Agilent 6634B 100 Watt Power Supply

Click image to enlarge
Agilent 6634B
|
|
Features:
- High throughput
- Precision low current measurement
- Low-output noise
- High-speed programming
- GPIB and RS-232 interface
- VXI plug&play drivers
- Order option J07 for use with inductive loads
Warranty:
1 year
|
Description
The Agilent 6634B System dc Power Supply is an ideal power source for demanding production test applications. This dc power supply offers high performance and precision measurements in a one-box solution. To reduce test-system development time and provide precision testing, the 6634B provides current measurement down to the microamp level. The 6634B also provides high-speed programming with under 4 millisecond response time.
This series of linear-regulated 80-100 W dc power supplies is designed to maximize the throughput of DUTs through the manufacturing test process. Both programming and measurement are optimized for speed. The active downprogrammer can sink up to the full rated current of the power supply, which quickly brings the power supply output to zero volts. The 6630B Series offers many advanced programmable features including stored states and status reporting. Programming is done using industry standard SCPI commands via the GPIB or RS-232. Test system integration is further simplified by using the VXIPlug&Play drivers. The optional relays simplify system design and troubleshooting.
The optional front panel binding posts make the 6630B Series convenient on the R&D lab bench. The built-in microamp measurement system helps the engineer to easily and accurately monitor the output voltage and current without a complicated test setup.
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| 6631B |
DC power supply, single-output. 0-8 V, 0-10 A, 80 W. GPIB, RS-232. |
| $2587.20 |  |
|
| 6632B |
DC power supply, single-output. 0-20 V, 0-5 A, 100 W. GPIB, RS-232. |
| $2587.20 |  |
|
| 6633B |
DC power supply, single-output. 0-50 V, 0-2 A, 100 W. GPIB, RS-232. |
| $2587.20 |  |
|
| 6634B |
DC power supply, single-output. 0-100 V, 0-1 A, 100 W. GPIB, RS-232. |
| $2587.20 |  |
|
| Model |
Options |
List Price |
Sale Price |
| 6634B-1CM |
Rackmount kit (p/n 5062-3974) |
| $43.46 |  |
|
| 6634B-1CP |
Rack mount and handle kit (p/n 5062-3975) |
| $99.51 |  |
|
| 6634B-760 |
DC isolation and polarity reversal relay |
| $217.35 |  |
|
| 6634B-A6J |
ANSI/NCSL Z540 calibration certificate |
| $366.45 |  |
|
| 6634B-J07 |
Stabilization for inductive loads |
| $270.90 |  |
|
| 6634B-UK6 |
Commercial calibration with test result data |
| $154.35 |  |
|
Specifications
(at 0º to 55ºC unless otherwise specified)
Supplemental characteristics
| |
6631B |
6632B |
6633B |
6634B |
| Number of outputs |
1 |
1 |
1 |
1 |
| GPIB |
Yes |
Yes |
Yes |
Yes |
| Output ratings |
| Voltage |
0 to 8 V |
0 to 20 V |
0 to 50 V |
0 to 100 V |
| Current |
0 to 10 A |
0 to 5 A |
0 to 2 A |
0 to 1 A |
| Programming accuracy at 25ºC ±5ºC |
| Voltage |
5 mV |
10 mV |
20 mV |
50 mV |
| + Current |
0.05% + |
4 mA |
2 mA |
1 mA |
0.5 mA |
| Ripple and noise (20 Hz to 20 MHz, with outputs ungrounded or with either terminal grounded) |
| Voltage Normal mode |
rms peak-to-peak |
0.3 mV 3 mV |
0.3 mV 3 mV |
0.5 mV 3 mV |
0.5 mV 3 mV |
| Fast mode |
rms peak-to-peak |
1 mV 10 mV |
1 mV 10 mV |
1 mV 15 mV |
2 mV 25 mV |
| Current |
rms |
3 mA |
2 mA |
2 mA |
2 mA |
| dc measurement accuracy via GPIB or front panel meters with respect to actual output at 25a86;C ±5ºC |
| Voltage |
0.03% + |
2 mV |
3 mV |
6 mV |
12 mV |
| Low current range |
–20 mA to +20 mA |
0.1% + |
2.5 μA |
2.5 μA |
2.5 μA |
2.5 μA |
| High current range |
+20 mA to + rated I –20 mA to –rated I |
0.2% + 0.2% + |
1 mA 1.6 mA |
0.5 mA 1.1 mA |
0.25 mA 0.85 mA |
0.25 mA 0.85 mA |
| Load regulation |
| Voltage |
2 mV |
2 mV |
4 mV |
5 mV |
| Current |
2 mA |
1 mA |
1 mA |
1 mA |
| Line regulation |
| Voltage |
0.5 mV |
0.5 mV |
1 mV |
1 mV |
| Current |
1 mA |
0.5 mA |
0.25 mA |
0.25 mA |
| Transient response time Less than 100 μs (50 μs in the fast mode) for the output voltage to recover to its previous level (within 0.1% of the voltage rating of the supply or 20 mV) following any step change in load current of up to 50% of the output current rating of the supply. |
|