Supplemental characteristics for all 663x 80-100W single output model numbers |
| dc Floating Voltage |
Output terminals can be floated up to ±240 Vdc maximum from chassis ground |
| Remote Sensing |
Up to two volts dropped in each load lead. Add 2 mV to the voltage load regulation specification for each one volt change in the positive output lead due to load current change. |
| Command-Processing Time |
Average time required for the output voltage to begin to change following receipt of digital data is 4 ms for the power supplies connected directly to the GPIB. (Display disabled). |
| Output-Programming Response Time |
The rise and fall time (10/90% and 90/10%) of the output voltage is less than 2 ms (400 μs in fast mode). The output voltage change settles within 1 LSB (0.025% x rated voltage) of final value in less than 6 ms (2 ms in the fast mode). |
| GPIB Interface Capabilities |
IEEE-488.2, SCPI command set and 6630A Series programming compatability |
| Measurement Time |
Average time to make a voltage or current measurement is 50 ms. |
| Input Power (full load) |
3.5 A, 250 W |
| Regulatory Compliance |
Complies with EMC directive 89/336/EEC (ISM 1B) |
| Warranty Period |
One year |
| Size |
425.5 mm W x 88.1 mm H x 364.4 mm D (16.8 in x 3.5 in x 14.3 in) |
| Weight |
Net, 12.7 kg (28 lb) net; 15.0 kg (33 lb) shipping |