHz
- Fluke 192= 60 MHz
Rise time:
- Fluke 199= 1.7 ns
- Fluke 196= 3.5 ns
- Fluke 192= 5.8 ns
Bandwidth limiter: User selectable 10 kHz or 20 MHz
Number of inputs: 2 plus external trigger. All isolated from each other and ground.
Input coupling: AC, DC with ground level indicator
Input sensitivity: 5 mV/div to100 V/div
Input voltage: See general specifications for maximum rating
Vertical resolution: 8 bit
Accuracy: ±(1.5% of reading + 0.04 x range/div)
Input impedance: 1 MW ± 1% // 15 pF ± 2 pF
|
|
Horizontal
|
Maximum real-time sample rate:
- Fluke 199= 2.5 GS/s
- Fluke 196= 1 GS/s
- Fluke 192= 500 MS/s
Number of digitizers: 2
Time base range:
- Fluke 199 and 196= 5 ns/div to 5 s/div
- Fluke 192= 10 ns/div to 5 s/div
Maximum record length: 1,000 points per input, 27,500 points per input in ScopeRecord-roll mode, (10 ms/div ... 2 min/div)
Accuracy: ± (0.01% of reading + 1 pixel)
Glitch capture: 50 nsec (5 µsec/div to 1 min/div) Faster time bases have higher sample rates than 20 MS/s
|
|
Display and acquisition *
|
Display modes: Input A, input B, dual, average, persistence, invert, replay
Waveform Mathematics(1): A+B, A-B, A*B all with user selectable scaling of the result ; A versus B (XY-mode)
Acquisition modes: Normal, auto, single shot, ScopeRecord, roll, glitch capture
|
| Trigger and delay |
Source: Input A, input B, external trigger input. All input references isolated from each other and ground.
Modes: Automatic Connect-and-View, free run, single shot, edge, delay, video, video line, selectable pulse width
Connect-and-View: Advanced automatic triggering that recognizes signal patterns, automatically sets up and continuously adjusts triggering, time base and amplitude. Automatically displays stable waveforms of complex and dynamic signals like motor drive and control signals.
Video triggering: NTSC, PAL, PAL+, SECAM. Includes field 1 and 2 and line select.
Pulse width triggering: Pulse width qualified by time. Allows for triggeringt, = t, ¹ t, where t is selectable in minimal steps of 0.01 div or 50 nsec.
Time delay: 9 divisions pre-trigger view to 1,000 divisions trigger delay
|
|
Automatic capture of 100 screens
|
The instrument ALWAYS memorizes last 100 screens (no user interaction or setup required). When an anomaly occurs on screen, theres 10 seconds to press HOLD and review it. If one sets up the instrument for triggering on glitches or intermittent anomalies the unit operates in baby-sit mode and will capture 100 events.
Replay: Manual or continuous replay. Displays the captured 100 screens as a live animation. The contents can also be viewed by manually scrolling backwards and forwards screen by screen.
Replay storage: Up to 2 sets of 100 screens can be saved for later recall and analysis.
|
|
Automatic scope measurements *
|
VDC, VAC rms, VAC+DC, Vpeak max, Vpeak min, Vpeak to peak, frequency (Hz), risetime(1) with markers, falltime(1) with markers positive pulse width, negative pulse width, positive duty cycle, negative duty cycle, amp AC, amp DC, amp AC+DC, power factor, watts, VA, VA reactive, phase, temperature °C, temperature °F, dBV, dBm into 50Ω and 600Ω
|
|
Cursor Measurements *
|
Source: Input A or B
Dual horizontal lines: Voltage at cursor 1 and 2, voltage between cursors
Dual vertical lines: Time between cursors, voltage between markers. Automatic risetime(1) with markers at 10% and 90%, falltime(1) with markers at 10% and 90%
Single vertical line: Min-Max and Average voltage at cursor position
|
|
Zoom
|
Up to 8x horizontal zoom
|
*Items noted by (1) are features of Firmware Version 4.0 or higher.
|
Meter Mode
|
|
General
|
Via 4 mm banana inputs. Fully isolated from scope inputs and scope ground. The specified accuracy is valid over the temperature range 18 °C to 28 °C (15 °F to 82 °F). Add 10 % of specified accuracy for each degree °C below 18 °C or above 28 °C (15 °F to 82 °F).
|
|
Maximum resolution
|
5,000 counts
|
|
Voltmeter ranges
|
500mV, 5V, 50V, 500V, 1,000V
|
|
Accuracy
|
VDC: ± (0.5 % + 5 counts)
VAC true rms
15 Hz to 60 Hz: ± (1 % + 10 counts)
60 Hz to 1 kHz: ± (2.5 % + 15 counts)
VAC+DC true rms
DC to 60 Hz: ± (1 % + 10 counts)
60 Hz to 1 kHz: ± (2.5 % + 15 counts)
|
|
Ohms
|
Ranges: 500Ω, 5kΩ, 50kΩ, 500kΩ, 5MΩ, 30MΩ
Accuracy: ± (0.6 % + 5 counts)
|
|
Other meter functions
|
Continuity: Beeper on < 30Ω (± 10Ω)
Diode test: Up to 2.8V
Amps: Amp DC, Amp AC, Amp AC+DC using an
optional current clamp or shunt. Scaling factors:
0.1 mV/Amp to 100 V/Amp
Temperature (°C, °F): With optional accessories.
