Agilent N3303A Electronic Load Module

Click image to enlarge
Agilent N3303A
|
|
Features:
- Stable operation down to zero volts
- Sinks up to 250 Watts, at up to 10 amps and up to 240 volts
- 16 bit programming and read-back
- Downloadable lists for increased speed
- Flexible triggering system
- Programmable measurement accuracy/speed
- Built-in pulse waveform generator
- Option UJ1 8mm screw connector for test systems
|
Description
The Agilent N3303A is a 250 W electronic load module for use in either the N3300A or N3301A electronic load mainframe in system applications. The N3303A load module is fast and accurate, for programming in constant current, constant voltage, or constant resistance modes, or for making voltage, current or power measurements. The N3303A can also digitize waveforms.
A single GPIB address is all you need for complete control and read-back of the N3303A load module within a single N3300A or N3301A mainframe. The N3303A electronic load provides fast operation and the accurate programming and read-back needed for high-volume DC power supply test systems.
Today’s high volume manufacturing requires optimization of test system throughput, to maximize production volume without increasing floorspace. The N3300A Series electronic loads can help you in a number of ways to achieve this goal.
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| N3300A |
DC Electronic Load Mainframe, 1800 W max, 6 slots |
| $2372.00 |  |
|
| N3301A |
DC Electronic Load Mainframe, 600 W max, 2 slots |
| $2015.00 |  |
|
| N3302A |
DC Electronic Load Module, 0-60V, 30A,150 W, for N330X mainframes, 1 slot |
| $1653.00 |  |
|
| N3303A |
DC Electronic Load Module, 0-240V, 10A, 250 W, for N330X mainframes, 1 slot |
| $2354.00 |  |
|
| N3304A |
DC Electronic Load Module, 0-60V, 60A, 300 W, for N330X mainframes, 1 slot |
| $2044.00 |  |
|
| N3305A |
DC Electronic Load Module, 0-150V, 60A, 500 W, for N330X mainframes, 2 slots |
| $3014.00 |  |
|
| N3306A |
DC Electronic Load Module, 0-60V, 120A, 600 W, for N330X mainframes, 2 slots |
| $2900.00 |  |
|
| N3307A |
DC Electronic Load Module, 0-150V, 30A, 250W, for N330X mainframes, 1 slot |
| $2194.00 |  |
|
| Model |
Options |
List Price |
Sale Price |
| N3303A-A6J |
Certificate of calibration - ANSI/NCSL Z540, printed |
| $350.00 |  |
|
| N3303A-UK6 |
Commercial calibration certificate with test data included |
| $147.00 |  |
|
Specifications
| |
N3302A |
N3303A |
N3304A |
N3305A |
N3306A |
N3307A |
| Input Ratings |
| Current |
0 - 30 A |
0 - 10 A |
0 - 60 A |
0 - 60 A |
0 - 120 A |
0 - 30 A |
| Voltage |
0 - 60 V |
0 - 240 V |
0 - 60 V |
0 - 150 V |
0 - 60 V |
0 - 150 V |
| Maximum Power @ 40ºC1 |
150 W |
250 W |
300 W |
500 W |
600 W |
250 W |
| Constant Current Mode2 |
| Low Range/High Range |
3 A/30 A |
1 A/10 A |
6 A/60 A |
6 A/60 A |
