BK 5370 Multimeter 0.05% TRUE RMS IEC1010 DMM

Click image to enlarge
BK 5370
|
|
Features:
- Precision 50,000 count instruments
- 0.05% Basic DCV accuracy
- True RMS reading
- Large LCD display provides easy readability, all test parameters in clear view. Backlight LCD display
- CE safety approved
|
Description
|
The 5370 handheld DMM is one of the most advanced digital multimeters available in it price range. This meter incorporates a special 50,000 count integrated circuit for high accuracy (0.05%) basic accuracy, high reliability design. Measure AC Volts (True RMS AC, AC + DC), Resistance, Frequency, Temperature and Capacitance. This meter also feature fuse and battery access without breaking a calibration seal, great for ISO facilities.
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| 5370 |
0.05% TRUE RMS IEC1010 DMM - Discontinued - suggested replacement: Protek 608 |
| Model |
Accessories |
List Price |
Sale Price |
| TL 130A |
General Purpose DMM Kit |
| $42.00 |  |
|
Specifications
| Basic Functions |
| Basic Accuracy |
0.05% |
| Voltage (Best Resolution) |
750 VAC/1000 VDC (10uV) |
| Current (Best Resolution) |
10AAC/DC (10nA) |
| Frequency, Temperature Probe, Duty Cycle, Backlight |
| CONTINUITY |
| Threshold |
10 to 20 Ohm |
| DIODE TEST |
Measures forward voltage drop of diode 0 to 1.999V~ |
| dB MEASUREMENT |
5390 and 5370 - Measures value relative to a reference from 1 to 9999, 5380 at 600 Ohm |
| PEAK HOLD |
Selection: Peak + or Peak - (surge or sag) ACQ, Time: < 1rms |
| DUTY CYCLE |
5390, 5370, 5360-Modes: Selectable %+, or %- Min |
| EVENT COUNTER (Model 5390) - Range |
0 to 99999 (Totalize) |
| MIN PULSE DURATION |
2us |
| PULSE WIDTH (Model 5390) Modes |
Selectable +, - Min. Pulse Width: 20 us Max Pulse Period: 12.5us |
| TEMPERATURE (Model 5370) |
Using Model TP-5300 Probe (Optional) |
| RANGE |
-200 0 to +800 0C |
| Computer Interface |
RS232 capability built-in. Requires AK-5300 |
| GENERAL |
| Display |
50000 count resolution LCD, 0.55'' (14mm) digit height, 34 segment analog bargraph |
| Measurement Rate |
2/sec., 20/sec - bargraph |
| Power |
9V (NEDA 1604) Battery Life: 500 hrs |
| Dimensions |
1.57x3.23x7.44'' (40x82x189mm) |
| Weight |
14 oz (397 g) |
| Three Year Warranty |
| AC VOLTAGE (Auto/Manual Ranging, True RMS) |
| Modes |
AC or AC + DC, selectable |
| Ranges |
4 kHz to 10 kHz |
10 kHz to 30 kHz |
30 kHz to 50 kHz |
Resolution |
Input Impedance |
| 500 mV |
5%R + 30D |
Not specified |
10uV |
10 MOHM/100 pF or 1 GOHM |
| 5V |
2%R + 30D |
2%R + 30D |
3%R + 30D |
100uV |
11 M/100 pF |
| 50V |
2%R + 30D |
2%R + 30D |
3%R + 30D 1 mV |
10 M/100 pF |
|
| 500V |
2%R + 30D |
2%R + 30D |
3%R + 30D 10 mV |
10 M/100 pF |
|
| 750V |
Not specified |
100 mV |
10 M/100 pF |
| 50 kHz to 100 kHz (5V, 50V, 500V ranges): +-6% FS typical |
| Crest Factor |
6 |
Unless otherwise stated, accuracy specifications apply from 18 0 C to 28 0 C. Specifications stated as n%R + nD, mean +-(n% of reading +n least significant digits). Accuracy specifications stated as n%FS, mean +-(n% of Full Scale) Factory Calibrated Only |
| DC VOLTAGE (Auto Manual Ranging) |
Accuracy |
| 500 mV |
0.05%R + 2D |
| 5V 0.025%R + 2D |
0.05%R + 2D |
| 50V |
0.05%R + 2D |
| 500V |
0.05%R + 2D |
| 1000V |
0.05%R + 2D |
| AC/DC VOLTAGE PROTECTION |
| Overvoltage Protection |
+-1100 V DC + AC Peak |
| Transient Protection |
6 kV for 10 us |
| DC CURRENT (Auto/Manual Ranging) |
| Ranges |
| 500 uA |
0.2%R + 5D |
| 5 mA |
0.2%R + 2D |
| 50 mA |
0.05%R + 2D |
| 500 mA |
0.2%R + 2D |
| 10 A |
0.5%R + 2D |
| AC CURRENT (Auto/Manual Ranging, True RMS) AC or AC + DC |
| Accuracy at greater than 10% of range, 50,000 counts at Full Scale: |
| Ranges |
10 kHz to 20 kHz |
20 kHz to 30 kHz |
Resolution |
Input Impedance |
| 500 uA |
1.0%R + 50D |
3.0%R + 50D |
10 nA |
700 mV |
| 5 mA |
1.0%R + 50D |
2.0%R + 50D |
100 nA |
700 mV |
| 50 mA |
1.0%R + 50D |
2.0%R + 50D |
1 uA |
700 mV |
| 500 mA |
Not specified |
10 uA |
1.5V |
| 10 A |
Not specified |
1mA |
500 mV |
| Crest Factor: 6 |
| RESISTANCE (Auto/Manual Ranging) |
| Accuracy |
| Ranges |
Resolution |
| 500 |
10 m |
| 5 k |
100 m |
| 50 k |
1 |
| 500 k |
10 |
| 5 M |
100 |
| 50 M |
1 k |
| FREQUENCY (Autoranging) |
| Range |
0.6 Hz to 500 kHz |
| Accuracy |
+-(0.03% rdg + 1 digit) |
| Ranges |
750V |
| 0.6Hz to 5 kHz |
100 V |
| 5 kHz to 50 kHz |
250 V |
| 50 kHz to 500 kHz |
Not specified |
| CAPACITANCE (Auto/Manual Ranging) |
| Range |
Measurement Resolution |
| 50 nF |
10 pF |
| 500 nF |
100 pF |
| 5 pF |
1 nF |
| 500 uF |
100 nF |
| 5000 uF |
1 uF |
| 50 mF |
10 uF |
| Accessories |
| SUPPLIED |
Test Leads, Spare Fuses, Instruction Manual |
| OPTIONAL |
TL-130A General Purpose DMM Kit TP-5300 Platinum Temperature Probe (for 5370) |
|