Compliance West HT-20000AC/DC AC Hi-Pot Tester
Click image to enlarge
Compliance West HT-20000AC/DC
|
| Model |
Description |
List Price |
Sale Price |
| HT-20kVac/dc |
Hipot Tester, 0-10kVac @ 10mA / 0-25kVdc @ 2mA - Discontinued - no replacement |
| Model |
Accessories |
List Price |
Sale Price |
| VF |
Voltage Hold on Failure |
| $200.00 |  |
|
| 240 |
220/240V only operation |
| $145.00 |  |
|
| MT |
120/220-240V operation |
| $225.00 |  |
|
| RI |
Relay Isolated Remote Interface board |
| $350.00 |  |
|
| MHC |
Manufactured Housing Cordset (2 red, 1 black) |
| $42.00 |  |
|
| MHC-RV |
Recreational Vehicle Cordset (2 red, 1 black, small alligator clips) |
| $40.00 |  |
|
Options:
Specifications:
|
Electrical
|
|
Output:
|
0-10000 Volts AC or 0-25000 Volts DC
|
|
Leakage Current:
|
0-5 mA AC from 0-9500Vac output, derated linearly to 2mA output current at 10kVac.
0-5 mA DC from 0-18kVdc output, derated linearly to 2mA output current at 25kVdc.
|
|
Pass/Fail Criteria:
|
Leakage Current: Pass/Fail point user adjustable.
Dielectric Breakdown: Separate high frequency detection circuit for breakdown spike detection
|
|
Test Time:
|
User adjustable 1-60 sec., defeatable.
|
|
Voltage Ramp Up Time:
|
User adjustable 1-5 sec., defeatable
|
|
Voltage Ramp Down Time:
|
Factory set 8 sec. maximum (open circuit)
|
|
Pass/Fail Repeatability:
|
± 2%
|
|
Duty Cycle:
|
100%
|
|
Test Adjustments:
|
Ramp Time (using Ramp Control), Test Time (using Timer Control), Leakage Limit (using Shutdown Limit), Voltage Adjust, Ramp Control ON/DEFEAT, Timer Control ON/DEFEAT, Shutdown Limit ON/DEFEAT
|
|
Environmental
|
|
Operating Temperature:
|
0-50°C at 5 – 45% RH
0-40°C at 5 – 75% RH
0-30°C at 5 – 95% RH
|
|
Non-Operating Temp:
|
-40° - 71°C
|
|
General
|
|
Input power requirements:
|
108-132 volts, 50/60 Hz, 2A max, or 220 volts, 50 Hz, 2A max.
|
|
Weight:
|
32 lbs.
|
|