12 A/120 A |
3 A/30 A |
| Regulation |
10 mA |
8 mA |
10 mA |
10 mA |
10 mA |
10 mA |
| Low Range Accuracy |
0.1% + 5 mA |
0.1% + 4 mA |
0.1% + 7.5 mA |
0.1% + 7.5 mA |
0.1% + 15 mA |
0.1% + 7.5 mA |
| High Range Accuracy |
0.1% + 10 mA |
0.1% + 7.5 mA |
0.1% + 15 mA |
0.1% + 15 mA |
0.1% + 37.5 mA |
0.1% + 15 mA |
| Constant Voltage Mode2 |
| Low Range/High Range |
6 V/60 V |
24 V/240 V |
6 V/60V |
15 V/150 V |
6 V/60 V |
15 V/150 V |
| Regulation |
5 mV |
10 mV |
10 mV |
10 mV |
20 mV |
10 mV |
| Low Range Accuracy |
0.1% + 3 mV |
0.1% + 10 mV |
0.1% + 3 mV |
0.1% + 10 mV |
0.1% + 3 mV |
0.1% + 10 mV |
| High Range Accuracy |
0.1% + 8 mV |
0.1% + 40 mV |
0.1% + 8 mV |
0.1% + 20 mV |
0.1% + 8 mV |
0.1% + 20 mV |
| Constant Resistance Mode2 |
| Range 1 (I >10% of current rating) |
0.067-4 ? |
0.2-48 ? |
0.033-2 ? |
0.033-5 ? |
0.017-1 ? |
0.067-10 ? |
| Range 2 (I >1% of current rating) |
3.6-40 ? |
44-480 ? |
1.8-20 ? |
4.5-50 ? |
0.9-10 ? |
9-100 ? |
| Range 3 (I >0.1% of current rating) |
36-400 ? |
440-4800 ? |
18-200 ? |
45-500 ? |
9-100 ? |
90-1000 ? |
| Range 4 (I >0.01% of current rating) |
360-2000 ? |
4400-12000 ? |
180-2000 ? |
450-2500 ? |
90-1000 ? |
900-2500 ? |
| Transient Generator |
| Frequency Range |
0.25 Hz - 10 kHz |
0.25 Hz - 10 kHz |
0.25 Hz - 10 kHz |
0.25 Hz - 10 kHz |
0.25 Hz - 10 kHz |
0.25 Hz - 10 kHz |
| Pulse Width |
50 μs ±1% to 4 seconds ±1% |
50 μs ±1% to 4 seconds ±1% |
50 μs ±1% to 4 seconds ±1% |
50 μs ±1% to 4 seconds ±1% |
50 μs ±1% to 4 seconds ±1% |
50 μs ±1% to 4 seconds ±1% |
| Current Measurement 2 |
| Low Range/High Range |
3 A/30 A |
1 A/10 A |
6 A/60 A |
6 A/60 A |
12 A/120 A |
3 A/30 A |
| Low Range Accuracy3 |
0.05% + 3 mA |
0.05% + 2.5 mA |
0.05% + 5 mA |
0.05% + 5 mA |
0.05% + 10 mA |
0.05% + 3 mA |
| High Range Accuracy3 |
0.05% + 6 mA |
0.05% + 5 mA |
0.05% + 10 mA |
0.05% + 10 mA |
0.05% + 20 mA |
0.05% + 6 mA |
| Voltage Measurement 2 |
| Low Range/High Range |
6 V/60 V |
24 V/240 V |
6 V/60 V |
15 V/150 V |
6 V/60 V |
15 V/150 V |
| Low Range Accuracy |
0.05% + 3 mV |
0.05% + 10 mV |
0.05% + 3 mV |
0.05% + 8 mV |
0.05% + 3 mV |
0.05% + 8 mV |
| High Range Accuracy |
0.05% + 8 mV |
0.05% + 20 mV |
0.05% + 8 mV |
0.05% + 16 mV |
0.05% + 8 mV |
0.05% + 16 mV |
| Power Measurement 2 |
| Accuracy |
0.1% + 0.5 W |
0.1% + 1.2 W |
0.1% + 0.5 W |
0.1% +1.5 W |
0.1% +1.2 W |
0.1% +0.5 W |
- Maximum continuous power available is derated linearly from 100% of maximum at 40ºC, to 75% of maximum at 55ºC.
- 2 Specification is ±(% of reading + fixed offset). Measurement is 1000 samples. Specification may degrade when the unit is subject to an RF field of 3 V/meter, the unit is subject to line spikes of 500 V, or an 8 kV electrostatic discharge.
- 3 dc current accuracy specifications apply 30 seconds after input current is applied.
|