company logo Fluke 116 HVAC Multimeter with Temperature and Microamps
     Search: HACKER SAFE certified sites prevent over 99.9% of hacker crime.  
 

 
 




Fluke 116 HVAC Multimeter with Temperature and Microamps



General specifications:

     Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 95 %.
The accuracy specifications take the form of: ± ( [ % of Reading ] + [ Counts ] )


Maximum voltage between any terminal and earth ground 600 V
Surge protection 6 kV peak per IEC 61010-1 600 V CAT III, Pollution Degree 2
Display Digital: 6,000 counts, updates 4/sec
Bar graph 33 segments, updates 32/sec
Operating temperature -10 °C to + 50 °C
Storage temperature -40 °C to + 60 °C
Battery 9 volt Alkaline, NEDA 1604A/IEC 6LR61
Battery life 400 hours typical, without backlight


Measurement Range Resolution Accuracy ± ([% of reading] + [counts])
DC millivolts 600.0 mV 0.1 mV 2.0 % + 3
DC volts 6.000 V 0.001 V  
60.00 V 0.01 V
600.0 V 0.1 V
Auto volts 600.0 V 0.1 V 2.0 % + 3 (dc, 45 Hz to 500 Hz)
4.0 % + 3 (500 Hz to 1 kHz)
AC millivolts1 true-rms 600.0 mV 0.1 mV 1.0 % + 3 (dc, 45 Hz to 500 Hz)
2.0 % + 3 (500 Hz to 1 kHz)
AC volts1 true-rms 6.000 V 0.001 V 1.0 % + 3 (45 Hz to 500 Hz)
2.0 % + 3 (500 Hz to 1 kHz)
60.00 V 0.01 V
600.0 V 0.1 V
Continuity 600 O 1 O Beeper on < 20 O off > 250 O; detects opens or shorts of 500 µs or longer.
Ohms 600.0 O 0.1O 0.9 % + 2
6.000 kO 0.001 kO 0.9 % + 1
60.00 kO 0.01 kO
600.0 kO 0.1 kO
6.000 MO 0.001 MO
40.00 MO 0.01 MO 1.5 % + 2
Diode test 2.000 V 0.001 V 0.9 % + 2
Capacitance 1000 nF 1 nF 1.9 % + 2
10.00 µF 0.01 µF
100.0 µF 0.1 µF
9999 µF 1 µF  
100 µF to 1000 µF   1.9 % + 2
> 1000 µF   5 % + 20
Lo-Z capacitance 1 nF to 500 µF   10 % + 2 typical
Temperature2 (Type-K thermocouple) -40 °C to 400 °C 0.1 °C 1 % + 102
-40 °F to 752 °F 0.2 °F 1 % + 182
AC µamps true-rms (45 Hz to 500 Hz) 600.0 µA 0.1 µA 1.5 % + 3 (2.5 % + 3 > 500 Hz)
DC µamps 600.0 µA 0.1 µA 1.0 % + 2
Hz (V or A input)3 99.99 Hz 0.01 Hz 0.1 % + 2
999.9 Hz 0.1 Hz
9.999 kHz 0.001 kHz
50.00 kHz 0.01 kHz

1 All ac voltage ranges are specified from 1 % to 100 % of range. Because inputs below 1 % of range are not specified, it is normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected from a circuit or are shorted together. For volts, crest factor of = 3 at 4000 counts, decreasing linearly to 1.5 at full scale. AC volts is ac-coupled. Auto-V LoZ, and ac mV are dc-coupled.
2 Temperature uncertainty (accuracy) does not include the error of the thermocouple probe.
3 Frequency is ac coupled, 5 Hz to 50 kHz for ac voltage. Frequency is dc coupled, 45 Hz to 5 kHz for ac current.




 

Welcome | Companies | Search by Type | How To Order | Leasing | About Us | Careers
  Tequipment.NET
205 Westwood Ave
Long Branch, NJ 07740
BBBOnline Reliability Seal, Click for confirmation page. Tequipment.NET
Phone: 1-877-742-TEST (8378) / 732-222-7077
Fax: (732) 222-7088
E-mail: salesteam@Tequipment.NET
 




Fluke Educational Grant Program