Agilent 34410A 6 1/2 Digit High Performance Digital Multimeter

Click image to enlarge
Agilent 34410A
|
|
Features:
- 6 1/2 digit resolution
- 10,000 readings/sec @ 5 1/2 digits to the PC
- LAN, USB & GPIB standard
- Capacitance & temperature measurements
- Data logger for improved usability
- 50,000 reading non-volatile memory
- Designed to comply with LXI Class C specification
Includes:
- (1) 34410A Digital Multimeter, 6 1/2 Digit High Performance
- (1) Power Cord
- (1) Test Lead Kit
- (1) USB Cable
- (1) Certificate of Calibration with Test Report
- Product Reference CD
- IO Libraries Media Suite CD
|
Description
The Best Just Got Better The Agilent 34410A and 34411A 61/2 Digit DMMs represent the latest generation of multimeters from Agilent Technologies. Building on the phenomenal success of the industry-standard Agilent 34410A, these new meters offer improved accuracy, expanded measurement capability, dramatically improved measurement speed and throughput, and modern computer interfaces including LAN and USB. The dual display offers both dual measurement capabilities and ease of use when setting up and configuring the DMM. Improvements have been made in every facet of the 34410A to make the best even better, whether you use it on the bench or in a system.
Dramatic Speed Improvements Dramatic Speed Improvements Whether it’s raw reading speed or fast system throughput, the 34410A sets a new benchmark in performance. Using a new A/D technology, the 34410A achieves an impressive 10,000 readings a second at 51/2 digits, and can stream readings to your computer at this same speed! Triggering is fast and precise, with both trigger latency and trigger jitter less than 1μ, while bus query response is less than 500 μ. ACV measurements are faster as well thanks to a digital measurement technique that additionally improves accuracy at high and low frequencies. For even greater reading speeds, select the 34411A, which achieves 50,000 readings a second at 41/2 digits.
Enhanced Measurement Performance The 34410A and 34411A offer Temperature and Capacitance capabilities, in addition to those measurements you have come to expect, such as DCV, ACV, DCI, ACI, 2-wire and 4-wire Resistance, Frequency, Period, Continuity and Diode Test. You also get Offset Compensated Ohms, allowing you to accurately measure resistance in the presence of voltages. Measurement ranges have been expanded as well; for example, DC and AC Current Ranges now go down to 100 μ, resulting in 100 pA resolution. Real-time math and statistics are included, and a peak-detect capability allows you to capture peaks as short as 20 μ.
Even Greater Performance
with the 34411A The 34411A has all the features of the 34410A, plus additional performance that makes it even more powerful. With the ability to make 50,000 readings per second at 41/2 digits, Analog Level Triggering, programmable Pre- and Post- Trigger and 1 million readings of volatile memory in addition to 50,000 reading of non-volatile memory, you now have the ability to capture low-frequency waveforms, characterize device performance and transfer results for analysis on your computer.
Data Logger Function A front panel data logger function allows you to set the meter up to make unattended, paced measurements over a fixed time or number of events, then pull up the results later for review or transfer to a computer for analysis. Set the meter up to take measurements every 10 seconds for an hour, go have lunch, and check the results upon your return. The contextual front panel sequences make setup and read back a breeze.
Improved Ease of Use From the inclusion of a second display to configuring setups for each of the measurement functions, these new DMMs offer significant improvements in usability. Simple things are simple; more complicated setups are easier than ever. There is even a new probe set designed to more easily probe today’s fine-pitch components. Finally, there is a built-in Graphical Web Interface that allows you to interactively control the DMM without the hassle of programming at all!
Modern I/O for Improved Connectivity When connecting to a computer, select LAN, USB or GPIB interfaces; all three are standard on the 34410A and 34411A. Concerned about the viability of your existing software programs? These new DMMs respond to Standard Commands for Programmable Instrumentation (SCPI), and there is even a 34410A Emulation Mode to ensure the easiest upgrade possible. Agilent’s I/O Library Suite ships with the 34410A and 34411A to help you quickly establish an error-free connection between your PC and instrument. It provides robust instrument control and works with the software development environment you choose.
LXI - Class C LAN Extensions for Instruments (LXI) provides the next generation I/O for system applications requiring the highest throughput. Transfer rates of well over 100,000 readings/sec are attainable ensuring even the most data intensive measurements are fast, without the overhead cost of an instrument cardcage. Both the 34410A and the 34411A have been designed for LXI - Class C compliance.
Built To LastOur new DMMs were designed to high standards of ruggedness and reliability. From the robust package with its shock absorbing bumpers to careful selection of components and conservative circuit design, these meters are built to last. Calculated Mean Time Between Failure (MTBF) is in excess of 100,000 hours. Backed by a 1-year warranty and a worldwide network of service centers, you can buy with confidence.
