Compliance West HT-25000DC DC Hi-Pot Tester

Click image to enlarge
Compliance HT-25kVDC
|
|
Features:
- DC output to 25Kv. Delivers up to 5 mA leakage current.
- Logic Controlled, One-button testing. LED indicators. Buzzer upon failure, test terminates. Test termination defeatable by switch on rear panel. Provided with arc detection circuit to properly detect hipot failures.
- Rear panel adjustments: leakage limit, ramp time, test time and voltage.
- Automatic voltage ramp-down circuit eliminates residual voltage.
- Designed for product development and laboratory use.
- High voltage test jack output.
- Calibrated to NIST standards.
- 1 year warranty
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| HT-25kVdc |
Hipot Tester, 0-25kVdc @ 5-2mA dc |
| $3995.00 |  |
|
| Model |
Accessories |
List Price |
Sale Price |
| VF |
Voltage Hold on Failure |
| $200.00 |  |
|
| LL2 |
2-200uA Leakage Limit Range |
| $350.00 |  |
|
| LL5 |
5-500uA Leakage Limit Range |
| $350.00 |  |
|
| 240 |
240V only operation |
| $200.00 |  |
|
| MT |
120/220-240V operation |
| $225.00 |  |
|
| RI |
Relay Isolated Remote Interface board |
| $350.00 |  |
|
| MHC |
Manufactured Housing Cordset (2 red, 1 black) |
| $40.00 |  |
|
| MHC-RV |
Recreational Vehicle Cordset (2 red, 1 black, small alligator clips) |
| $40.00 |  |
|
Options:
Specifications:
|
Electrical
|
|
Output:
|
0-25000 Volts DC
|
|
Leakage Current:
|
0-5 mA DC from 0-18kVdc output, derated linearly to 2mA output current at 25kVdc.
|
|
Pass/Fail Criteria:
|
Leakage Current: Pass/Fail point user adjustable.
Dielectric Breakdown: Separate high frequency detection circuit for breakdown spike detection
|
|
Test Time:
|
User adjustable 1-60 sec., defeatable.
|
|
Voltage Ramp Up Time:
|
User adjustable 1-5 sec., defeatable
|
|
Voltage Ramp Down Time:
|
Factory set 8 sec. maximum (open circuit)
|
|
Pass/Fail Repeatability:
|
± 2%
|
|
Duty Cycle:
|
100%
|
|
Test Adjustments:
|
Ramp Time (using Ramp Control), Test Time (using Timer Control), Leakage Limit (using Shutdown Limit), Voltage Adjust, Ramp Control ON/DEFEAT, Timer Control ON/DEFEAT, Shutdown Limit ON/DEFEAT
|
|
Environmental
|
|
Operating Temperature:
|
0-50°C at 5 – 45% RH
0-40°C at 5 – 75% RH
0-30°C at 5 – 95% RH
|
|
Non-Operating Temp:
|
-40° - 71°C
|
|
General
|
|
Input power requirements:
|
108-132 volts, 50/60 Hz, 2A max, or 220 volts, 50 Hz, 2A max.
|
|
Weight:
|
32 lbs.
|
|
Safety Agency Topics
|
|
Transformer Output:
|
< 500VA
|
|
Visual Indication of Voltage Output:
|
Provided by front panel meter, directly connected to high voltage output
|
|
Failure Indication:
|
Audible, provided by internal buzzer
Visual, provided by red LEDs on front panel
Test can be automatically terminated on failure
|
|
Leakage Test:
|
Provided; 2 mA DC factory set pass/fail point, user adjustable. 120Kohm calibration resistor provided on rear panel.
|
|