Compliance West ATS-8 AC/DC Hi-Pot Tester
Click image to enlarge
Compliance West ATS-8
|
Features:
- AC Output to 2000 volts; delivers up to 20 milliamps or DC output to 2800 volts: delivers up to 5 milliamps.
- Performs continuity check, leakage test and high voltage dielectric withstand test with one button operation.
- Automatically tests up to 8 channels, in any combination.
- Adjustable voltage ramp-up, test time and leakage limits
- Performs to UL, CE, IEC, CSA, TUV safety agency requirements for production line testing.
- Other configurations / options available as well as higher voltages and leakage currents. Consult factory.
- 1 year warranty.
|
|
The ATS-8 is ideal for custom testing of multi-phase motors, wiring harnesses and other applications where up to 8 points must be checked for hipot or continuity. Differential mode testing allows any point in the circuit to be tested to any other point, up to 28 tests per sequence. Easy rear-panel programming. Custom configurations feature AC or DC testing, custom reporting and up to 5000 VAC or 7200 VDC. This unit performs all the production line dielectric withstand tests required by UL CSA and CE Standards. Provided with custom setup to customer requirements, manual and pigtail leads. 2-channel tester is stock; 8 channels max.
|
| Model |
Description |
List Price |
Sale Price |
| ATS-8 |
Please call or email us for pricing and availability. |
Click here to download the ATS-8 Data Sheet in PDF format
Specifications:
|
Electrical
|
|
Input:
|
114-127 volts, 50/60 Hz
|
|
Output:
|
0-2000 volts AC or 0-2800 Volts DC between channels programmed for dielectric withstand by rear panel DIP switches. ATS-8 equpped with high voltage leads for configuration to customer test jigs. Other termination options available. Consult factory.
|
|
Dielectric Channels:
|
Dielectric tests can be completed between channels as programmed by Dielectric Test Channel DIP Switches on rear panel.
|
|
Continuity Channels:
|
0-8 continuity tests can be completed on the two wires of each channel as programmed by Continuity Test DIP Switches on rear panel.
|
|
Maximum Number of Tests:
|
36; any combination of Continuity and Dielectric.
|
|
Leakage Current:
|
0-5 mA
|
|
Pass/Fail Criteria:
|
Continuity: 3Ω ± 5%* (defeatable)
Leakage Current: User adjustable 1-5 mA
Dielectric Breakdown: Separate high frequency detection circuit for breakdown spike detection.
|
|
Test Time:
|
User adjustable 1-60 sec.
|
|
Voltage Ramp Time:
|
User adjustable .5-5 sec.
|
|
Pass/Fail Repeatability:
|
± 3%
|
|
Duty Cycle:
|
100%
|
|
Meter Accuracy:
|
±3% Full Scale
|
|
Discharge Time:
|
Voltage Dependent; discharge progress displayed on front panel meter
|
|
Test Adjustments:
|
Tamper-proof operation available.
|
|
Signal I/O:
|
Remote control and dry contact relays available. Consult factory.
|
|
Environmental
|
|
Operating Temperature:
|
15-40°C
|
|
Relative Humidity Range:
|
0-90% non-condensing
|
|
Dimensions:
|
Standard rackmount chassis. 19” x 5.25” x 15”
|
|
Safety
|
|
Transformer Output:
|
<500 VA
|
|
Visual Indication Of Voltage Output:
|
Provided by front panel meter, directly connected to high voltage output.
|
|
Failure Indication:
|
Audible, provided by internal buzzer
|
|
Continuity Test:
|
Visual, provided by red LED’s on front panel Test automatically terminated on failure.Provided; 3Ω pass/fail point.
|
|
Leakage Test:
|
Provided; user adjustable.
|
|