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| 34410A |
34410A - Digital multimeter, 6.5 digit |
| $1295.00 |  |
|
| 34411A |
34411A - Digital multimeter/Digitizer, 6.5 digit |
| $2003.00 |  |
|
| Model |
Options |
List Price |
Sale Price |
| 11059A |
Kelvin probe set |
| $156.00 |  |
|
| 11060A |
S.M.D. test probe |
| $30.56 |  |
|
| 11062A |
Kelvin clip |
| $28.89 |  |
|
| 34131A |
Transit case for half-rack 2U high instruments (e.g., 34401A) |
| $301.00 |  |
|
| 34162A |
Accessory pouch for 33220A, 34410A and 34411A |
| $43.66 |  |
|
| 34171A |
Input terminal block for the 34401A DMM (qty. two per) |
| $103.00 |  |
|
| 34172A |
Input calibration short for the 34401A DMM (qty. two per) |
| $103.00 |  |
|
| 34330A |
Current Shunt, 30 Amp, for HP Dmm's |
| $80.76 |  |
|
| E2308A |
Thermistor temperature probe 0-70 degrees C, 5k ohm |
| $61.12 |  |
|
| 5061-9694 |
Please call or email us for pricing and availability. |
| 5063-9212 |
Please call or email us for pricing and availability. |
| 34410A-1CM |
Rackmount Kit |
| $61.00 |  |
|
| 34410A-A6J |
ANSI Z540 compliant calibration |
| $155.00 |  |
|
Specifications
| DC Voltage |
| Measurement Method |
Continuously integrating multi-slope IV A/D converter |
| Linearity |
0.0002% of reading (10 V range) + 0.0001% of range |
| Input Resistance |
0.1 V, 1 V, 10 V 10 MΩ or > 10 GΩ Ranges (Selectable) 100 V, 1000 V 10 MΩ ±1% Ranges (Fixed) |
| Input Bias Current |
< 30 pA at 25°C |
| Input Protection |
1000 V |
| DC CMRR |
140 dB1 |
| True RMS AC Voltage |
| Measurement Method |
AC-coupled True RMS measurement. Digital sampling with anti-alias filter. |
| Crest Factor |
No additional error for crest factors < 10. Limited by peak input and 300 kHz bandwidth. |
| Peak Input |
300% of range or 1100 V |
| Overload Ranging |
Will select higher range if peak input overload is detected during auto range. Overload is reported in manual ranging. |
| AC CMR |
70 dB2 |
| Maximum Input |
400 Vdc, 1100 Vpk |
| Input Impedance |
1 MΩ ± 2% in parallel with < 150 pF |
| Input Protection |
750 Vrms all ranges |
| Resistance |
| Measurement Method |
electable 2-wire or 4-wire. Current source referenced to LO input. |
| Offset Compensation |
Selectable on the 100 Ω, 1 kΩ, and 10 kΩ ranges |
| Max. Lead Resistance (4-wire) |
10% of range per lead for 100 Ω, 1 kΩ. 1 kΩ per lead on all other ranges. |
| Input Protection |
1000 V on all ranges |
| DC Current |
| Current Shunt |
200 Ω for 100 μA, 1 mA 2 Ω for 10 mA, 100 mA 0.1 Ω for 1 A, 3 A |
| Input Protection |
3 A, 250 V fuse |
| True RMS AC Current |
| Measurement Method |
AC-coupled True RMS measurement. Directly coupled to the fuse and shunt. Digital sampling with anti-alias filter. |
| Current Shunt |
200 Ω for 100 μA, 1 mA 2 Ω for 10 mA, 100 mA 0.1 Ω for 1 A, 3 A |
| Maximum Input |
The peak value of the DC + AC current must be <300% of range. The RMS current must be <3 A including the DC current content. |
| Input Protection |
3 A, 250 V fuse |
| Frequency and Period |
| Measurement Method |
Reciprocal-counting technique. AC-coupled input using the AC voltage measurement function. |
| Input Impedance |
1 MΩ ± 2% in parallel with < 150 pF |
| Input Protection |
750 Vrms all ranges |
| Capacitance |
| Measurement Method |
Current input with measurement of resulting ramp. |
| Connection Type |
2-wire |
| Temperature |
| Thermistor |
2.2 kΩ, 5 kΩ, and 10 kΩ |
| RTD |
a = 0.00385 Ro from 49 Ω to 2.1 kΩ |
| Continuity/Diode Test |
| Response Time |
300 samples/sec with audible tone |
| Continuity Threshold |
Fixed at 10 Ω |
| Triggering and Memory |
| Reading Hold Sensitivity |
1% of reading |
| Samples per Trigger |
1 to 50,000 (34410A) 1 to 1,000,000 (34411A) |
| Trigger Delay |
0 to 3600 sec (20 μs step size) |
| External Trigger |
Programmable edge, Low-power TTL compatible |
| Delay |
< 1 μs Max rate: 5,000/sec |
| Jitter |
> 1 μs Min Pulsewidth: 1 μs |
| Voltmeter Complete |
3 V Logic output, 2 μs pulse with programmable edge |
| Nonvolatile Memory |
50,000 readings |
| Volatile Memory |
50,000 readings (34410A) 1,000,000 readings (34411A) |
| Sample Timer |
| Range |
0 to 3600 sec (20 μs step sizes) |
| Jitter |
< 100 ns |
| General Specifications |
| Power Supply |
100 V/120 V/220 V/240 V ±10% |
| Power Line Frequency |
45 Hz to 66 Hz and 360 Hz to 440 Hz, Automatically sensed at power-on |
| Power Consumption |
25 VA peak (16 W average) |
| Operating Environment |
Full accuracy for 0°C to 55°C, 95% R.H. at 40°C non-condensing |
| Storage Temperature |
-40°C to 70°C |
| Weight |
3.72 kg (8.2 lbs) |
| Safety |
IEC 61010-1, EN 61010-1, UL 61010-1, CAN/CSA-C22.2 No. 61010-1, Refer to Declarations of Conformity for current revisions. Measurement CAT II 300 V, |
| CAT I 1000 V. Pollution Degree 2 EMC |
IEC 61326, EN 61326, CISPR 11, ICES-001, AS/NZS 2064.1, Refer to Declaration of Conformity for current revisions. |
| Vibration & Shock |
MIL-T-28800E, Type III, Class 5 (Sine Only) |
| Warranty |
1 year |
|