Agilent 34405A 5 1/2 Digit Digital Multimeter

Click image to enlarge
Agilent 34405A
|
|
Features:
- 120000 counts resolution
- 16 built-in measurement functions including temperature and capacitance
- 8 built-in math functions; Null, dBm, dB, MinMax, Limit and Hold
- Up to 70 readings per second
- 0.025% DC voltage accuracy
- USB 2.0 interface (TMC-488.2 compliant)
- SCPI compatible
- Agilent IO Library Suite, DIMM Intuilink connectivity software, and IVI-COM and LabVIEW drivers
Includes:
- (1) 34405A Digital Multimeter
- (1) Test Lead Kit
- (1) Test Report
- (1) USB Interface Cable
- Quick Start Guide
- Product Reference CD-ROM
- Agilent IO Library Suite CD-ROM
|
Description
Affordable and Feature-RichMeasurement Tool The 34405A represents the latest member in the Agilent’s DMM family and this expands Agilent’s offerings in theelectronics measurement tools. It provides a broad range of features and measurement functions such as DCvoltage, DC current, true-RMS AC voltageand AC current, 2-wire resistance,frequency, diode test and continuity which are designed to meet general industrial needs. Furthermore, it can measure temperature ranging from -80oC to 150oC. The true value is more evident with its capability to measure capacitance ranging from 1000pF to 10000mF. Agilent 34405A also improves efficiency and accuracy with its 6 built-in math operations: Null, dBm, dB, MinMax, Limit and Hold.
Quick Connection to the PC with USB 2.0 Interface For those with a need to control and take preset measurements with a PC, the builtin USB 2.0 interface provides an easy and robust connection between the PC and DMM. The USB interface that is compliant with the TMC-488.2 Standards, works seamlessly with Agilent Connectivity software and can be controlled remotely via industry standard SCPI commands or through DMM Intuilink Connectivity software. IVI-COM and LabVIEW drivers are included to ensure an easy integration with different programming environments.
Bright Display, Fast Reading Speed and Configuration Storage When high throughput and productivity are the priority, Agilent 34405A VFD dual display feature allows users to take more than one measurement and display them simultaneously on the front panel. For speed critical applications, Agilent 34405A can take up to 19 readings/sec at 4.5 digits resolution directly to the PC. In addition, the user can configure and store complete instrument setups and recall them at anytime from any of the four built in storing states.
Rugged and Reliable The 34405A is designed and tested according to major Safety and Regulatory Standards. In addition, the shock absorbing bumpers are designed to prevent physical damage from your day-to-day use.
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| 34405A |
Digital multimeter, 5.5 digit |
| $749.00 |  |
|
| Model |
Accessories |
List Price |
Sale Price |
| 34132A |
Deluxe test lead set for 34401A |
| $49.04 |  |
|
| 34133A |
Precision electronic DMM test lead |
| $27.78 |  |
|
| 34330A |
Current Shunt, 30 Amp, for HP Dmm's |
| $80.76 |  |
|
| E2308A |
Thermistor temperature probe 0-70 degrees C, 5k ohm |
| $61.12 |  |
|
Specifications
| DC Voltage |
| Measurement Method |
Sigma Delta A-to-D converter |
| Input Resistance |
10MW\2% range (typical) |
| Input Protection |
1000V on all ranges/td>
|
| Resistance |
| Measuring Method |
2-wire Ohms |
| Open circuit voltage |
Limited to < 5V |
| Input Protection |
1000V on all ranges |
| DC Current |
| Shunt Resistance |
0.1W to 10W for 10mA to 1.2A ranges 0.01W for 12A range |
| Input Protection |
200 Ω for 100 μA, 1 mA Front Panel: 1.25A, 500V fuse for I terminal Internal: 15A, 600V fuse for 12A terminal |
| Continuity/Diode Test |
| Measuring Method |
Uses 0.83mA\0.2% constant current source, <5V open circuit voltage |
| Response Time |
70 samples/second with audible tone |
| Continuity Threshold |
10W fixed |
| Input Protection |
1000V |
| Temperature |
| Measurement Method |
2-wire Ohms measurement of 5kW thermistor sensor (YSI 4407) with computer conversion Auto-ranging measurement, no manual range selection |
| Input Protection: |
1000V |
| Measurement Noise Rejection |
| CMRR (Common Mode Rejection) for 1kW unbalance LO lead |
DC: 120 dB AC: 70 dB |
| NMR (Normal Mode Rejection) For 60 Hz (50Hz) \ 0.1% |
5 1/2 digit: 65 dB (55 dB) 4 1/2 digit: 0dB |
| AC Voltage |
| Measurement Method |
AC coupled true-RMS - measure the ac component with up to 400 Vdc bias any range |
| Crest Factor |
Maximum 5:1 at full scale |
| Input Impedence |
1MW\2% in parallel with < 100pF of all ranges |
| Input Protection |
750Vrms on all ranges |
| AC Current |
| Measuring Method |
DC coupled to the fuse and current shunt, AC coupled true-RMS measurement (measure the AC component only) |
| Shunt Resistance |
0.1W to 10W for 10mA to 1.2A range 0.1W for 12A range |
| Input Protection |
Externally accessible 1.25A, 500V fuse for I terminal Internally replaceable 15A, 600V fuse for 12A terminal |
Frequency |
| Measurement Method |
Reciprocal counting technique. AC coupled input using AC voltage function. |
| Signal Level |
10% of range to full scale input on all ranges Auto or manual range selection |
| Gate Time |
0.1 second or 1 period of the input signal, whichever is longer. |
| Input Protection |
750Vrms on all ranges |
| Technical Specifications |
| Math Functions |
Null, dBm, dB, Min/Max/Avg, Hold, Limit Test |
| Trigger and Memory |
Single trigger, 1 reading memory |
| Remote Interface |
USB 2.0 full speed, USBTMC class device (GPIB over USB) |
| Programming Language |
SCPI, IEEE-488.1, IEEE-488.2 |
| General Specifications |
| Power Supply |
100V/120V(127V)/220V(230V)/240V \ 10% AC line frequency 45Hz - 66Hz and (360Hz - 440Hz, 100/120V operation) |
| Power Consumption |
16VA maximum, <11W average |
| Operating Environment/td>
| Full accuracy at 0 oC to 55 oC Full accuracy to 80% RH at 30 oC (non-condensing) Altitute up to 3000 meters |
| Storage Compliance |
- 40 oC to 70 oC |
| Safety Compliance< |
Certified to CSA for IEC/EN/CSA/UL 61010-1 2nd Edition |
| Measurement Category |
CAT II, 300V: CAT I 1000Vdc, 750Vac rms, 2500Vpk transient over voltage Pollution degree 2 |
| EMC Compliance |
Certified to IEC/EN 61326:2002, CISPR 11, and equivalents for Group 1, Class A |
| Shock and Vibration |
Tested to IEC/EN 60086-2 |
| Dimension(HxWxD) |
Rack: 88.5mm x 212.6mm x 272.3mm Bench: 103.8mm x 261.2mm x 303.2mm |
| Weight |
3.75 kg, 8.27 lb |
| Warm up time |
30 Minutes |
| Warranty |
1 year |
|