Compliance West HT-10000DC DC Hi-Pot Tester
Click image to enlarge
Compliance West
HT-10000DC
|
Features:
- DC output to 14000 Volts. Delivers up to 5 milliamps.
- Logic controlled, one button test. LEDs indicate test progress. Upon failure, buzzer sounds and testing is terminated.
- Leakage current limit, voltage ramp-up time and test duration are controlled from the rear panel. Output voltage adjustable from front and rear.
- Automatic voltage rampdown circuit eliminates residual voltage.
- Designed for product development and laboratory use.
- High voltage test jack output.
- Voltage ramp, shutdown on failure and test timer are defeatable.
- 1 year warranty
|
| The HT-10000 is available for very high voltage dielectric withstand testing, in either an AC Voltage or DC Voltage output design. The front panel voltage dial-up makes this unit easy to use. As with our HT-5000, test termination on failure, voltage ramp time, and the test timer may be disabled for specific research. Provided with leads, manual, and calibration certificate. |
| Model |
Description |
List Price |
Sale Price |
| HT-10kVDC |
Hipot Tester, 0-14kVdc @ 5mA |
| $2495.00 |  |
|
| HT-10kVAC |
Hipot Tester, 0-10kVac @ 10mA |
| $2395.00 |  |
|
| HT-10kac/dc |
Hipot Tester, 0-10kVac @ 10mA / 0-15kVdc @ 5 mA |
| $3095.00 |  |
|
| Model |
Accessories |
List Price |
Sale Price |
| TM |
TestMinder HT |
| $350.00 |  |
|
| 240 |
220/240V only operation |
| $145.00 |  |
|
| MT |
120/220-240V operation |
| $225.00 |  |
|
| RI |
Relay Isolated Remote Interface board |
| $350.00 |  |
|
| LC |
Leakage Current Display |
| $350.00 |  |
|
| MHC |
Manufactured Housing Cordset (2 red, 1 black) |
| $42.00 |  |
|
| MHC-RV |
Recreational Vehicle Cordset (2 red, 1 black, small alligator clips) |
| $40.00 |  |
|
Click here to download the HT-10000 Data Sheet in PDF format
| Electrical |
| Output: |
14,000 Volts DC. Leakage current 0-5 mA DC |
| Pass/Fail Criteria: |
Leakage Current: Pass/Fail point user adjustable
Dielectric Breakdown: Separate high frequency detection circuit for breakdown spike detection |
| Test Time: |
User adjustable 1-60 sec., defeatable. |
| Voltage Ramp Time: |
User adjustable .5-5 sec., defeatable. |
| Voltage Ramp Down Time: |
Factory set 8 sec. maximum |
| Pass/Fail Repeatability: |
± 3% |
| Duty Cycle: |
100% |
| Meter Accuracy: |
±3% Full Scale |
| Test Adjustments: |
Rear panel: Ramp Time / Test Time / Leakage Limit / Voltage Adjust / Voltage Ramp Defeat / Timer Duration Defeat / Shutdown on Failure Defeat. |
| Environmental |
| Operating Temperature: |
15-40°C |
| Relative Humidity Range: |
0-90% non-condensing |
| General |
| Input Power Requirements: |
114-127 volts, 50/60 Hz |
| Weight: |
20 lbs. |
| Dimensions: |
11.250(W) x 4.250(H) x 12.000(L) |
| RESET and TEST Lamp Type: |
Replace with type 73 14V lamp |
| Safety |
| Transformer Output: |
<500 VA |
| Visual Indication Of Voltage Output: |
Provided by front panel meter, directly connected to high voltage output. |
| Failure Indication: |
Audible, provided by internal buzzer. Visual, provided by red LEDs on front panel. Test automatically terminated on failure. |
| Leakage Test: |
Provided; 5mA AC factory set pass/fail point, user adjustable. 120KΩ calibration resistor provided on rear panel. |
|