Supplemental characteristics for all 668x model numbers
|
| dc Floating Voltage |
Output terminals can be floated up to ±60 Vdc maximum from chassis ground |
| Remote Sensing |
Up to half the rated output voltage can be dropped in each load lead. The drop in the load leads subtracts from the voltage available for the load. |
| Command Processing Time |
Average time required for the output voltage to begin to change following receipt of digital data is 20 ms for power supplies connected directly to the GPIB. |
| Modulation |
(analog programming of
output voltage and current) |
| Input Signal |
0 to –5 V for voltage,
0 to +5 V for current |
| Input Impedance |
30 k&Ohm;/or greater |
| ac Input (47 to 63 Hz) |
180 to 235 Vac (line-to-line, 3 phase), 27.7 A rms maximum worst case, 21.4 A rms nominal; 360 to 440 Vac, 14.3 A rms maximum worst case, 10.7 A rms nominal (maximum line current includes 5% unbalanced phase voltage condition.) Output voltage derated 5% at 50 Hz and below 200 Vac. |
| Input Power |
7350 VA and 6000 W maximum; 160 W at no load |
| GPIB Interface Capabilities |
SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, E1, and C0. IEEE-488.2 and SCPI command set. |
| Size |
425.5 mm W x 221.5 mm H x
674.7 mm D (16.75 in x 8.75 in x 25.56 in) |
| Weight |
Net, 51.3 kg (113 lbs) Shipping, 63.6 kg (140 lbs) |