Compliance West 10x700-2.5 MegaPulse Impulse Tester
Click image to enlarge
Compliance West 10x700-2.5
|
Features:
- Generates either of the waveforms described in EN60950, Annex N, Table N.1, for testing electrical separation (MegaPulse 10/700 Series) or creepage in floating secondary circuits (MegaPulse 1/2/50 Series). Use a general purpose oscilloscope to read and record test results.
- Theoretical graphs and actual graphs showing the output of your tester are included with each MegaPulse, allowing easy sanity checks and proof to visiting agency engineers of test setups.
- 1000:1 BNC Oscilloscope output option allows quick connections and easy evaluation of output waveforms.
- Reverse polarity waveforms available by pushing a button on the front panel. No need to change the test setup. Positive polarity indication minimizes errors.
- 1 year warranty
|
The 10/700-2.5 is designed to test the separation of TNV-1 and -3 circuits from user-contacted parts as described by the test in Paragraph 6.4.2.1. The waveform generated is in accordance with the instructions in Annex N, Table N.1, line 1. In accordance with these requirements, the 10/700-2.5 can generate the required waveform with up to 2,500 volts peak A front panel switch reverses output polarity. Use with a general purpose oscilloscope for waveform capture.
The MegaPulse 10/700-2.5 ships with cables, graphs of theoretical and actual waveforms, and a Calibration Certificate. One year warranty. Options include relay outputs for PLC control, and a 1000:1 oscilloscope output BNC jack.
|
| Model |
Description |
List Price |
Sale Price |
| 10x700-2.5 |
2500V, 10x700-2.5 Voltage Waveform, 39 ohm 1 chnl |
| $2795.00 |  |
|
| Model |
Accessories |
List Price |
Sale Price |
| IER |
IEC 65 Figure 7b Switch |
| $1650.00 |  |
|
| IFD |
MegaPulse Insulation Fault Detector, Adjustable trip |
| $860.00 |  |
|
Options:
Specifications:
|
Specifications
|
|
Voltage Rise Time:
|
10 μs
|
|
Voltage Duration:
|
700 μs |
|
Current Rise Time:
|
6 μs
|
|
Current Duration:
|
300 μs
|
|
Max Voltage:
|
2500V
|
|
Meter Accuracy:
|
±120V
Better than 1% of full-scale reading
|
|
Operating Temperature:
|
15-40°C
|
|
Relative Humidity Range:
|
0-90% Non-Condensing
|
|
Input Power Requirements:
|
120 V AC, 50/60 Hz, 1 Amp
230 V optional; check with factory
|
|
Dimensions:
|
11.250 (W) x 12.000 (D) x 5.000 (H) in.
|
|
Weight:
|
16 lbs.
|
|