Altek 934 Multifunction Loop Calibrator
   Search: HACKER SAFE certified sites prevent over 99.9% of hacker crime.
   
   
   




Altek 934 Multifunction Loop Calibrator



Altek 21037E
Click image to enlarge

 Altek 21037E

Features:
  • Simultaneous Source & Read — Isolated to 500 V DC. Dual 5 digit Liquid Crystal Display. Fuseless protection to 120 V AC/DC.
  • Multifunction — Source to 24.000 mA, 999.90 mV or 20.00 V DC, read to 24.000 and 52.00 mA, 999.90 mV or 99.990 V DC. Plus simulate and power/measure 2-wire transmitters.
  • High Resolution — Up to 1 microamp and 10 microvolt. Convenient ranges for % of 4-20 mA, % DP flow and user adjustable engineering units.
  • High Accuracy — ±0.02 % of reading. Source to within 0.004 mA, 0.2 mV or 0.004 V.
  • QUIK-CHEK Switch — Instantly recall any 3 outputs for each range. Fast, easy adjustment with our multi-speed "digital potentiometer."
  • AUTO-CHEK — Automatically sequenced outputs in 2, 3, 5 or 10 equal steps at intervals of 5 to 900 seconds.
Additional Photos
Click image to enlarge




  Description

Simultaneously source and read any combination of process milliamp and voltage signals. Altek's Model 934 combines all the functionality of a milliamp and a voltage calibrator! This dual function calibrator continuously displays the output and input signals. The top half of the display indicates the output setting while the lower half of the display shows the signal being measured.

Calibrate state of the art equipment such as smart transmitters and indicators to 41/2 or more digits using the Model 934. Full 5 digit resolution lets you check any brand of process control equipment. Your calibrations are correct with accuracy surpassing 0.02% of reading.

Use the scaleable display to match the readings on the Model 934 to that of the equipment being calibrated. For example, if your 4 to 20 mA recorder scale is 0.0 to 300.0 gallons/hour you can set the 934 to display 0.0 for 4 mA and 300.0 for 20 mA. No more need to use your calculator or "cheat sheet" to come up with the correct milliamp values.

Six "AA" alkaline batteries provide more than 20 hours sourcing 20 mA continuously . . . more than 3 months of typical use. From the time the low battery indicator first comes on you have hours of power remaining to get you through a shift without interruption. The 934 comes configured to automatically power down after 30 minutes of inactivity to save battery life.

Calibrate loop converters and isolators by simultaneously providing the input signal and measuring the output of the converter or isolator. Internal optoisolators withstand up to 500 VDC between the source and read leads. Calibrate and checkout all your process milliamp and voltage instruments with the "benchtop accurate" Model 934.

Use at any point in your 4 to 20 milliamp loop. Source milliamps directly into your receivers with loop loads up to 1200 Ohms. Simulate 2-Wire Transmitters for checkout of field wiring and total loop testing. Power and measure 2-Wire or 4-Wire transmitters on the bench before installation. Read Transmitter & Controller outputs up to 52.00 mA. Display directly in milliamps, percent of 4 to 20 mA, percent of DP Flow or engineering units to match the scale of your process.

Calibrate and checkout all your DC Voltage instrumentation. Source up to 999.90 millivolts to simulate sensors and up to 20.000 Volts to check your process signal inputs. Measure voltages from -99.990 to +99.990 VDC to check your power supplies and the outputs from any process voltage equipment. Calibrate any 1 to 5 Volt device in a live 4 to 20 milliamp loop without disconnecting any wires. The Model 934 will automatically source or sink 0 to >16 milliamps to clamp the test voltage in all source ranges. Output impedance is less than 0.2 Ohm to let you calibrate voltage instruments in milliamp loops without removing the input load resistors.





  Models, Options, & Accessories


Model Description List Price Sale Price
21037E Altek Model 934 Calibrator, Item # 21037E $2036.00
$1812.04 