Scale factors 1 mV/°C or 1 mV/°F
Input impedance: 1 MΩ ± 1% // 10 pF ± 2 pF
Advanced meter functions: Auto/manual ranging, relative measurements (Zero reference), TrendPlot recording
|
|
Recorder Mode
|
|
ScopeRecord-roll mode
|
Dual input waveform storage mode.
Source and display: Input A, Input B, Dual
Memory depth: 27,500 points per input. Each point consist of Min-Max pair.
Min-Max values: Min-Max values are measured at high sample rate ensuring capture and display of glitches
Sample Recording Settings:
Set Time base range = 10 ms/div to 1 min/div:
- Recorded timespan: 11 sec to 15 hrs
- Glitch capture: 50 nsec
- Sample rate: 20 MS/s
- Resolution: 400 µsec to 2 sec
Set Time base range = 2 min/div:
- Recorded timespan: 30 hrs
- Glitch capture: 250 nsec
- Sample rate: 4 MS/s
- Resolution: 4 sec
Recording modes: Single sweep, continuous roll, start on external trigger
Horizontal scale: Time from start, time of day
Zoom: Up to 100 x zoom
Memory: Up to 2 dual input ScopeRecordings can be saved for later recall and analysis
|
|
TrendPlot Recording
|
Dual input electronic paperless chart recorder. Plots, displays and stores meter and scope measurements.
Source and display: Input A, Input B and DMM input
Memory depth: 13,500 points record per input. Per record point a minimum, maximum and average value, and a time and date stamp are stored.
Ranges: 10 s/div to 20 min/div in normal view mode. 10 min/div to 24 hour/div in view-all mode (overview of total record)
Recorded timespan: Up to 8 days with a resolution of 1 minute
Recording mode: Continuous roll
Measurement speed: 2.5 measurements per second maximum
Horizontal scale: Time from start, time of day
Zoom: Up to 64x zoom
Memory: Up to 2 TrendPlot recordings can be saved for later recall and analysis
|
|
Cursor measurements - All Recorder Modes
|
Source: Input A, B or DMM input
Dual vertical lines: Min-Max or Average voltage. Time between cursors.
Single vertical line: Min-Max or Average voltage. Absolute date and time or time from start.
|
|
General Specifications
|
|
Case
|
Design: Rugged, shock proof with integrated protective holster
Drip and dust proof: IP51 according to IEC529
Shock and vibration: Shock 30g, Vibration 3g
according to MIL-T28800E, Type III, Class 3, Style B
|
|
Display
|
Bright LCD with CCFL backlight, 35/60 cd/m2 without/with power adapter.
Size: 105 x 79 mm (4.1 x 3.1 inches)
Resolution: 240 x 240 pixels
Contrast and brightness: User adjustable, temperature compensated
|
|
Memory Save and Recall
|
Scope memories: 10 memory locations that each can contain two waveforms plus corresponding setup.
Recorder memories: 2 memory locations that each can contain 100 captured dual input scope screens, or a dual input ScopeRecord (27,500 Min-Max pairs per input), or a dual input Trendplot (13,500 Min-Max pairs per input).
|
|
Real-time Clock
|
Time and date stamp for ScopeRecord, 100 captured screens and Trendplots.
|
|
Power
|
Line power: Country specific line voltage adapter/battery charger included
Battery power: Rechargeable NiMH (installed)
Battery operating time: 4 hours
Battery charging time: 4 hours
Battery power saving functions: Auto power down with adjustable power down time. On screen battery power indicator
|
|
Mechanical data
|
Size: 256 x 169 x 64 mm (10.1 x 6.6 x 2.5 inches)
Weight: 1.95 kg (4.3 lbs)
|
|
Safety
|
Compliance: EN61010-1 (1993) Pollution degree 2, UL 3111-1 (1994), CAN/CSA C22.2 No.1010.1 (1992), ANSI/ISA S82.01 (1994)
|
|
Input voltage ratings
|
Maximum probe voltage: 1,000V CAT II, 600V CAT III (Maximum voltage between 10:1 probe tip (VPS200) and reference lead)
Floating voltage: 1,000V CAT II, 600V CAT III (Maximum voltage between earth ground and any terminal (signal input or shielding))
Independently isolated: 1,000V CAT II, 600V CAT III inputs (Maximum voltage between any terminal of one input or probe (VPS200) and any other terminal of another input or probe (VPS200))
Maximum voltage on BNC input directly (input A or B): 300V CAT III
Maximum voltage on meter input: 1,000V CAT II, 600V CAT III
|
|
Environmental
|
Operating temperature: 0 °C to +50 °C
Storage temperature: -20 °C to +60 °C
Humidity 10 °C to 30 °C: 95% RH non condensing 30 °C to 40 °C: 75% RH non condensing 40 °C to 50 °C: 45% RH non condensing
Maximum operating altitude: 3,000 m (10,000 feet)
Maximum storage altitude: 12 km (40,000 feet)
Electro-Magnetic Compatibility: EN 61326-1 for emission and immunity
|
|
Optically isolated PC/Printer interface
|
To printer: Supports HP Laserjet®,Deskjet®,Epson FX/LQ® and postscript printers via optional PAC 91
To PC: Transfer instrument settings, screen images and waveform data, compatible with FlukeView®software for Windows® via optional PM9080.
|
|
Warranty
|
3 years
|
|