AEMC H216 2129.63 Hipot Tester

Click image to enlarge
AEMC 2129.63
|
|
Features:
- AC Hipot programmable to 5kV @ 40mA
- DC Hipot programmable to 6kV @ 10mA (H216)
- Continuity measurement @ 100mA
- Insulation resistance to 9900MW @ 500VDC and 1000VDC and to 1990mW @ 50V/100V
- ARC current detection with programmable limits
- Programmable ramp, test and wait times, output voltages and trip limits
- Selectable AC test frequency (50Hz or 60Hz)
- Set trip current without loads or high voltage present
- Automatic testing - store up to 10 tests (up to 16 steps per test)
- Line and load regulation for linear output voltage
- Highly visible warning light indicator for HV output
- Pass and fail lights with adjustable buzzer
- Program lock-out feature to prevent user error or program changes
- PLC remote control (9 pin connector); Serial communication port (9 pins)
- Includes tester, high voltage test pistol, return lead with retainer, US AC 110V power chord, and manual
|
Description
|
The Hipot Tester Model H216 is designed for AC/DC Withstanding Voltage Testing and Insulation Resistance Testing (IR) providing a safe and accurate test environment for the operator. The Model H216 complies with the requirements of the electrical equipments and appliance testing standard, UL, CSA and other international standards. The Model H216 can be used for withstanding voltage tests of the various types of electrical equipment and components. The Model H216 provides AC hipot programmable to 5kV @ 40mA and DC hipot programmable to 6kV @ 10mA.
|
Models, Options, & Accessories
| Model |
Description |
List Price |
Sale Price |
| 2129.63 |
Hipot Model H216 (5kVAC/6kVDC and IR w/SP02) - Discontinued - suggested replacement: Vitrek V63 |
| 2129.60 |
Hipot Model H111 (5kVAC w/SP02) - Discontinued - suggested replacement: Vitrek V61 |
| Model |
Accessories |
List Price |
Sale Price |
| 2117.89 |
Lead HV Test Lead 9 ft (unterminated, 3m) for use with Hipot Testers - Discontinued - suggested replacement: Vitrek TL-20-60 |
| 2119.45 |
Cable PC RS-232, DB9 F/F 6 ft Null Modem Cable (for use with Megohmmeter Models 1060/5060/5070) |
$16.95
 |
| You may qualify for better pricing. |
| Please call 877-742-TEST (8378) |
| or Email for details. |
|
|
Specifications
| AC Hipot |
| Voltage Range |
100V to 5000V AC (5kV) |
| Voltage Step Resolution |
5V |
| Voltage Regulation |
±1% of Reading ± 5V |
| Accuracy |
±1% of Reading ± 10V (>500V) |
| Test Frequency ; 50Hz or 60Hz (selectable) |
Test Frequency ; 50Hz or 60Hz (selectable) |
| Current |
0.10mA to 40mA in 0.02mA steps |
| Max Current Output |
40mA @ 5kV |
| Accuracy |
±1% of Reading ±50V/A |
| Ramp and Test Time |
0.000s to 999.9s |
| DC Hipot |
| Voltage Range |
100V to 6000V DC (6kV) |
| Voltage Step Resolution |
5V |
| Voltage Regulation |
±1% of Reading ±5V |
| Accuracy |
±1% of Reading ±10V(>500V) |
| Trip Current |
0.10mA to 10mA in .01mA steps |
| Max Current Output |
10mA @ 6kV DC |
| Accuracy |
±1% of Reading ± 50µA |
| Ramp and Test Time |
0.000s to 999.9s |
| Insulation Resistance |
| DC Test Voltage |
50 & 100V; 500V & 1000V |
| Measurement Range |
1 to 1990MW / 1 to 9900MW |
| Accuracy |
1 to 500MW ±5% of Reading / 501 to 2000MW ±10% of Reading |
| Test Time |
0.000s to 999.9s |
| Continuity Check |
| Resistance Range |
0.001W to 1.500W |
| Test Current |
100mA (0.1A) |
| Detect Accuracy |
0.1W @ 1W |
| Test Time |
0.000s to 999.9s |
| Arc Detection |
| Detect Current |
Selectable in 1mA steps from 2 to 40mA |
|