Agilent 8114A High Power, Single-Channel Pulse Generator 100V/2A

Click image to enlarge
Agilent 8114A
|
|
Features:
- 100Vpp(2A) into open (or from 1KΩ into 50Ω
- 7ns Transitions (fixed)
- 15MHz Repetition Rate
- Variable Pulse-Width
- Variable Pulse Baseline ±25 V(optional)
- 3 digit resolution
- SCPI Programming Commands
- External Triggering and Gathing
- Counted Burst Mode
- Graphic Display
|
Description
Today's telecommunications and computer systems increasingly take advantage of devices such as laser and IR diodes, EEPROMS and flash memories which operate at high voltages or currents.
- Fast, clean positive or negative pulses up to 100 V at frequencies up to 15 MHz
- User-definable limits for voltage, current and duty cycle
- TLL input: allows the output to be enabled or disabled
- Transitions < 12 ns for high impendance source into 50 Ohm load. Baseline OV.
- In 50 Ohm/50 Ohm mode, the maximum voltage is 50 Vpp with transitions < 7ns
8114A discontinuance is postponed to end of summer 2009.
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| 8114A |
100V 2A programmable pulse generator |
$25392.00 |
| $25138.08 |  |
|
| 81101A |
10V 50 MHz pulse generator |
$8990.00 |
| $8900.10 |  |
|
| 81104A |
80 MHz Pulse Generator |
$6254.00 |
| $6191.46 |  |
|
| 81110A |
165 / 330 MHz Pulse / Pattern Generator |
$9191.00 |
| $9099.09 |  |
|
| 81130A |
400/660 MHz Pulse / Pattern Generator |
| $13293.00 |  |
|
| Model |
Options |
List Price |
Sale Price |
| 8114A-001 |
Var. baseline and current programming |
$3483.00 |
| $3448.17 |  |
|
| 8114A-013 |
Front and rear panel connectors (front connected) |
$357.00 |
| $353.43 |  |
|
| 8114A-014 |
Front and rear panel connectors (rear connected) |
$357.00 |
| $353.43 |  |
|
| 8114A-1BP |
Mil std 45662A calibration with test data |
$1008.00 |
| $997.92 |  |
|
| 8114A-UK6 |
Commercial cal. certificate w/ test data |
$245.00 |
| $242.55 |  |
|
Specifications
Timing Characteristics
Measured at 50% amplitude at fastest transitions in continuous mode and 50Ω source impedance.
| Frequency range | 1Hz to 15.00MHz |
| Timing resolution (period, width) | 3 digits, 100ps best case |
| RMS jitter (period, width, delay) | 0.03% ±25 ps (±0.05% ±25 ps for period < 100ns) |
| Period Range | 66.7 ns to 999 ms |
| Accuracy | ±5% ±100 ps |
| Width Range | 10.0 ns to 150 ms (max: period - 10.0 ns) |
| Accuracy | ±5% ±500 ps |
| Fixed Delay | 42 ns typical {1} |
| Additional Variable Delay Range | 0.00 ns to 999 ms (max: period - 4 ns) |
| Accuracy | ±5% ±1 ns |
| Resolution | 3 digits, 10 ps best case |
| Transition Time (10/90) | 50Ω into 50Ω: <7ns (amplitude > 5V) High-Z into 50Ω: <12ns (amplitude > 10V) |
NOTE: [1] Measured between Trigger Output and Output
Level/Pulse Performance Characteristics
| | Agilent 8114A |
| Amplitude | 1.00 V to 50.0V or 20.0 mA to 1.00A (values approximately double for source or load resistance > 1 kΩ) |
| Accuracy | ±1% of amplitude ±100 mV |
| Resolution | 3 digits, best case 10 mV |
| Baseline |
| Standard Instrument | 100 mV + 0.5 of amplitude |
| Option 001 | -25.0 V to 25.0 V[1] |
| Accuracy | ±1% ±100 mV ±0.5% of amplitude |
| Pulse Window | ±50 V |
| Baseline Noise | 10 mV RMS typ. |
| Output Connectors | BNC |
| Baseline Noise | Selectable 50Ω or High Impedance (> 1kΩ) |
| Short Circuit Current | 2A |
| Overshoot/preshoot/ringing | < ±5 of amplitude ±100 mV |
| Settling Time | < 100 ns typical |
NOTE: [1] 50Ω source must be selected for variable baseline
|