  Specifications

Milliamp Source  
Ranges: 0.00 to 24.000mA
-25.00 to 125.00% of 4 to 20mA
-50.0 to 111.8% of % DP Flow
Accuracy: ±(0.012% of Reading + 0.008% of 24.000 mA)
Typical Drive Capability: 1200 Ohms @ 20.00 mA
Power to Drive Receiver: Nominal 24 VDC
Compliance Voltage: 0 to 285 Ohm Load: 10V nominal
285 to 1200 Ohm Load: 24V nominal
Power & Measure 2-wire Transmitters  
Ranges: 0.00 to 24.000mA
-25.00 to 125.00% of 4 to 20mA
-50.0 to 111.8% of % DP Flow
Accuracy: ±(0.012% of Reading + 0.008% of 24.000 mA)
Output Current: Up to 24.00 mA
Typical Drive Capability: 1200 Ohms @ 20.00 mA
Power to External 2-Wire Transmitter: nominal 24 VDC
Overload Protection: Current limited to 25 mA
2-wire Transmitter Simulator  
Ranges: 3.000 to 24.000 mA
-6.25 to 125.00% of 4 to 20 mA
-25.0 to 111.8% of % DP Flow
Accuracy: ±(0.012% of Reading + 0.008% of 24.000 mA)
Loop Voltage Limits: Minimum: 3 VDC
Maximum: 45 VDC
Overload Protection: Current limited to 25 mA nominal
Milliamps Read  
Ranges: -52.00 to -24.001
-24.000 to +24.000 mA2
24.00 to 52.00 mA1
-25.00 to 125.00% of 4 to 20 mA
-50.0 to 111.8% of % DP Flow
Accuracy: ±(0.012% of Reading + 0.008% of 52.00mA)
±(0.012% of Reading + 0.008% of 24.000 mA)
Input Impedance: 25 Ohms
Voltage Burden: 0.2 V at 4 mA, 0.6 V at 20 mA, 1.4 V at 50 mA
Measuring Across a Diode: Add ±0.2% to specification from 0 to 20 mA (Diode 1N 914 or equivalent)
Voltage Source  
Ranges: 0.000 to 999.90mVDC
0.000 to 20.000 VDC
Accuracy: ±(0.012% of Reading + 0.008% of 999.90 mV)
±(0.012% of Reading + 0.008% of 20.000 V)
Source Current: >24mA
Sink Current: >16mA from 1 to 5 VDC
Output Impedance: Greater than 0.2 Ohms
Load Resistance: > 3 Ohms
Short Circuit Duration: Infinite
Overload Protection: Current limited to 50 mA nominal
Voltage Read  
Ranges: -99.990 to -0.999 VDC
-999.90 to 999.90 mVDC
0.999 to 99.990 VDC
Accuracy: ±(0.012% of Reading + 0.008% of 99.900 V)
±(0.012% of Reading + 0.008% of 999.90 mV)
Input Resistance: 2 Meg Ohm nominal
Source Resistance Effect: 0.01% error per 200 Ohms
Maximum Voltage Limit: 120 Volts DC to 100 Hz AC
General  
Accuracy: ±(0.012% of Reading + 0.008% of Full Scale Reading)
Warm up Time: 10 Seconds to specified accuracy, 2 minutes to maximum accuracy
Temperature Effect: ±0.008%/°C based on 23°C ±25°C
Batteries: Six "AA" (R6) batteries (Alkaline supplied and recommended)
Battery Life: Milliamp Source and 2-wire Modes: Nominal 33 hours at 12 mA, 20 hours at 20 mA with 250 ohm load
Other Functions: Nominal 40 hours
AC Adapters: Optional, nominal 115 or 230 VAC, 50/60 Hz
Low Battery Indication: "BAT" indication on the display, approximately 4 hours left
Overvoltage Protection: Protected to 120 Volts AC or DC in all ranges without fuses for 30 seconds
Output Noise: ±1 LSD at frequencies less than 10 Hz
Normal Mode Rejection Ratio: 50dB @ 50/60 Hz
Common Mode Rejection Ratio: Not applicable for battery operation. 120 dB @ 50/60 Hz with AC Adaptor
Operating Temperature Range: -5 to +130 °F (-20 to +55°C)
Storage Temperature Range: -13 to +130 °F (-25 to +55°C)
Relative Humidity: 10 to 90%, non-condensing for 24 hours
Dimensions: 170mm x 102mm x 75mm
Weight: 1lb 9oz (0.709 kg)



   
  Welcome | Companies | Search by Type | How To Order | Leasing | About Us  
  Tequipment.NET
205 Westwood Ave
Long Branch, NJ 07740
For our Electronic Whiteboard division visit www.touchboards.com
BBBOnline Reliability Seal. Tequipment.NET
Phone: 1-877-742-TEST (8378) / 732-222-7077
Fax: (732) 222-7088
E-mail: salesteam@Tequipment.NET