company logo Compliance West
HT-25000DC DC Hi-Pot Tester
     Search: HACKER SAFE certified sites prevent over 99.9% of hacker crime.  
 
 
 



Compliance West
HT-25000DC DC Hi-Pot Tester



Compliance HT-25kVDC
Click image to enlarge

 Compliance HT-25kVDC

Features:
  • DC output to 25Kv. Delivers up to 5 mA leakage current.
  • Logic Controlled, One-button testing. LED indicators. Buzzer upon failure, test terminates. Test termination defeatable by switch on rear panel. Provided with arc detection circuit to properly detect hipot failures.
  • Rear panel adjustments: leakage limit, ramp time, test time and voltage.
  • Automatic voltage ramp-down circuit eliminates residual voltage.
  • Designed for product development and laboratory use.
  • High voltage test jack output.
  • Calibrated to NIST standards.
  • 1 year warranty



  Models, Options, & Accessories


Model Description List Price Sale Price
HT-25kVdc Hipot Tester, 0-25kVdc @ 5-2mA dc
$3995.00 
Model Accessories List Price Sale Price
VF Voltage Hold on Failure
$200.00 
LL2 2-200uA Leakage Limit Range
$350.00 
LL5 5-500uA Leakage Limit Range
$350.00 
240 240V only operation
$200.00 
MT 120/220-240V operation
$225.00 
RI Relay Isolated Remote Interface board
$350.00 
MHC Manufactured Housing Cordset (2 red, 1 black)
$40.00 
MHC-RV Recreational Vehicle Cordset (2 red, 1 black, small alligator clips)
$40.00 



    Options:




    Specifications:

Electrical
Output: 0-25000 Volts DC
Leakage Current: 0-5 mA DC from 0-18kVdc output, derated linearly to 2mA output current at 25kVdc.
Pass/Fail Criteria: Leakage Current: Pass/Fail point user adjustable.
Dielectric Breakdown: Separate high frequency detection circuit for breakdown spike detection
Test Time: User adjustable 1-60 sec., defeatable.
Voltage Ramp Up Time: User adjustable 1-5 sec., defeatable
Voltage Ramp Down Time: Factory set 8 sec. maximum (open circuit)
Pass/Fail Repeatability: ± 2%
Duty Cycle: 100%
Test Adjustments: Ramp Time (using Ramp Control), Test Time (using Timer Control), Leakage Limit (using Shutdown Limit), Voltage Adjust, Ramp Control ON/DEFEAT, Timer Control ON/DEFEAT, Shutdown Limit ON/DEFEAT


Environmental
Operating Temperature: 0-50°C at 5 – 45% RH
0-40°C at 5 – 75% RH
0-30°C at 5 – 95% RH
Non-Operating Temp: -40° - 71°C


General
Input power requirements: 108-132 volts, 50/60 Hz, 2A max, or 220 volts, 50 Hz, 2A max.
Weight: 32 lbs.


Safety Agency Topics
Transformer Output: < 500VA
Visual Indication of Voltage Output: Provided by front panel meter, directly connected to high voltage output
Failure Indication: Audible, provided by internal buzzer
Visual, provided by red LEDs on front panel
Test can be automatically terminated on failure
Leakage Test: Provided; 2 mA DC factory set pass/fail point, user adjustable. 120Kohm calibration resistor provided on rear panel.




 

Welcome | Companies | Search by Type | How To Order | Leasing | About Us | Careers
  Tequipment.NET
205 Westwood Ave
Long Branch, NJ 07740
BBBOnline Reliability Seal, Click for confirmation page. Tequipment.NET
Phone: 1-877-571-7901 / 732-222-7077
Fax: (732) 222-7088
E-mail: salesteam@Tequipment.NET
 



Live chat by BoldChat
Need help? We have a staff
of more than 30 people to help you.
Let us show you why over 100,000
customers choose Tequipment.NET