| Voltage DC |
| Maximum Voltage |
1000 V |
| Accuracy |
±(0.09%+2) |
| Maximum Resolution |
0.001V |
| Voltage AC |
| Maximum Voltage |
1000 V |
| Accuracy |
±(2%+3) |
| AC Bandwidth |
5kHz with low pass filter; 3db @ 800 Hz |
| Maximum Resolution |
0.1 mV |
| Current DC |
| Maximum Amps |
400mA |
| Amps Accuracy |
±(0.2%+2) |
| Maximum Resolution |
0.01 mA |
| Current AC |
| Maximum Amps |
400mA |
| Amps Accuracy |
±(1.5%+2) 1kHz bandwidth |
| Maximum Resolution |
0.01 mA |
| Resistance |
| Maximum Resistance |
50 MΩ |
| Accuracy |
±(0.9%+2) |
| Maximum Resolution |
0.1Ω |
| Capacitance |
| Maximum Capacitance |
9,999 μF |
| Accuracy |
±(1.2%+2) |
| Maximum Resolution |
1 nF |
| Frequency |
| Maximum Frequency |
100 kHz |
| Accuracy |
±(0.1%+1) |
| Maximum Resolution |
0.01 Hz |
| Temperature Measurement |
-40.0 °C to 537 °C -40.0 °F to 998 °F excluding probe |
| Diode Test |
| Range |
6 V |
| Resolution |
1 mV |
| Accuracy |
±(2 % + 1) |
| Insulation Test |
| Minimum Test Current @ 1kΩ/V |
1mA |
| Test Voltages |
50, 100, 250, 500, 1000 V |
| Maximum Resistance per Test Voltage |
100V: 2 GΩ |
| Maximum Resolution per Test Voltage |
50V: 0.01 MΩ 100V: 0.01 MΩ 250V: 0.1 MΩ 500V: 0.1 MΩ 100V: 0.1 MΩ |
| Accuracy per Test Voltage |
50V: ±(3 % + 5) 100V: ±(3 % + 5) 250V: ±(1.5 % + 5) 500V: ±(1.5 % + 5) 100V: ±(1.5 % + 5) to 600 MΩ, ±(10 % + 3) above 600 MΩ |
| Environmental Specifications |
| Operating temperature |
–20 °C to +55 °C |
| Storage temperature |
-40°C to +60°C |
| Humidity (Without Condensation) |
0% - 95% (10 °C – 30 °C) 0% - 75% (30 °C – 40 °C) 0% - 40% (40 °C – 55 °C) |
| Operating Altitude |
2000 m |
| Safety Specifications |
| Overvoltage category |
Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 NO. 61010-1-04, and IEC/EN 61010-1 2nd Edition for measurement CAT III 1000 V and CAT IV 600 V |
| Agency Approvals |
2000 m CAT III 1000 V, CAT IV 600 V; 3000 m CAT II 1000 V, CAT III 600 V |
| Mechanical & General Specifications |
| Size |
203 x 100 x 50 mm (with holster) |
| Weight |
624 g |
| Warranty |
3 years |
| Battery life Alkaline |
Meter use 1000 hours |
| Insulation test use |
Meter can perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into 1 MΩ with a duty cycle of 5 seconds on and 25 seconds off. |
| IP Rating |
IP40 |