| Frequency Characteristics (Standard Waveforms) |
| Sine |
0.1 Hz to 20 MHz |
| Square |
0.1 Hz to 20 MHz |
| Triangle, Ramp |
0.1 Hz to 2 MHz |
| Accuracy |
0.01% (100 ppm) |
| Resolution |
4 digits or 10 mHz |
| Output Characteristics |
| Amplitude Range |
10 mV to 10 Vp-p into 50 Ω |
| Resolution |
3 digits (1000 counts) |
| Amplitude Accuracy |
±2% ± 20 mV of the programmed output from 1.01 V - 10 V |
| Flatness |
0.5 dB at 1 MHz 1 dB to 20 MHz |
| Offset Range |
±4.5 V into 50 Ω, depending on the Amplitude setting |
| Offset Resolution |
10 mV with 3 digits resolution |
| Offset Accuracy |
±2% ± 10 mV into 50 Ω |
| Output Impedance |
50 Ω |
| Output Protection |
The instrument output is protected against short circuit or accidental voltage practically available in electronic laboratories, applied to the main output connector |
| Waveform Characteristics |
| Harmonic Distortion |
0 - 20 kHz, -50 dBc 20 kHz - 100 kHz, -45 dBc 100 kHz - 1 MHz, -40 dBc 1 MHz - 20 MHz, -30 dBc |
| Spurious |
DC - 1 MHz, < -55 dBc |
| Square Rise/Fall Time |
< 20 ns (10% to 90%) at full amplitude into 50 Ω |
| Variable Duty Cycle |
20% to 80% to 2 MHz for Square and 10% - 90% for Triangle |
| Symmetry at 50% |
< 1% |
| Operating Modes |
| Continuous |
Output continuous at programmed parameters. |
| Triggered |
Output quiescent until triggered by an internal or external trigger, then one waveform cycle is generated to programmed parameters, up to 2 MHz |
| Gate |
Same as triggered mode, except waveform is executed for the duration of the gate signal. The last cycle started is completed. |
| Trigger Source |
Trigger source may be internal, external or manual. Internal trigger rate 10 µs to 10 s. |
| Modulation Characteristics |
| Amplitude Modulation |
| Internal |
Sine signal of 1000 Hz Variable modulation from 0% to 100% in 1% steps |
| External |
5 Vp-p for 100% modulation, 10 kΩ input impedance, DC to 20 kHz bandwidth. |
| Frequency Modulation |
| Internal |
Sine signal of 1000 Hz |
| External |
5 Vp-p for 100% deviation, 10 kΩ input impedance, DC to 20 kHz bandwidth. |
| Sweep Characteristics |
| Sweep Shape |
Linear and Logarithmic, up or down |
| Sweep Time |
10 ms to 50 s. |
| Inputs and Outputs |
| Trigger In |
TTL compatible. Max. rate 2 MHz. Minimum width 50 ns. |
| Sync Out |
TTL pulse at programmed frequency, 50 Ω source impedance. |
| Modulation In |
5 Vp-p for 100% modulation. 10 kW input impedance. DC to > 20 kHz minimum bandwidth. |
| General |
| Dimensions |
8.4'' (213 mm) wide 3.5'' (88 mm) high 8.3'' (210 mm) deep |
| Weight |
Approx 2.5 kg. |
| Power |
90 V - 264 V, 30 VA max |
| Temperature |
Operating 0°C to +50°C Non-operating -10°C to +70°C |
| EMC |
According to EN55011 for radiated and conducted emissions. |
| Electrical Discharge Immunity |
According to EN55082 |
| Safety Specifications |
According to EN61010 |