Altek 422 14-Type Universal Thermocouple Calibrator
   Search: HACKER SAFE certified sites prevent over 99.9% of hacker crime.
   
   
   




Altek 422 14-Type Universal Thermocouple Calibrator



Altek 5086T
Click image to enlarge

 Altek 5086T

Features:
  • 14 T/C type plus mV - J, K, T, E, R, S, B, N, G, C, D, P, L, U & mV
  • Temperature input & output - Reads directly in degrees
  • "Quik-Chek™ " switch - Instantly recall three outputs; HI, LO & Set
  • 45 output memories - Three for each T/C type & mV
  • Accuracy ±(0.008% of Reading+0.006 mV) - Typical accuracy of 0.4°C or 0.7°F
    Field selectable 0.1° or 1µV resolution
Additional Photos
Click image to enlarge





  Description

Source and read T/C’s over the entire industrial temperature range with Altek’s Model 422 Thermocouple Calibrator. Use with transmitters, recorders, controllers, alarms, indicators, data acquisition and computer systems. Switch between 14 T/C types or millivolts.

High accuracy and stability is achieved through Altek’s exclusive isothermal block. Make your connections directly with thermocouple wire or with miniature thermocouple connectors.

Field customize the Model 422 to lock-in 0.1° or 1° resolution, fixed °F or °C or front panel selectable °F/°C operation. A shrouded miniature thermocouple connector receptacle plus terminal crews allow for easy hookups. Built in protection guards the Model 422 against mis-connection to 120 Volts AC or DC, in any mode.

The Model 422 turns on to the T/C type last used. Other T/C types may be selected each time the unit is turned on. If you use less than 14 T/C types, you can create your own T/C List. By following a simple setup procedure you can deselect unwanted T/C types to customize T/C types available at turn on.

Select resolution of 0.1° or 1° for the full listed range of all thermocouple types. Millivolts allows 1 microvolt resolution from -99.999 to 99.999mV. The Model 422 simulates key temperatures for repetitive calibrations. “Quik-Chek” function stores three output temperatures for each T/C type (45 total) for real convenience. Three memories are retained for each thermocouple type even when the power is off.

Turn the knob to check trip points, controller action or hysteresis. The fast response 422 sets quickly without overshoot but allows slow changes at your own rate.

The Model 422 display gives you fast, accurate temperature measurement with 0.1 and 1 degree or with 0.001 millivolt resolution. High resistance or open T/Cs and leads are detected and indicated on the LCD display. Two readings per second track fast moving temperatures.

“MAX” and “MIN” memories are continuously updated from turn-on or whenever the “RESET” pushbutton is pressed. The Model 422 gives you a handy tool to monitor temperatures for drift or control deviation. Just flip the “Quik-Chek” switch to display the minimum and maximum temperature since reset.

Each time the Model 422 is turned on, the LCD will display all segments for 1 second. It then displays the currently selected thermocouple type or mV for approximately 3 seconds. If °C/°F operation has been selected the currently selected temperature scale of °C or °F will display.





  Models, Options, & Accessories


Model Description List Price Sale Price
5086T Universal Thermocouple, Item # 5086T $1343.00
$1195.27 



  Specifications

Thermocouple Simulator (Source)  
Output Impedance: Less than 0.1 ohms
Source Current: Up to 8 mA (drives 80mV into 10 ohms)
Output Noise: Less than 4 microvolts p-p for frequencies of 10 Hz or below
Overload: Indicates OVER and blanks digits on the display
Read a Thermocouple  
Input Impedance: > 10 Megohms
Open Thermocouple Detection: 450 millisecond check pulse. Nominal threshold, 10 K ohms.
Displays "_ _ _ _" for open circuit.
Normal Mode Rejection: 50/60 Hz, 50 dB
Common Mode Rejection: 50/60 Hz, 120 db
General  
Accuracy: ±(0.008% of reading + 0.006 mV)
Cold Junction Compensation: Built-in for specific thermocouple type, characterized to T/C curve
Cold Junction Temperature Effect: Within 0.05° per °C change in ambient temperature over operating range
Operating Temperature Range: -5 to +140° F (-20 to +60° C)
Storage Temperature Range: -22 to +175° F (-30 to +80° C)
Relative Humidity: 10 to 90%, non-condensing
Zero Stability: Included in cold junction effect
Warm up Time: 1 minute to full rated accuracy
Overload Protection: 120 Volts AC/DC for 30 seconds on connecting leads, in any mode
Battery Life: 9-volt Alkaline Battery - Nominal 40 hours
Low Battery Indicator: "BAT" indication on LCD at 7 Volts nominal, approximately 10 hours left; Batteries should be removed when storing the unit for longer than three months.
Reference Drift: Less than 10 PPM/°C
Dimensions: 63.5mm x 66.7mm x 130mm
Weight: 10.9oz (0.31 kg)
Accessories: Carrying Case included with zipper, belt loop, and shoulder strap.



   
  Welcome | Companies | Search by Type | How To Order | Leasing | About Us  
  Tequipment.NET
205 Westwood Ave
Long Branch, NJ 07740
For our Electronic Whiteboard division visit www.touchboards.com
BBBOnline Reliability Seal. Tequipment.NET
Phone: 1-877-742-TEST (8378) / 732-222-7077
Fax: (732) 222-7088
E-mail: salesteam@Tequipment.NET